EMPOWER

INJECTOR
SYSTEM
SITE SURVEY &
SITE INSTALLATION
INSTRUCTIONS

E-Z-EM, INC., WESTBURY, NY 11590-5021


Phone: 516-333-8230 Fax: 516-333-8278
EMPOWER INJECTOR SYSTEM

PROPRIETARY INFORMATION NOTICE

This document contains information proprietary to E-Z-EM, Inc. All


rights are reserved.

No part of this document may be reproduced, transmitted, processed or


recorded by any means or form, electronic, mechanical, photographic or
otherwise, nor be released to any third party without the express written
consent of E-Z-EM, INC.

Site Survey Doc. 018-0913, Item 1476913, Rev A ii


Empower Injector System

Table of Contents

SECTION 1: INTRODUCTION .............................................................................. 1-1


1.1 PURPOSE OF THIS MANUAL ............................................................................ 1-1
1.2 MANUAL ORGANIZATION ................................................................................ 1-1
1.3 MANUAL CONVENTIONS ................................................................................. 1-2

SECTION 2: SITE SURVEY.................................................................................. 2-1


2.1 PURPOSE .................................................................................................... 2-1
2.2 GENERAL SITE CONSIDERATIONS .................................................................... 2-1
2.3 FLOOR STAND .............................................................................................. 2-2
2.4 OVERHEAD (CEILING/WALL) MOUNT ................................................................ 2-3
2.5 SITE SURVEY CHECK LIST .............................................................................. 2-5

SECTION 3: INSTALLATION INSTRUCTIONS .................................................... 3-1


3.1 INSTALLATION OVERVIEW .............................................................................. 3-1
3.2 CEILING MOUNT COLUMN INSTALLATION .......................................................... 3-1
3.3 WALL MOUNT BRACKET INSTALLATION ............................................................. 3-2
3.4 REMOTE CONTROL WALL MOUNT BRACKET ...................................................... 3-3
3.5 POWER SUPPLY............................................................................................ 3-4
3.6 CABLE LENGTH CONFIGURATIONS ................................................................... 3-5

SECTION 4: DRAWINGS...................................................................................... 4-1


4.1 ARTICULATING ARM OPERATING RANGE - WALL MOUNT ..................................... 4-1
4.2 ARTICULATING ARM OPERATING RANGE - CEILING MOUNT .................................. 4-2
4.3 WALL MOUNT HOLE LOCATIONS – POWER SUPPLY BRACKET .............................. 4-3
4.4 WALL MOUNT HOLE LOCATIONS- REMOTE CONTROL BRACKET ............................ 4-4
4.5 CEILING COLUMN MOUNTING HOLE LOCATIONS ................................................. 4-5

Site Survey Doc. 018-0913, Item 1476913, Rev A iii


Section 1: Introduction Empower Injector System

Section 1:
Introduction
1.1 PURPOSE OF THIS MANUAL

This document details information regarding the Empower family of Injector


Systems. The Empower will be shown in photos. This document provides
site survey instructions and the site survey check list for the Empower Injector
System. In addition, the document provides installation instructions for
mounting of the Ceiling Column, Wall Mount for the Articulating System,
Remote Wall Mount Bracket and Wall Mounted Power Supply Assembly.
Through use of this document, the site will be ensured to receive the proper
configuration for their installation.

1.2 MANUAL ORGANIZATION

This document is organized into the following sections:

Section 1: Introduction Identifies the purpose and organization of


this manual.

Section 2: Site Survey Contains site survey instructions, the Site


Survey Check List

Section 3: Site Installation Provides detailed installation instructions


Instructions for Injector wall and ceiling installations
and Power Supply and Remote Control
Wall installations.

Section 4: Drawings Provides illustrations showing the


Articulating Arm operating range for wall
mount and ceiling mount configurations.
Provides dimensional drawings to mount
the Ceiling Column, wall mounting the
Power Supply and wall mounting the
Remote Control.

Site Survey Doc. 018-0913, Item 1476913, Rev A 1-1


Section 1: Introduction EMPOWER INJECTOR SYSTEM

1.3 MANUAL CONVENTIONS

This manual uses the following conventions:

♦ Bold indicates emphasis or heading.


♦ Note is used to set off important information from the rest of the text.

This manual also uses the following warning format:

The WARNING symbol alerts you to a hazard


that may result in equipment damage or
WARNING personal injury.

Site Survey Doc. 018-0913, Item 1476913, Rev A 1-2


Section 2: Site Survey Empower Injector System

Section 2: Site
Survey
2.1 PURPOSE

The Site Survey is to be used to determine site requirements for the most
optimal installation of your Empower Injector System.

Important!

Please complete and submit the Site Survey Check List prior to
requesting shipment of the Empower Injector System.

