You are on page 1of 21

ACCUSTAT™ STATIONARY

&
ACCUSTAT™ BMS

Differential Pressure Monitors

Installation and Operation Manual

Read and Save these Instructions

Manufactured by

749 Hope Road, Suite A- Eatontown, NJ 07724 USA


Tel: 800-224-9768, 732-389-8922 Fax: 732-389-8821
www.biologicalcontrols.com
12/20/2013
SAFETY INSTRUCTIONS
Read and Save these Instructions
Understanding the signal words: DANGER, WARNING and CAUTION.

These words are universally used for overall safety:

Signifies the most serious hazards, which will result in severe personal
injury or death.

Signifies hazards, which could result in personal injury or death.

Signifies unsafe practices, which would result in minor personal


injury or product and property damage.

READ INSTRUCTION MANUAL THOROUGHLY AND FOLLOW ALL


DANGER, WARNING OR CAUTION NOTES AND LABELS ATTACHED TO THE UNIT
BEFORE STARTING INSTALLATION OR MAINTENANCE ACTIVITIES.
IMPROPER INSTALLATION, ADJUSTMENT, ALTERATION, SERVICE,
MAINTENANCE, OR USE CAN CAUSE FIRE, ELECTRICAL SHOCK, OR OTHER
CONDITIONS THAT MAY CAUSE PERSONAL INJURY OR PROPERTY DAMAGE.
WEAR SAFETY GLASSES AND WORK GLOVES AND FOLLOW ALL SAFETY LOCAL
BUILDING AND ELECTRICAL CODES. CONSULT A QUALIFIED INSTALLER,
SERVICE AGENCY OR YOUR SUPPLIER FOR INFORMATION OR ASSISTANCE.
Table of Contents
Section:
Page#

1. INTRODUCTION.................................................................................................................................. 1
2. Remote Monitoring ................................................................................................................................ 1
3. Keyed Switches....................................................................................................................................... 1
3.1. Power ...................................................................................................................................................... 2
3.2. Positive/Negative Pressure Mode ........................................................................................................... 2
4. OPERATION and SET-UP.................................................................................................................... 2
4.1. Quick-Reference Instruction Guide ...................................................................................................... 2
4.1.1. Power ...................................................................................................................................................... 2
4.1.2. Display Activity ...................................................................................................................................... 2
4.1.3. Display (Units)........................................................................................................................................ 2
4.2. PRESSURE ALERT FUNCTION ......................................................................................................... 2
4.2.1. Alert Mode Switch ................................................................................................................................. 3
4.2.2. Alert Set-Point Adjustment.................................................................................................................... 3
4.2.3. Hospital Configuration Example: .......................................................................................................... 3
4.3. “ZERO” Display Adjustment ................................................................................................................ 3
4.4. Typical Pressure Alert Set-Up Examples: ............................................................................................. 4
4.4.1. Negative-Pressure................................................................................................................................... 4
4.4.2. Positive-Pressure .................................................................................................................................... 4
4.5. Audible Mute.......................................................................................................................................... 5
4.6. Alarm Contacts for Remote Options ..................................................................................................... 5
5. INSTALLATION ................................................................................................................................... 5
5.1. ACCUSTAT™/ BMS Mounting ............................................................................................................ 5
5.2. Remote Panel Mounting......................................................................................................................... 5
5.3. Sample Tube Wall Mount Kit................................................................................................................ 5
6. MAINTENANCE INFORMATION...................................................................................................... 6
6.1. General ................................................................................................................................................... 6
6.2. Cleaning.................................................................................................................................................. 6
7. WARRANTY ......................................................................................................................................... 6
7.1. Making a Warranty Claim .................................................................................................................... 6
8. SPECIFICATIONS ................................................................................................................................ 7
9. APPENDIX A ......................................................................................................................................... 8
9.1. Printed Wiring Boards........................................................................................................................... 8
9.1.1. EDI 091009 REV Model......................................................................................................................... 8
9.1.2. ACCUSTAT™ Earlier Model (REV J 2008).......................................................................................... 8
9.1.3. ACCUSTAT™ Earlier Model (REV 2002)............................................................................................. 9
9.1.4. Building Management System Option (BMS) ....................................................................................... 9
9.2. Enclosure Dimensions ...........................................................................................................................10
9.3. Schematics .............................................................................................................................................11
9.3.1. Building Management System Option (BMS) ......................................................................................11
9.4. Remote Panels .......................................................................................................................................12
10. CALIBRATION PROCEDURES.........................................................................................................14
10.1. W.C. (H2O) Calibration ........................................................................................................................14
10.2. Pascal (Pa) Calibration Procedure .......................................................................................................15
11. Troubleshooting.....................................................................................................................................16
12. Wall Mounting Template ......................................................................................................................17
ACCUSTAT™ STATIONARY and ACCUSTAT™ BMS
Differential Pressure Monitors

