Professional Documents
Culture Documents
DOCUMENT TITLE
DOCUMENT NUMBER
HT –PAUT-ASME-01
PREPARED
M.SATHIASEELAN
& PCN NDT LEVEL II-UT/PAUT
REVIEWED BY
S.ESWARAN
APPROVED BY ASNT NDT LEVEL III
RT/UT/MT/PT/ET/VT
TECHNICAL MANAGER
REVISION STATUS
TABLE OF CONTENTS
1.0 PURPOSE
3
2.0 SCOPE
3
3.0 REFERENCES
3
4.0 ULTRASONIC PHASED ARRAY EXAMINATION EQUIPMENT
3
5.0 PHASED ARRAY INSTRUMENT LINEARITY
5
6.0 PHASED ARRAY PROBE ELEMENT OPERABILITY VERIFICATION
6
7.0 PHASED ARRAY SYSTEM CALIBRATION
6
8.0 SURFACE PREPARATION
10
9.0 EXAMINATION COVERAGE AND SCANNING TECHNIQUE
11
10.0 RECORDING AND AMPLITUDE DETERMINATION
12
11.0 REPORT OF EXAMINATION
14
12.0 POST EXAMINATION CLEANING
15
13.0 DOCUMENTATION
15
APPENDICES
17
APPENDIX-1: PHASED ARRAY SCAN PLANS- FULL VOLUMETRIC COVERAGE PIPE TO PIPE
WELDS ....................................................................................................................................................................18
APPENDIX-2 : TYPICAL GRAPHICAL SCAN PLAN FOR PIPE WELDS...............................................................19
APPENDIX-3 : BASIC CALIBRATION BLOCK.............................................................................................................21
APPENDIX-4 : PHASED ARRAY CALIBRATION & QUALIFICATION BLOCK.....................................................22
APPENDIX-5 : ACCEPTANCE CRITERIA -ASME SEC VIII DIV.1.............................................................................23
APPENDIX- 6: SAMPLE PAUT INSPECTION REPORT.............................................................................................24
1.0 PURPOSE
1.1 This procedure provides requirements for Semi-Automatic OR fully automated with
encoded Ultrasonic Examination of pipe welds and base materials using Phased
Array.
1.2 Project specific procedures complying with ASME Sec VIII DIV 1 requirements are
attached in the Appendices.
2.0 SCOPE
2.1 This procedure establishes the specific phased array ultrasonic examination
requirements that shall be used to examine Carbon & Alloy Steel pipe to pipe welds
prior to or after Post Weld Heat Treatment (PWHT).
2.2 This procedure is applicable for welds that are 8.0 mm and above in thickness with
outer diameter (OD) from 4 inch and above.
2.3 This procedure is designed to demonstrate the Olympus NDT Omni Scan Phased
Array System used in this procedure as qualified UT System in accordance with the
ASME Sec VIII DIV.1.
2.4 A non-blind test shall be used as a Performance Demonstration for the Omni Scan
Phased Array System.
2.5 The Phased Array Scan Plan for each demonstration is outlined in the Appendix 2 &4
with specific requirements for this job.
3.0 REFERENCES
3.1 ASME Sec VIII Division 1: Rules for construction of pressure vessel.
3.2 ASME Boiler and Pressure Vessel Code, Section V, Article 4, Edition 2010.
3.3 ASME Code for Pressure Piping, B31.3 (Latest Edition)
3.4 BV Rules Part B Ch 12 Section 1
3.5 HT-WP-01 (2010) - Company written practice for qualification and Certification of
NDE personnel.
4.3 Examination personnel may use a real-time sector scan (also called an S-Scan) image
during scanning to assure that proper data has been collected. Sector scan images contain
signal amplitude and reflector depth information projected for the refracted angle of the
ultrasonic beam. The Omni Scan Phased Array system provides a variety of analysis
capabilities including A-scan display and parameter readout associated with software
cursors. Images produced by B, C and Sectorial Scan images are a useful aid in evaluation.
4.4 The Omni Scan Phased Array system has on-board focal law generation software that
permits direct modification to ultrasonic beam characteristics. The Omni Scan Phased Array
System requires the use of an external storage device, or flash card or USB memory Stick. A
remote portable PC connected via Ethernet may be used for this purpose.
4.5 In addition to data storage, the PC will also be used by the Data Analysis Personnel for
analyzing data subsequent to the completion of data collection. Data display software
compatible to that residing on the Omni Scan Phased Array System will also be used on the
remote PC for data playback. Reference the manufacturer operating manuals for instrument
operation specifics.
