You are on page 1of 2

The values used for the data treatment under the wavelength of maximum

are shown in Table 1. The values were used absorption of 541nm.


to create a calibration curve that will find the
Table 2. Absorbance of the unknown
equation of the best fit line. Using this line,
concentration
the concentration of the unknown solution
can be determined by measuring its Concentration
absorbance. of Stock
Trial Absorbance
Sample Cu
Table 1. Calibration Curve Values
(II) ppm
Volume of
Concentration 1 0.111 267
Working
of Standard Absorbance 2 0.110 264.5
Standard
Cu(II) ppm 3 0.110 264.5
Solution
Average 0.110333333 265.6666667
0.00 0 0.002
2.00 100 0.042
All of the three trials for the solution
4.00 200 0.087 containing the unknown concentration had
6.00 300 0.118 an absorbance of 0.036. Replacing y with
8.00 400 0.157 0.036 in the equation y=0.0001x, the
10.00 500 0.196 concentration of the unknown sample would
Using the equation y = mx + b, be 360ppm. Since all of the trials had the
absorbance could be determined by same absorbance values, it can be inferred
equating it to y. This is determined by the that the data acquired is precise.
spectrophotometer. Meanwhile, x is Possibles sources of errors for this
equivalent to the concentration of the experiment would be random, systematic
solution, m is equal to the absorptivity, and b and gross. Random errors are often
would be the y-intercept. Using Microsoft unavoidable. They result from the usage of
Excel, the calibration curve is shown as seen glassware and may affect the precision of the
in (Figure 1) (refer to Appendix 2). In the results. Systematic errors could occur from
equation, y = 0.0001x thus, the absorbance method or instrument. The absorbance
is equal to 0.0001. Moreover, since there is values obtained from the spectrophotometer
no y-intercept in the data acquired, it signifies may increase if the reaction cell was held
that there is no deviation from the theoretical improperly. This is because of the
value. In the y-intercept, there is no fingerprints may be left in the cell. These
concentration of Cu(II) because x = zero. fingerprints may absorb light from the
Theoretically, the absorbance at this point spectrophotometer which increases
would be zero because there is no colored absorbance value. Since the same cell was
solution due to the lack of copper which used for all tests in this experiment, it is
means that no light should be absorbed due important that it is washed with the solution
to the solution being colorless. The presence about to be tested throughly. This is done to
of a y-intercept even after negating the ensure that the correct concentration of the
effecting of the cell and the ammonia solution Cu(II) is being analyzed otherwise the
on the values of absorbance signals that absorbance value may increase or decrease
something else is absorbing light due to the depending on the contaminated
presence of an absorbance value even at concentration. The reagent blank test must
zero ppm concentration. In this case, a y- be performed in order to negate the effects
intercept would deviate from the theoretical of the cell and the ammonia solution on
value by increasing the values of absorbance otherwise the absorbance
absorbance. The data acquired were done values will increase. Instrumental errors may
occur in this experiment from the
spectrophotometer. Any stray light in the concentration of the sample. The curve must
spectrophotometer will decrease the be constructed as linear as possible. When
absorbance value. This is due to the testing standardized solutions for the
additional presence of unabsorbed light. The construction of the calibration curve, some
cell used for analysis must be standardized tests may be repeated to find the most
because a wider cell results in a longer path accurate result to reduce any deviations in
length. The longer path length will increase the linear graph. Any instrument error may be
absorbance values, and the graph may addressed through proper inspection of the
deviate from its linear form. The spectrophotometer before the performance
spectrophotometer must also be inspected of the experiment. In addition, it is important
before use to ensure it is still effective for that all solutions used for the construction of
analysis. Gross errors can result from the calibration curve are properly prepared
contaminated glassware especially during and uncontaminated during solution
solution preparation. Unwanted reagents preparation to ensure that the obtained
may enter the prepared stock solutions and absorbance values are accurate.
affect the concentration of copper which will
affect the absorbance value. This will greatly
affect the construction of the calibration
curve which will affect the calculations for the
concentration of the unknown sample. All
concentrations must be uncontaminated
before testing to prevent such errors from
occurring.
4. Conclusion and Recommendations

The quantitative analysis of the copper


solution via spectrophotometry was
successful. An accurate calibration curve
was constructed with the wavelength at
maximum absorption at 541 nm. The graph
was nearly completely linear which allows for
an accurate equation of the line which was
found at y = 0.0001x. Based on the
absorbance value of 0.036, the calculated
concentration of the copper solution was
found at 360 ppm which is true for all trials.
This also ensures all values are precise to
one another for the sample analysis.
However, the calculated concentration of the
copper solution may slightly deviate from the
theoretical value due to the calibration curve
not being completely linear which may have
resulted in some minor errors.

To increase the accuracy of the results, it


is recommended to address the systematic
errors. During the testing cell must be
handled properly and with utmost care to
prevent unwanted errors. The cell must also
be throughly washed to prevent
contamination and dilution from affecting the

You might also like