Sampling Medium:
Air or non-corrosive, non-explosive gas only

Do Not Exceed Overload Pressure: ±138” WC (H2O) or ±34K Pa Maximum


1. INTRODUCTION
The ACCUSTAT™ Stationary model is designed specifically to monitor the air-pressure conditions
within an enclosed room relative to the surrounding environment. Controlling and monitoring
pressurization is a vital step in the creation of a compliant, controlled, and safe environment within
medical facilities and other critical areas. Since both pressure ports are accessible the unit may be
mounted in one area and tubing can be run to the areas of interest to the ACCUSTAT™.

The ACCUSTAT™ (BMS) Unit is designed to connect to a Building Management System. The
BMS unit provides the additional computer interface circuitry required by a Central Computer and
Control System to monitor and react appropriately to pressure changes within a facility.
See the Specifications Section and Appendix for a listing of features, functions and schematics.

2. Remote Monitoring
Two Remote options are available to monitor the Alarm Status of an ACCUSTAT™ unit. A wall
mounted plate having a Red and Green Led is used to remotely monitor a single ACCUSTAT™.
A second larger wall mounted unit is used to monitor up to six (6) ACCUSTAT™ units from a
central location such as a nursing station. The larger unit is the same size and uses the same wall
mounting plate as an ACCUSTAT™. It is equipped with per-room Status LEDs and a mutable
buzzer. See the Appendix for schematics and installation information.

3. Keyed Switches
The key-switch is located on the top surface of the unit and may be removed in either select-position
to prevent unauthorized use. All ACCUSTAT™ Key-Switches use the same key within a model-
vintage.

Copyright 2014 Biological Controls


1 01/07/2014
3.1. Power
The Power on/off Key Switch is standard on stationary units (BMS units have no switch since power
is controlled by the central computer).

3.2. Positive/Negative Pressure Mode


This factory option switch is utilized in applications where the ACCUSTAT™ is anticipated to make
frequent changes in the two pressure monitoring modes.
The switch allows the user to change the pressure monitoring mode of the device without having to
open the unit to either move the internal selection jumper (JP1) or change the air sample tube on the
pressure ports.
4. OPERATION and SET-UP
4.1. Quick-Reference Instruction Guide
4.1.1. Power
The ACCUSTAT™ is powered with either twelve volts (12VDC) for the standard unit or twenty-four
volts (24VDC) for the Building Management System (BMS) option. A power-module is provided
with each ACCUSTAT™ to convert 90 to 250VAC @50/60 Hz wall-plug power to the 12 VDC.
The power-module is equipped with either a US-domestic AC wall-plug or an international plug
adaptor-kit. A 12VDC modular-connector is supplied on the power-module cable to connect to the
associated ACCUSTAT™ input plug. When 12VDC screw terminals are provided on the
ACCUSTAT™ the insulation is stripped back by approximately 5mm and the wires are pushed into
the connector and secured using a jewelers-screwdriver. Polarity (+/-) is not an issue with the
12VDC input power terminals since the design allows for connection without regard to
polarity. NOTE: Polarity must be observed for the 24VDC inputs to the BMS units.
The BMS unit is powered from the facility Building Management System circuits that must provide
the 24VDC source. A euro-type screw connector is provided on the BMS unit to interface both the
power and loop-monitoring cable-wires.
Note: If for any reason an ACCUSTAT™ BMS unit is removed from the central 24VDC power
source, a separate 24VDC power-module will be required to power the BMS unit to make it function
as a standard stand-alone ACCUSTAT™.
4.1.2. Display Activity
It is NORMAL for the display to fluctuate periodically. The air-pressure in most buildings can be
dynamic at times. Even slight pressure changes are detectable by the ACCUSTAT™. Fluctuations
are likely due to activity in the hallways caused by the opening and closing of stairway, room, or fire
doors. Elevator operation and air-conditioning systems can also cause pressure fluctuations.
The display will be updated constantly with the room pressure readings approximately once a second.
Positive pressures are displayed with no sign. Negative pressures are displayed with a negative sign (-
).
4.1.3. Display (Units)
The ACCUSTAT™ can display pressure-units in either Pascal’s (Pa) or Water Column (H2O, WC).
Place the Units-Selector jumper on the rear of the circuit board in the proper position for the intended
application. See the Appendix section to locate the jumper and the correct position.