4.6 Any control, which affects the instrument linearity (e.g., Reject) shall be in the off or
minimum position for instrument calibration, system calibration and examination.
4.7 If any control (e.g., filters, averaging, pulse duration, etc.) are used in calibration, then
these controls shall not be adjusted afterwards since they may affect the Omni Scan System
Calibration.
4.8 The Ultrasonic Phased Array System shall be calibrated for linearity in accordance with
Paragraph 5.0.
4.9 Ultragel II, Sonotrace 40, Sonatech, glycerine, diluted Wall paper paste, water or
equivalent may be used as couplant when performing calibrations and examinations.
4.10 The same couplant material used for calibration shall be used for examinations.
4.11 Ultrasonic transducer configurations are specified by the technique used for
examination. Linear Phased Array Probe configurations may include from 10 to 128 elements.
4.12 The Phased Array probe frequency shall be between 2 and 10 MHz depending
upon material type, thickness.
4.13 Phased Array Wedges should be of a design to accommodate the aforementioned
Phased Array Probes. Nominal refracted wedge angles shall be 45, 55, 60 or 70 degree to
ensure coverage of the weld and heat affected zone (HAZ).
4.14 An encoder interfaced with the Phased Array instrument may be used to track the
Phased Array Probe movement. The encoder shall be calibrated by moving a minimum
distance of 500mm and the distance recorded shall be within +/- 1% of actual distance
moved. After the initial verification, further verification shall be performed not exceeding one
month from the first test and the same test shall be repeated.
7.5.5 With the transducer remaining in the peaked position, measure the metal
path of the 100 mm radius reflector using a cursor in the A-scan Display.
7.5.6 The value should measure to be 100 +/- 2.5 mm. If this measurement is
less than 97.5 mm, increase the value of the Delay parameter until the
measurement is correct. If this value is greater than 102.5 mm, decrease the
Delay parameter until the measurement is correct. The requirements for focal
law verification, time base verification and sensitivity adjustments are listed
below. Instrument linearity verification and ultrasonic beam spread is not
required.
7.5.7 As an alternative, other calibration blocks may be used to perform the Time
Base or Wedge Delay calibration.
7.6.4.1 Peak up this signal from the calibration reflector and scan the
phased array probe backwards through all the different angles or focal
laws.
7.6.4.2 Scan forward over the calibration reflector through all the
refracted angles or focal laws.
7.6.4.3 The Omni Scan system will calculate the required gain needed at
each focal law to adjust the amount needed
7.7 A Time Corrected Gain (TCG) calibration shall be used to compensate for attenuation
in the material at the sound paths utilized during calibration and examination.
7.8 As an alternative, DAC may be used for electronic Scans (E-Scans) of specific angles,
e.g., 45, 60 or 70 degrees.
7.9 The examination system calibration may be stored in the Omni Scan System with
electronic memory or on an external chip or data storage device. This calibration may
be used at a later date provided the system calibration is verified prior to examination.
7.10 A complete Ultrasonic System Calibration shall be performed at least once prior to the
examination.
8.2 Weld Surfaces - the weld surface should be free of irregularities that could mask or
cause reflections from defects to go undetected and should merge smoothly into the
adjacent base materials.
8.3 Conditions which do not meet these requirements shall be recorded as limitations on
the Ultrasonic Examination Data Report Form.
8.4 Testing of Base Material & Transverse Defects :
8.4.1 Following inspection shall be considered when appropriate prior to PA
scanning:
8.4.2 Prior to PA inspection, the detection for lamination or inclusion in the base
material need not be foreseen, when this has already been carried out by
the plate manufacturer or the fabricator when plates are cut. However,
when laminar defects are suspected, the entire base metal through which
ultrasound must travel to test the weld shall be tested for laminar reflectors
using a straight-beam search unit.
8.4.3 To ensure the coverage for the detection of transverse defects, transverse
notches shall be introduced as depicted on the Qualification block (See
appendix4). These transverse notches (both Cap and the Root) shall be
detected by the PA technique during the calibration / validation exercise.
Satisfactory detection of the transverse notches will ensure sufficient
coverage.
8.4.4 Alternatively, manual scanning with shear waves 70° or 60° may be
substituted for the detection of transverse defects. The 70° or 60° angle
beam shall be directed essentially parallel to the weld axis. The search unit
shall be manipulated so that the ultrasonic energy passes through the
required volume of weld and adjacent base material. The search unit shall
be rotated 180 degree and the examination shall be repeated. Wherever,
the weld cap is not machined or ground flat, the examination shall be
performed from the base metal on both sides of the weld cap in both weld
axis directions.