2.2 GENERAL SITE CONSIDERATIONS

The following considerations must be reviewed prior to installation:

Power Requirements – The Empower Injector System requires a standard,


hospital grade power source with a grounded outlet. Line cords compatible
with local voltages are provided with the system. A power source with a
dedicated circuit is recommended for all E-Z-EM Injector systems. Two
dedicated outlets are required for use with the Empower Injector System.
The first outlet is for the power supply assembly and the second outlet is for
the Remote Control. Before installation, take careful note to where the
outlets are located. These locations may affect the length of cables required
of the installation. Power outlets should be either 120 VAC/60 Hz or 240
VAC/50 Hz.

Access – Patient access to the gantry (left and right sides) must not be
impeded by the Injector system.

WARNING
Cables on the floor are a potential hazard and should not be in the
patient or hospital staff access path.

Site Survey Doc. 018-0913, Item 1476913, Rev A 2-1


Section 2: Site Survey Empower Injector System

There are two types of Injector mounting possibilities: Floor Stand and
Overhead (Ceiling/Wall) Mount.

2.3 FLOOR STAND

The floor stand assembly is comprised of a pole, base with casters,


counterweight, and shroud. The Injector and Extravasation Detection
Accessory (EDA), are mounted on the pole. Casters on the base facilitate
movement and placement of the Injector system throughout the scanner
room.

Typical Floor Installation

Site Survey Doc. 018-0913, Item 1476913, Rev A 2-2


Section 2: Site Survey Empower Injector System

2.4 OVERHEAD (CEILING/WALL) MOUNT

The overhead mount assembly is comprised of a ceiling column or wall


mount bracket, articulating arm, and Injector mounting arm. The Injector and
the Extravasation Detection Accessory (EDA) are mounted on the Injector
mounting arm.

Prior to installation of the overhead mount assembly, a suitable structure (for


example: uni-strut) must be located that will support the load of the Empower
Injector System and mount. Facility Engineering or other knowledgeable
personnel should be consulted regarding, but not limited the installation of the
wall mount and ceiling mount systems, power supply assembly wall mounted
and Remote Control wall mounted.

Important!
The customer is responsible for the installation of the Injector System
Ceiling Mount Column or Wall Mount bracket.

It is recommended that two (2) people perform the installation on the Injector
System Ceiling Mount Column.

The Customer must provide fasteners and anchors for mounting the Ceiling
Mount Column and Wall Mount Bracket:

• Wall mount systems (2) each.


• Ceiling mount systems (4) each.

Lengths – Three different column lengths are available S (11.25"), M (21.50"),


and L (34.50").

Clearance – Once installed, the horizontal member of the Articulating Arm


must be able to clear the top of the scanner.

Minimum Height – The minimum height of the bottom of the Ceiling Mount
Column should be 84”(7 feet) from the floor. The bottom of the Injector
should be a minimum of 36” (3 feet) off the floor. The minimum height of the
bottom of the Wall Mount should be 96” (8 feet) off the floor.

Site Survey Doc. 018-0913, Item 1476913, Rev A 2-3


Section 2: Site Survey Empower Injector System

Typical Ceiling Installation

Site Survey Doc. 018-0913, Item 1476913, Rev A 2-4


Section 2: Site Survey Empower Injector System

2.5 SITE SURVEY CHECK LIST

Prior to shipment of an Empower Injector System including accessories, please complete


the form below and return to the following address:

E-Z-EM, Inc.
Attn: Customer Service
717 Main Street
Westbury, NY 11590
Phone: 1-800-544-4624 ext.2393 Fax: 1-800-443-6362

1. Site Name: __________________________

2. Injector is being placed in:

New CT Suite Remodeled CT Suite Existing CT Suite

3. Scanner manufacturer:

GE Philips /Marconi Siemens Toshiba

Other ____________________________ Scanner Model: _________________

4. Is the system displacing an existing Injector system? Yes No

If Yes, which manufacturer:

Medrad Liebel E-Z-EM Other_________________

Model: ___________________________

5. Mount style requested: Floor Stand Ceiling Mount Wall Mount

For Ceiling Mount System Installations Only:


Ceiling Mount Length Required:

Short (11.25") Medium (21.50") Long (34.50")

Height from Mount Support Structure (usually uni-strut) to Ceiling: _________ Inches

Height from Ceiling to Floor: _________Inches

6. Cable Lengths:

Remote to Power Supply Cable: Standard 20 ft. / Optional 50 ft

Injector to Power Supply Cable: Standard 50 ft. / Optional 20 ft

7. Remote Mount Required: Pedestal Mount Wall Mount

8. Approximate installation date: _______________________

Special Instructions/Requirements:
____________________________________________________________________________________
____________________________________________________________________________________

Submitted by:_________________________________________Date:____________________

Site Survey Doc. 018-0913, Item 1476913, Rev A 2-5


Section 3: Installation Empower Injector System

Section 3: Installation
Instructions
3.1 INSTALLATION OVERVIEW

After the site survey form has been completed and sent to E-Z-EM, Customer
Service will review the form and the order will be entered in accordance with
the information provided on the form. The appropriate system and
accessories will be sent to the site.