4.2. PRESSURE ALERT FUNCTION


A visual or audible alert feature is available to indicate when pressure-levels are attained or
exceeded (without the need to look at the LCD display).

Copyright 2014 Biological Controls


2 01/07/2014
4.2.1. Alert Mode Switch
This three-position switch when in the left-most “Set” position causes the LCD Display to
switch from showing the differential pressure across the pressure-ports (“P1” and “P2”) to
displaying the desired Alert Set-Point. The center switch-position will mutes the Alert buzzer
and the right-most position allows the buzzer to sound on any red alert level after a one minute-
delay. The switch must be returned to the center or right-most position after adjusting the
Alert Set-Point to allow the LCD to display the differential pressure.
4.2.2. Alert Set-Point Adjustment
Slide the “Mode” switch to the left “Set” position. The LCD Display will now show the alarm
“Set-Point” pressure-level that establishes either a “Green” or “Red” LED alert depending upon
which pressure-port is used for the sampling. (See example below.) Use a small jeweler’s
screwdriver to change the pressure set-point potentiometer through the “Adj” access hole in the
front panel below the LCD display. After setting the Alert pressure-level, always return the
Mode Switch to either the “Mute” or “Aud” position so the LCD display can return to the
normal mode of showing the differential pressure between the ACCUSTAT™ location and the
area that is connected to the sampling tube.
4.2.3. Hospital Configuration Example:
A sampling-tube is attached to the ACCUSTAT™ “P2” pressure-port and the “P1” port is open. The
Pressure Alert set-point is adjusted to -.012” WC. The ACCUSTAT™ is installed in the hallway
outside the room to be monitored. The “P2”sampling tube is routed thru a ¼ inch hole thru the wall
into a patient room. A wall-plate kit is included with each ACCUSTAT™ to cleanly terminate the
sampling tube on the patient-room wall.
In this configuration, when pressures in the patient-room are more positive than the alert set-point the
Alert LED will start to flash red and conversely more negative pressures will cause the LED to glow
green. If the Alert LED remains flashing red for one minute or more an internal buzzer will sound
and relay contacts will close to signal remote-station circuits (if equipped) of the situation. If the
sampling tube port “P2” is used instead of “P1”, then pressures more negative than zero will cause
the Alert LED to flash red and positive pressures will cause the LED to turn green.
When the patient-room is used for an isolation function the air-pressure in the room should be
negative relative to the hallway such that contaminated patient-room air does not flow into the
hallways. For operating rooms just the opposite is desired to prevent contaminated air from entering
the surgery-area. Since “more-negative” or “more-positive” terms are relative to what pressure-port
(P1 or P2) is used for pressure sampling, always consult the facility engineer to determine what room-
pressure alerts need to be set and monitored for a particular application.
An optional POS-NEG key-switch is available to easily change the meter to function as either a
positive or negative pressure monitor without having to move the sampling tube ports. If the external
switch is not equipped a “jumper” on the inside of the unit can be used to change the pressure
monitoring mode. See the Appendix for correct jumper settings.