9.6 Scanning shall be performed using a Line Scan Technique. Each Line scan shall be
parallel to the weld using a Sector Scan, (S-Scan) and/or Electronic Scan, (E-Scan).
9.7 Appropriate refracted angles as accepted by ASME will define the positions of the
arrays, and hence the angles. These should be detailed in the Ultrasonic Scan Plan.
9.8 A minimum of two (2) line scans shall be performed at two (2) different index points
from the center of the weld from both sides of the weld where practical to ensure
coverage of the weld and heat affected zone, (HAZ).
9.9 Scans shall be parallel to the weld at Skew 90° and 270° scan axes (where accessible).
9.10 Raster scanning will aid in Flaw Characterization.
9.11 The rate of search unit movement shall be limited to a maximum of 50 mm per second
for Pipe to pipe circumferential welds unless calibration has been verified at a higher
speed.
9.12 Perform the examination from both sides of the weld, where practical, or from one
side as a minimum. All examination volume limitations shall be documented on the
Ultrasonic Examination Data Report Form.
9.13 The entire examination volume comprising the weld and adjacent base metal shall be
examined with the beam parallel to the weld on either side of the weld where
accessible, in two directions, i.e., clockwise and counter clockwise. Phased Array
Wedges contoured to the radius of the pipe may be used in the examination to ensure
proper contact.
9.14 Full A-Scan & S-Scan data shall be recorded and stored when using encoders.
10.6.1 Interpret the area containing the reflector in accordance with the applicable
examination instruction;
10.6.2 Plot and verify the reflector coordinates, provide a cross-sectional display
showing the reflector position and surface discontinuity such as root or
counter bore.
10.6.3 Review fabrication or weld prep drawings when accessible.
10.6.4 All recordable indications shall be resolved to determine the shape,
identity, and location of the reflector. Final evaluation and disposition of the
indication is the responsibility of owner/user of the component to be
examined.
10.6.5 Amplitude Determination - Signal amplitude shall be measured as a
percent of the calibrated DAC or TCG.
10.6.6 The appropriate Code or Specification shall be used to determine final
acceptance of the weld inspection using the Omni Scan.
The remaining couplant shall be wiped from the surface at the completion of the
examination.
13.0 DOCUMENTATION
13.1 Results of ultrasonic examination shall be reported on the PA UT Report Form shown
in the appendix. All Phased Array examination data for each recordable indication, A-
scans, B, C, and S Scans shall be saved as a Report File and recorded on a CD/DVD.
13.2 The examiner shall record the results of the examination on the PA Summary Report
from (Appendix 6). All recordable indications shall be documented on the
examination report. All blocks should be completed with the appropriate data or
"NA".
13.3 The Ultrasonic Scan Plan, Ultrasonic Calibration, and Ultrasonic Instrument Linearity
verification (if required) shall be considered part of the examination report.
SCREEN HEIGHT
AMPLITUDE CONTROL LINEARITY
LINEARITY
INSTRUMENT WITH INDICATION dB INDICATION ACTUAL % OF FULL SCREEN **
FINE dB CONTROL LIMITS % OF FULL
SET AT % OF CONTROL SCREEN
CHANGE HI GAIN LOW GAIN
FULL SCREEN SETTING
SETTING
FIRST SECOND INITIAL FINAL INITIAL FINAL
SIGNAL
SIGNA IN %
L IN % * ____dB ____dB ____dB ____dB
100 100
90 90
80 -6Db 32 TO 48
80 40 80 40
70 80 -12 Db 16 TO 24 70 80 -12 Db
60 60
50 40 +6 Db 64 TO 96 50 40 +6 Db
40 40
30 20 +12 dB 64 TO 96 30 20 +12 dB
20 20
*READING MUST BE 50% OF FIRST SIGNAL AMPLITUDE WITHIN 5% OF FULL SCREEN HEIGHT.
COMMENTS:
APPENDICES
APPENDIX-1 : PHASED ARRAY SCAN PLANS- FULL VOLUMETRIC COVERAGE PIPE TO PIPE WELDS
Note:
I) Scanning From OD
Wedge: SA12-N55S-IHC
Primary Height 1st Widt
Velocity Length Angle
Offset Element h
40m
2.33mm/µs 56.8mm 11.02mm 68.53mm 36o
m
Transducer: 5L64-A12
Total Aperture: 38.4mm Number of Elements: 64 Element Pitch: 0.6mm
75 mm
CL