It is the responsibility of the site to install the Ceiling Column or Wall Mount
Bracket for the Articulating Arm, Remote Wall Mount Bracket and to mount
the power supply assembly to the wall if required.

The use of non-E-Z-EM Ceiling Column or Wall mounts is not


recommended.

3.2 CEILING MOUNT COLUMN INSTALLATION

Install the Ceiling Mount Column as follows:

1. Remove the Ceiling Mount Column from its packaging.

Note: Position of the Pivot Shaft (see Fig. 3-1b) is critical to the range of
motion (approximately 270 degrees) of the articulating arm. The pivot shaft
can be orientated in any position prior to securing the ceiling column to its
support structure. Select the pivot shaft orientation such that the resulting
range of motion best addresses the operational requirements of the site (See
Fig. 4-2).

2. Mark the locations of the holes by using the Ceiling Mount Column
drawing (see Figure 4-5). Drill mounting holes at the locations marked.

3. Install the Ceiling Mount Column using the appropriate fasteners and
anchors (NOT PROVIDED) (see Figure 3-1a). Ensure that the base of the
column is level.

Site Survey Doc. 018-0913, Item 1476913, Rev A 3-1


Section 3: Installation Empower Injector System

Pivot Shaft

Figure 3-1a: Ceiling Mount Column


(Pattern for Competitor column not shown)

(A)
Figure 3-1b: Pivot Shaft
(B)
(2) 9/16 in. dia. holes

7 ½ in.

(C)

Figure 3-2a: Wall-Mount Bracket Figure 3-2b: Wall-Mount Bracket

3.3 WALL MOUNT BRACKET INSTALLATION

Install the Wall Mount Bracket as follows:

1. Remove the Wall Mount Bracket from its packaging.

2. Unsnap and remove plastic cover (A), plastic housing (B), and plastic
plug (C) (see Figure 3-2a).

3. Place the Wall Mount Bracket against the surface to which it will be
mounted. Mark the location of the holes by using the bracket as a
template (see Figure 3-2b).

4. Drill mounting holes at the appropriate points.

Site Survey Doc. 018-0913, Item 1476913, Rev A 3-2


Section 3: Installation Empower Injector System

5. Install the Wall Mount Bracket using the appropriate fasteners and
anchors (not provided). Replace the plastic covers and plug removed in
Step 2, above. Make certain that the bracket is level.

3.4 REMOTE CONTROL WALL MOUNT BRACKET

The Remote Control is typically table mounted in the control room in close
proximity to the scanner console. The Remote Control can also be mounted
on the wall in the control room.

Location – The Remote Control will usually be connected to the Power


Supply by a 20-ft. cable. An optional 50 ft cable is available. Take into
consideration the cable routing when locating the Remote Control and the
Power Supply.

Location of power outlet – The preferred location of the power source is a


wall outlet that must be within reach of the Remote Control’s line cord
(2 meters; approx. 6-ft.).

A Wall Mount is provided for wall mounting installations. The Wall Mount
consists of the following three parts that require assembly:

• wall mount bracket


• main body
• top cover

To wall mount the Remote Control, install the wall mount bracket as follows:

1. Remove the Wall Mount components from packaging.

2. Select a suitable wall location and do the following:

Note: Make careful measurements before making permanent


installations. The Remote Control’s AC line cord must reach a wall outlet.

a. Using the Remote Wall Mount Drawing (see Figure 3-4), mark four
hole locations on the wall.

Note: If mounting to a masonry or cement wall, use anchors provided. If


mounting into any other material, the proper anchors (not supplied) must
be utilized and capable of supporting a minimum of 30 lbs.

b. Drill four holes into wall. Hole size is determined by anchor type.
Consult anchor manufacturer’s literature for recommended size.

c. Before driving in the anchors (if using anchors provided), push the
rounded end up into the sleeve as much as possible. Use a hammer
to drive in the anchors up to the top of the sleeve. Warning: driving
the exposed threads on the bolt may damage them.

d. Remove the nuts and washers. Place the Wall Mount Plate against
the wall then replace nuts and washers. Tighten nuts.

Site Survey Doc. 018-0913, Item 1476913, Rev A 3-3


Section 3: Installation Empower Injector System

3.5 POWER SUPPLY

The Power Supply may be placed on the floor or may be wall mounted.