4.3. “ZERO” Display Adjustment


The ACCUSTAT™ display may need to be adjusted to “zero” after an installation. The unit is
calibrated at the factory to display .000”W.C. or 0.00 Pa for differential pressure monitoring when
both the P1 and P2 monitoring ports are at the same pressure. Each unit is also calibrated at
±.250”WC and ±6.25 Pa. See the Calibration section of the Appendix for factory calibration methods
and other information.

Copyright 2014 Biological Controls


3 01/07/2014
Allow the unit to be powered and reach room temperature before making adjustments.
Example: W.C. (H2O) ZERO adjustment.
• Power the ACCUSTAT™.
• Allow 5-15 minutes warm-up to stabilize the circuits.
• Set the Pressure Alert Mode Slide Switch to the “Mute” or “Aud” position (normal). The
“set” position is used only to display or set the desired Alert Pressure Level. Always
return this switch to either the Mute or Aud position for normal pressure monitoring.
The following “zero” display calibration cannot be performed if the Alert Mode
switch is in the “set” position.
• For the highest accuracy, loop the tubing from transducer Port “P1” to “P2” to prevent
outside interference during the Zero adjustment. If the unit is already mounted on a wall,
a long piece of tubing may be attached to the room sampling tube hole and extended
temporarily to be next to the ACCUSTAT™. If this is not possible, make sure that the
door is opened between the ACCUSTAT™ location and the monitored area during the
procedure.
• Turn the “Zero” adjustment potentiometer located near the top edge of the unit and to
the right of the display using a jewelers screwdriver. Adjust until the display reads -.000
and then very slowly (display updates once a second) continue until the minus (-) sign
just disappears (.000). Never force the potentiometer past the left or right stops. See
the Appendix Calibration section for instructions on adjusting an internal coarse-
adjustment pot when the fine-adjustment is out of range.

4.4. Typical Pressure Alert Set-Up Examples:

4.4.1. Negative-Pressure
Isolation Room Example: The ACCUSTAT™ is mounted on the wall in the hallway outside
an Isolation Room and a sampling tube from the ACCUSTAT™ (-) port is connected to the
pressure port labeled “P2”. The tube runs through the wall into the Isolation Room where it is
terminated on a stainless-steel wall plate mounted on the wall of the Isolation Room. (A wall-
plate kit is included with each ACCUSTAT™). The ACCUSTAT™ pressure (+) port labeled
“P1”is not used (this port senses the hallway pressure).
In this example, the Isolation Room is checked for a negative-pressure of at least -0.012WC
relative to the hallway pressure. To be alerted to the fact that the pressure is at or below this
level (more negative) without having to look at the LCD display set the Alert Select switch to
the right-most position. In this mode the Pressure Alert LED will start flashing red and after a
one-minute delay the audible alarm will sound when the room pressure goes positive relative
to the set-point. If the sampled pressure is -0.012WC or more negative; the LED will remain
green.
4.4.2. Positive-Pressure
Operating Room Example: In a hospital the operating rooms are checked from the hallway
with the same sampling tube set-up described above to detect positive-pressures at or above
the set-point of +0.012”WC.
The user adjusts the ACCUSTAT™ Alert Set-Point to (positive) +0.012WC.
Under these circumstances the Alert LED is Green when the sampled pressure is at or more-
positive than the set-point. When the pressure is more-negative than the set-point the Alert
LED will flash red (and ultimately set off the audible alarm after 1 minute).
Copyright 2014 Biological Controls
4 01/07/2014
4.5. Audible Mute
The Pressure Alert Mode Switch can be moved to the Mute position to turn-off the audible sound
without affecting the LED or remote monitoring functions.

