Instrument :
Serial Number:
HMI SW Version:
SMART SW Version:
Department:
Tel: Email:
Distributor in charge:
Tel: Email:
IQP-TF Test Form IQ -TF01
DAYmate S PRE-REQUIREMENTS
A preventive
2 maintenance plan is Date of the next
or will be available maintenance
each 6 months
The completeness of
The packaging list is
6 the checklist has
complete.
been confirmed with
(form 02310)
date and signature.
IQP-TF Test Form IQ -TF01
DAYmate S SECURITY INSPECTION
The DAYmate S is
located in a place Around the machine
with enough place is a minimal place of
4 for ventilation to 150 mm.
avoid overheating.
All software and
The products are
accessories used
delivered and/or
5 with the DAYmate S
officially approved
are approved from
from DAY medical.
DAY medical.
Every PC, which will An antivirus and
be connected to firewall is installed
6 the DAYmate S, is on the PC.
protected against (only if connected to
Virus and Malware. internet)
All sources of
The DAYmate S is
electromagnetic
not under influence
7 emissions with high
of electromagnetic
intensity are not in
waves of high
proximity of the
intensity.
DAYmate
Connect the · A green check
DAYmate S to a appears above in the
8 power outlet and middle on the HMI.
start the machine · No error messages
completely (both / warnings appear in
power switches) the fixed-bar.
IQP-TF Test Form IQ -TF02
DAYmate S RESULTS
All the Finding are under control (fill out the table “List of finding”) Yes No
Non-blocking
N/A
blocking
Non-blocking
N/A
blocking
Non-blocking
N/A
blocking
Non-blocking
N/A
blocking
Non-blocking
N/A
IQP-TF Test Form IQ -TF02
DAYmate S RESULTS
Final Result
passed Yes
Failed No
Yes
No
Remarks: ……..……………………………………………………………………………………………………………………………………
………………………………………………………………………………………………………………………………………………………….
………………………………………………………………………………………………………………………………………………………….
………………………………………………………………………………………………………………………………………………………….
………………………………………………………………………………………………………………………………………………………….
Distributor
Final customer