Location – The Power Supply will be usually connected to the Injector by a


50-ft cable and to the Remote Control by a 20-ft. cable. There is an option for
a 20 ft Injector to power supply cable. Take into consideration the cable
routing when locating the Power Supply, Injector, and the Remote Control.

Location of power outlet – The preferred location of the power source is a


wall outlet that must be within reach of the Power Supply’s AC line cord
(3 meters; approx. 9-ft.).

To wall mount the Power Supply, install as follows:

1. Remove the Power Supply and AC line Cord from the Injector package.

2. Select a suitable wall location.

Note: Make careful measurements before making permanent


installations. A 50-ft. cable is used to connect the Power Supply to the
Injector, and a 20-ft. cable is used to connect the Power Supply to the
Remote Control. The Power Supply AC line cord must also reach the
power outlet.

a. Using the Power Supply drawing (see Figure 3-3) mark 4 hole
locations on the wall at the center of each “+” cutout.

b. Drill mounting holes in the wall and insert the proper mounting
hardware such that the space between the wall and the bottom of the
head of the screw is 1/4”.

Note: Fasteners (not provided) must be rated for 30 lb. load, minimum.

c. Using a pair of pliers, remove the rubber feet on the bottom of the
Power Supply.

d. To mount the Power Supply, align the mounting screws with the
mounting holes and slide the Power Supply down. Insure that the
Power Supply is closely fit against the wall. If it is not, remove the
Power Supply and slightly decrease the distance between the head of
the mounting screw and the wall. Replace Power Supply.

Site Survey Doc. 018-0913, Item 1476913, Rev A 3-4


Section 3: Installation Empower Injector System

3.6 CABLE LENGTH CONFIGURATIONS

Two cables are provided that are sufficient for most installations:

! Remote to Power Supply Cable – 20 ft. cable for


connecting the Remote Control to the Power Supply.
An optional 50 ft cable is available.
! Injector to Power Supply Cable – 50 ft. cable for
connecting the Injector to the Power Supply. An
optional 20 ft cable is available.

Important: You cannot connect two Remote to Power Supply Cables or two
Injector to Power Supply Cables together for the purposes of increased
length. Please make sure that the cables provided will be sufficient for the
installation.

Site Survey Doc. 018-0913, Item 1476913, Rev A 3-5


Section 4: Drawings Empower Injector System

Section 4: Drawings
4.1 ARTICULATING ARM OPERATING RANGE - WALL MOUNT

WEIGHT SPECIFICATIONS:

EmpowerCT Injector with: EmpowerCTA Injector with:


Mounting Arm, 200ml syringe, Mounting Arm, 200ml syringe,
EDA, heater and cables: 20 lbs EDA, heater and cables: 28 lbs
Wall Mount: 5 lbs Wall Mount: 5 lbs
Articulating Arm: 22 lbs Articulating Arm: 22 lbs
Injector Mounting Arm: 5 lbs Injector Mounting Arm: 5 lbs
Total System Weight: 52 lbs Total System Weight: 60 lbs

Site Survey Doc. 018-0913, Item 1476913, Rev A 4-1


Section 4: Drawings EMPOWER INJECTOR SYSTEM

4.2 ARTICULATING ARM OPERATING RANGE - CEILING MOUNT

WEIGHT SPECIFICATIONS:

EmpowerCT Injector with: EmpowerCTA Injector with:


Mounting Arm, 200ml syringe, Mounting Arm, 200ml syringe,
EDA, heater and cables: 20 lbs EDA, heater and cables: 28 lbs
Wall Mount: 5 lbs Wall Mount: 5 lbs
Articulating Arm: 22 lbs Articulating Arm: 22 lbs
Injector Mounting Arm: 5 lbs Injector Mounting Arm: 5 lbs
Total System Weight: 52 lbs Total System Weight: 60 lbs

Site Survey Doc. 018-0913, Item 1476913, Rev A 4-2


Section 4: Drawings EMPOWER INJECTOR SYSTEM

4.3 WALL MOUNT HOLE LOCATIONS – POWER SUPPLY BRACKET

(DRAWING NOT TO SCALE)

(ALL DIMENSIONS ARE INCHES)

Site Survey Doc. 018-0913, Item 1476913, Rev A 4-3


Section 4: Drawings EMPOWER INJECTOR SYSTEM

4.4 WALL MOUNT HOLE LOCATIONS- REMOTE CONTROL BRACKET

(DRAWING NOT TO SCALE)

(ALL DIMENSIONS ARE INCHES)

Site Survey Doc. 018-0913, Item 1476913, Rev A 4-4


Section 4: Drawings EMPOWER INJECTOR SYSTEM

4.5 CEILING COLUMN MOUNTING HOLE LOCATIONS

(DRAWING NOT TO SCALE)

(ALL DIMENSIONS ARE INCHES)

Site Survey Doc. 018-0913, Item 1476913, Rev A 4-5