4.6. Alarm Contacts for Remote Options


The remote relay contacts are available on connector P3. Reading from left-to-right, Pin 2 is NO
(normally open). Pin 3 is NC (normally closed) and Pin 4 is the C (common) pin. Jumper JP2 will
reverse the “normal” action of the contacts to provide a fail-safe mode that may be required for
certain applications. The other relay contacts on P3 and Jumper JP3 are reserved for custom
blower-control options.

5. INSTALLATION

5.1. ACCUSTAT™/ BMS Mounting


Mount the ACCUSTAT™ such that the pressure sampling-tubing leaving the transducer can line-up
(in straight-line) with the sampling tubing wall kit in the adjacent room. If the sample tube is run
though an anteroom, over a ceiling or another non-direct method; the tubing must be routed such that
there will be no kinks, sharp bends or debris forced inside the tubing. Use silicone calk around the
hole that enters the monitored-room to prevent air-leaks.
Use a level to scribe a 7” inch line on the wall where the unit will be mounted to represent the bottom
horizontal edge of the chassis. Remove the four (4) Philips-head screws that hold the two stainless-
steel chassis parts together. Separate the wall anchor-plate (rear-cover). Align the bottom-edge of
the rear cover chassis with the 7 inch line and trace-out the L-Shaped cut-out and the three (3) anchor
screw positions onto the wall. (A Wall Anchor-Plate Template found in the Appendix may be cut out
and taped to the wall for the same purpose.)
Use a ¼” drill-bit to drill the center-positions of the three mounting screws when utilizing the plastic
anchors and screws supplied with the unit. Otherwise, select the proper drill-bit for customer-selected
mounting hardware. The wiring and pressure tubing should be routed within the L-shaped cut-out
area. Provide enough slack in the wiring to enable the unit to be removed from the wall to access and
view the rear of the circuit board if needed.
A small slot is provided in the top cover of the unit for low voltage wiring that may be sent from a
power module mounted in a ceiling outlet and preferably routed through surface mounted conduit.

5.2. Remote Panel Mounting


Use the same mounting procedures as the ACCUSTAT™ for the larger unit. The smaller single
Accustat monitor plate attaches to the wall using the provided electrical box.

5.3. Sample Tube Wall Mount Kit


Drill a hole for the sampling tubing using the ¼” drill bit. Push the sampling tube onto the
ACCUSTAT™ pressure transducer port (usually P2) and then through the wall. To install the
sampling-tube wall mount kit in the room, pull the tubing through the grommet so the wall-plate is
flush with the wall. Mark the centers of the two mounting holes on the wall and drill a 7/32” hole for
each of the two plastic wall anchors found in the kit. Tap the plastic anchors into the holes with a
small hammer. Mount the wall-plate using the screws in the kit.

Copyright 2014 Biological Controls


5 01/07/2014
6. MAINTENANCE INFORMATION

Remove power from the unit during all maintenance routines.


6.1. General
The ACCUSTAT™ requires no maintenance other than cleaning surfaces that may become dirty
during normal use.
6.2. Cleaning
Keep water away from the unit and wall plug power module. If the unit comes into contact with
water remove power from the unit until all of the liquid has evaporated. Take care not to scratch the
bezel lens covering the LCD display. Treat the bezel like a camera-lens or eye glasses by avoiding
both harsh cleaners and tools that may scratch the surface.
Never use spray applicators to apply solvents. Use a soft damp (not wet) cloth to wipe the surfaces of
the unit as required. The cloth may be soaked in a mild dish detergent or denatured alcohol and
wrung-out thoroughly to assure that no running liquid will drip into the ACCUSTAT™.

7. WARRANTY
Limited Warranty:
Biological Controls warrants that the ACCUSTAT™ to be free of defects in workmanship and materials
during normal use and service for a period of Twelve (12) months from the date of purchase by the original
end user.
If at any time during the warranty period the product is defective or malfunctions, Biological Controls or its
dealer or distributor, from whom the product was purchased, shall at the option of Biological Controls replace
or repair the defective part or component.
This warranty shall not apply if it is shown that the defect or malfunction was caused by damage due to
shipment, improper electrical connections, or improper use or abuse of the product.
The sole responsibility of Biological Controls shall be to repair or replace the product within the terms stated
above. Biological Controls shall not be liable for any loss or damage of any kind, including any incidental or
consequential damages resulting, directly or indirectly, from any breach of warranty, expressed or implied, or
any other failure of this product. (Some areas do not allow the exclusion or limitation of incidental or
consequential damages, so this limitation may not apply to you.) The warranties set forth are exclusive and
Biological Controls expressly disclaims all other warranties, whether written or oral, implied or statutory,
including but not limited to any warranties of merchantability, workmanship, or fitness for a particular use.
In our continuing effort to produce the highest quality products, we reserve the right to change or alter
product specifications and materials without notice.
This warranty gives you specific legal rights and you may have other rights, which vary, from state to state or
country to country.
7.1. Making a Warranty Claim
To make a warranty claim or if you have questions about the warranty policy, contact the distributor from
whom you purchased the product.
NOTE: Do not return any products or components directly to the factory without a factory issued “Return
Merchandise Authorization (RMA) number” issued by the Biological Controls Customer Service Department.
Any products returned without the issuance of the RMA number will be refused and returned to shipper.
For questions related to this warranty call or write:
Manufacturer:
BIOLOGICAL CONTROLS TEL: 800-224-9768
749 Hope Road Suite A FAX: 732-389-8821
Eatontown, NJ 07724 WEBSITE: www.biologicalcontrols.com

Copyright 2014 Biological Controls


6 01/07/2014
8. SPECIFICATIONS

ACCUSTAT™ Specifications

ELECTRICAL OPERATIONAL FEATURES

• Voltage typical: 12V Min. / 15V Max. • LCD Display Refresh Rate: 1.0 Second
(International 12VDC Power Module) • Visual Alert LED: Green / Red Alert Modes
Typical Current: 50 mA (70mA max ) • Alert Set Point Storage: Non-Volatile Memory
• Alert Set Point: User Defined; Front Panel
Maximum Power: <1.3W @12VDC
Adjusted
• BMS option: Voltage 24VDC±10% • Set Point Alerts (Flashing/Steady Modes)
Current: 120mA max • Audible Alarm: One Minute Delayed Buzzer
(Power supplied by Bld. Mngmt. System) • Zero Calibration: Front Panel Adjusted
Power: Typ.<2W • Accuracy of Alarm Output: ± 1% of Set Point
• On/Off Key Switch (not on BMS) • Standard Application Range: ± 0.1” H2O (± 25 Pa)
• Resolution: 0.001" H2O (0.1 Pascal’s)
TRANSDUCER • Tamperproof Controls: On/Off Key Switch
• Piezoresistive Silicon • Pos/Neg Key Switch (optional)
Calibrated: -20 °C to 85 °C [-4 °F to 185 °F] • Wall Mounting Hardware Kit
• Medium: Air or Non-Corrosive, Non-Explosive • Selectable Pos/Neg Pressure Modes
Gas • Both Pressure Ports Accessible for Probe tubing
• Application Pressure Range: -0.100 to +0.100”
H2O and -6.25Pa to +6.25Pa Remote Panel or Remote Monitor
• Overload Pressure: ±138” WC or ±34K Pa Outputs:
Maximum
• Accuracy of Reading: ± 1% F.S. • SPDT Relay Contacts - 1A max (Resistive) @ 30
VDC or 120 VAC
MECHANICAL DATA
Building Maintenance System Option
• Wall Mount Enclosure: Stainless Steel #4 Polish (BMS):
• LCD Display: ½ inch High Single Line
• Dimensions: 7.367" (W) x 3.25" (H) x 2.25" (D) Selectable Modes:
• Weight: 22 oz. (BMS option 24 oz.) • 4-20ma Current Loop Output
• Operating Temperature: 55 - 105°F (13°C to 41°C) • 0-5VDC, 2.5V at Zero Pressure 2mA max
• Sampling Tube: Requires a 1/4 Inch Hole into the • 0-10VDC, 5V at Zero Pressure 2mA max
Sampling Area. A Stainless Probe Plate kit w/ 1’ of
¼” tubing included. • Max Loop Resistance is 580 OHMS
• Operating Humidity: 10% to 94% RH Non-
condensing

Copyright 2014 Biological Controls


7 01/07/2014
9. APPENDIX A
9.1. Printed Wiring Boards
9.1.1. EDI 091009 REV Model

9.1.2. ACCUSTAT™ Earlier Model (REV J 2008)

Copyright 2014 Biological Controls


8 01/07/2014
9.1.3. ACCUSTAT™ Earlier Model (REV 2002)

9.1.4. Building Management System Option (BMS)

Copyright 2014 Biological Controls


9 01/07/2014
9.2. Enclosure Dimensions
ACCUSTAT™ and Remote Panel Front Covers. The Wall Mounting Plate is the same for both.
Dimension units are in inches.

Copyright 2014 Biological Controls


10 01/07/2014
9.3. Schematics

9.3.1. Building Management System Option (BMS)

Internal BMS Circuit Board

Copyright 2014 Biological Controls


11 01/07/2014
9.4. Remote Panels

Copyright 2014 Biological Controls


12 01/07/2014
Copyright 2014 Biological Controls
13 01/07/2014
10. CALIBRATION PROCEDURES

Follow standard electrostatic discharge (ESD) practices and procedures (i.e.; use a properly grounded
wrist-strap) during calibration or maintenance activities to avoid static-electricity discharge currents
that can damage the unit.

Note: The Pressure Alert Mode-Switch must be in the “Mute” or “Audible” position to
calibrate the differential pressure zero setting.
If the unit cannot be adjusted per the instructions, contact the factory technical support for advice
before possibly having to return the unit for calibration. Always obtain an RMA number from the
factory before returning the unit.

When possible calibrate the ACCUSTAT™ after installation so variables due to the mounting position
and environmental conditions within the application-area are eliminated.

When calibration is performed on a bench, place the unit flat on the table (in a horizontal position) so that the
LCD display is positioned in the upper left hand corner, the on/off key-switch is on top of the unit, the black
three (3) position slide switch is under the display in the center (alarm-mute) position.

10.1. W.C. (H2O) Calibration


1. Connect the DC power module supplied with the unit or a 12-14VDC power source to the two (2) pin power
connector shown on the schematic. The connector is located on the back of the circuit board in the lower left
hand corner. See the Appendix for schematic diagram. BMS units require a 24VDC power source.

2. Place jumper JP6 on the rear of the circuit board in the correct position for the unit to display WC (H2O) or
Pa pressure-units. (See Appendix diagrams for location information).

3. Place jumper JP1 on the rear of the circuit board in the correct position for the unit to be in the negative (-)
pressure-mode. If a pressure select key-switch is equipped there is no need to locate JP1. Set the switch to
the negative (-) pressure-mode position.

4. Using the power key-switch, turn the ACCUSTAT™ “on” for at least 15 minutes to allow for temperature
stabilization. Check the voltage at the input connector pins with a DC voltmeter (under load) to verify the
correct voltage. The internal voltage-regulator will easily compensate for ±10-20% input voltage tolerances.
Note: Since the power source for BMS units is controlled by the facility maintenance system, an “on”-
“off” switch does not normally exist on BMS units. Use an external 24VDC, 200ma power source for
calibration.

5. Adjust the unit to display .000”W.C. or 0.00 Pa for differential pressure monitoring when both P1 and P2
ports are at the same pressure. Each unit is also calibrated at the higher levels of ±.250”WC and ±6.25 Pa
by applying these pressure-levels to transducer-port “P2” while “P1” remains open.
• Open P1 and P2 ports or tie them together with a short piece of tubing to assure that both are at the
same pressure and that local air-currents do not affect the procedure. Check the display for the
zero reading.

• Turn the zero adjustment screw with a flat-head jeweler’s screwdriver to obtain a display reading
of .000. The screw is accessed through a small hole above and to the right of the LCD display
Copyright 2014 Biological Controls
14 01/07/2014
located along the top edge of the unit. If the display is reading a negative (-) value rotate the screw
clockwise. If a positive (+) value is displayed then rotate counter-clockwise. Do not force the
screw in either direction.

If the “Zero” (Fine- Adjustment) is out of range a Coarse-Adjustment (pwb marking C/Z) on the
inside of the unit can be altered. Adjust the “Zero” (fine adjustment) on the front panel such that it
is centered (screw-slot running up and down in the vertical direction).
To perform this procedure the ACCUSTAT™ cover is removed by unscrewing the four (4) screws
attaching the front cover to the wall plate. Then two screws that mount the printed wiring board
must be removed to gain access to the rear of the ACCUSTAT™ circuit board (pwb).
See the proper circuit diagram in the Appendix and locate the Coarse-Adjustment potentiometer
accessible after the cover and circuit board are removed. A hole in the BMS pwb (marked C/Z) is
used to access the potentiometer screw.
Carefully turn the adjustment screw left or right to get the meter to read close to .000. The
adjustment does not have to be exact since the final zero setting will be done with the fine
adjustment on the front after the cover is replaced.

6. Apply -.250” W.C. pressure from a calibrated source to the P2 port (upper port) located on the black
pressure-transducer. If -.250” W.C. does not display on the LCD then adjust VR1 (trimmer pot) located
near the white AXICOM relays. Set the VR1 so that -.250” W.C. is displayed on LCD. When the BMS pwb
is present and covers the pot, adjustment is made through the pwb hole marked “H2O” just above the pot.

7. Remove pressure from ACCUSTAT™ P2 transducer port and the display should read .000.

8. Repeat the last 2 steps of the procedure to verify repeatability.

10.2. Pascal (Pa) Calibration Procedure

1. Follow steps 1 thru 5 from the WC Calibration procedure above. Make sure the JP6 is in the Pa position.

2. Apply -62.5 Pa. from a calibrated pressure source to the P2 port (upper port) located on the black pressure
transducer. If -62.5 Pa. does not display on the LCD then adjust VR2 (trimmer pot) located near the white
AXICOM relays. Set VR2 so that -62.5 Pa is displayed on LCD. When the BMS pwb is present the
adjustment is made through the pwb hole marked “Pa” just above the pot.

3. Remove pressure from ACCUSTAT™ and 0.00 should now be displayed.

4. Repeat the last 2 steps of the procedure to verify repeatability.

5. Calibration procedure is complete.

Copyright 2014 Biological Controls


15 01/07/2014
11. Troubleshooting

Problem Solutions
No Display Check Power Voltage and Connections
Display remains neutral Is room completely sealed, is exhaust greater
than supply (check with velometer)?
The 3-position slide switch must be in the
center or to the right.
Display does not move during Alert set point The 3-position slide switch must be set to the
adjustments rightmost position. (return to center or left
after adj.
Alert LED remains Green for alarm condition Check Alert set-point. Check slide switch for
proper position. Is sampling tube on proper
transducer Port?
Erratic Display, will not stabilize when setting Sampling tube must be disconnected or
“Zero” extended to be next to the ACCUSTAT™.
Check power connections and voltage
stability.
Erratic Display when sampling tube is Check for proper connection of the sampling
connected tube to the ACCUSTAT™ and the wall mount
fixture. Inspect tubing for bends, cracks,
pinching or obstructed. Check power
connections and voltage stability.
Audio Alarm (buzzer) does not sound when Check the 3-position slide switch to make sure
set-point is reached and the LED is flashing it is not in the “Mute” position. Allow 1
red. minute delay between an alarm condition and
the sound. Every time the alarm condition
disappears the minute delay is reset and starts
counting again.

Copyright 2014 Biological Controls


16 01/07/2014
12. Wall Mounting Template

The same template is used for both the ACCUSTAT™ and the REMOTE PANEL

Copyright 2014 Biological Controls


17 01/07/2014

You might also like