You are on page 1of 76

{MiscellaneousPUB}{Production}{Health Group}{Confidential}

Publication No. 8H8443


2012-08-13

CARESTREAM DIRECTVIEW Vita/Vita LE/Vita XE


CR System

Troubleshooting Guide

Confidential

© CARESTREAM HEALTH, INC. 2012


TROUBLESHOOTING GUIDE

CARESTREAM DIRECTVIEW Vita/Vita LE CR SystemTroubleshooting Guide


Document No.: 8H8443
Copyright Carestream Health, Inc. 2012
All rights reserved. No part of this manual may be reproduced or copied in any form by any means—graphic,
electronic or mechanical, including photocopying, typing, or information retrieval systems—without written
permission of Carestream Health.

0086
Use of Manual
CARESTREAM CR Systems are designed to meet international safety and performance standards. Personnel
operating the unit must have a thorough understanding of the proper operation of the system. This manual has been
prepared to aid medical and technical personnel to understand and operate the system. Do not operate the system
before reading this manual and gaining a clear understanding of the operation of the system. If any part of this manual
is not clear, please contact your Carestream Health representative for clarification.

Authorized Representatives
Carestream Health France
1, rue Galilée
93192 NOISY-LE-GRAND CEDEX
FRANCE

Carestream Health, Inc.


Rochester, NY 14608
CARESTREAM is a trademark of Carestream Health.
© Carestream Health, Inc., 2012

ii 2012-08-13 – 8H8443
Table of Contents
Description Page

Safety and Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1


Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Document Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
General Safety Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Electrical Hazards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Explosion and Implosion Hazards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Overheating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Vita CR System Laser Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Device-Specific Safety Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Health and Safety Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
CE Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
IEC Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Labelling Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Recycling the Scanner . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
How to Use the Vita CR System Troubleshooting Guide . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
The Chapters in this Manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Vita CR System Hardware Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Troubleshooting Vita CR System Hardware Error Messages. . . . . . . . . . . . . . . . . . . 13
#7 Linear motor stall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
<#5, #6, #9, #10> Rotation motor encoder failure . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
#11 W0 sensor is in wrong state. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
#12 W1 sensor in wrong state . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
#13 Z0 sensor in wrong state . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
#14 Z4 sensor in wrong state . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
#15 Loader motor didn't reach back sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
<#21, #22, #23> Roller motor encoder failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
#24 Screen didn’t reach W0 sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
#41 Cassette release failure. Manually remove the cassette . . . . . . . . . . . . . . . . . . . 27
#61 Linear homing cycle was not completed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
#62 Loader homing cycle was not completed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
#69 The loaded cassette is not secured properly. Please insert the cassette, making sure
that the inserted cassette reached the end. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
#71 Erasing LED module is disconnected . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
#73 Erasing LED module temperature out of range . . . . . . . . . . . . . . . . . . . . . . . . . . 32
#74 Erasing LED module works at reduced capacity . . . . . . . . . . . . . . . . . . . . . . . . . 33
#77 Scanner cover is not in place. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
#78 Internal Controller board communication failure . . . . . . . . . . . . . . . . . . . . . . . . . 35
No License Detected. Click OK for Settings Screen. Click Cancel to Exit . . . . . . . . . 36
Software Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Troubleshooting Vita CR System Software Error Messages . . . . . . . . . . . . . . . . . . . 37
Vita CR Systems Software Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
General Operation Failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
When the cassette is pushed, the CR System Scanner does not load the screen. . . 43
No communication with the CR System scanner, disconnected message. . . . . . . . . 43
No power in the CR System scanner . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Barcode misreading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
The Screen is left outside the cassette after the unloading cycle. . . . . . . . . . . . . . . . 44
The Vita CR System works as Vita at 40 cassettes per hour, when it should work as Vita
XE at 60 cassettes per hour . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Image Artifacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Black Dots on the Image. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
White Dots on the Image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
8H8443 – 2012-08-13 iii
TROUBLESHOOTING GUIDE

Shifted Image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Black Surround Mask (BSM) Failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Grid Lines on the Image. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Inverted Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Vertical White Lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
The Image Becomes Black after Reprocessing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
The Image is with white spots . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
White artifact line across image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
White Band across image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Vertical White strips across the image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Vita CR System Troubleshooting Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Sensors Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Controller Fuses Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Controller Board Test Point—Voltages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Controller Board Dip Switch and LED Indication . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
DIP Switch Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Electric Schematic Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Publication History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71

iv 2012-08-13 – 8H8443
Safety and Regulatory Information

Section 1: Safety and Regulatory Information

Introduction
The information contained herein is based on the experience and knowledge relating to the subject matter gained by
Carestream Health prior to publication. No patent license is granted by this information.
Carestream Health reserves the right to change this information without notice, and makes no warranty, express or
implied, with respect to this information. Carestream Health shall not be liable for any loss or damage, including
consequential or special damages, resulting from any use of this information, even if loss or damage is caused by
Carestream Health’s negligence or other fault.

Document Conventions
In this document the CARESTREAM CR Vita Family Systems will be referred to as “Vita CR System”, “CR Scanner”,
or “System”.

Note
Notes provide additional information, such as expanded explanations, hints, or reminders.

Caution
Cautions point out procedures that you must follow precisely to avoid damage to the system or any of its components,
yourself or others, loss of data or corruption of files in software applications.

Important
Important highlights critical policy information that affects how you use this manual and this product.

General Safety Guidelines


• This product is designed and manufactured to ensure maximum safety of operation. Operate and maintain it in
strict compliance with the safety precautions and operating instructions contained in this manual.
• This product meets all the safety requirements applicable to medical equipment. However, anyone attempting to
operate the system must be fully aware of potential safety hazards.
• Aside from cleaning the phosphor screen and the system’s rollers, there are no user serviceable parts in this
system. The product must be installed, maintained, and serviced by qualified service personnel according to
procedures and preventive maintenance schedules in the product service manual. If your product does not
operate as expected, contact your Service Representative.
• Do not modify this product in whole or in part without prior written approval from Carestream Health.
• Personnel operating and maintaining this system should receive training and be familiar with all aspects of
operation and maintenance.
• To ensure safety, read all user manuals carefully before using the system and observe all Cautions, Importants,
and Notes located throughout the manual.
• Keep this manual with the equipment.
• Reading this manual does not qualify you to operate, test, or calibrate this system.
• Unauthorized personnel are not allowed access to the system.
• If the product does not operate properly or fails to respond to the controls as described in this manual:
– Follow the safety precautions as specified in this manual.
– Stop using the unit and prevent any changes to it.
– Immediately contact the service office, report the problem, and await further instructions.
• Use only legally marketed cassettes. Check periodically the quality of the cassettes, and replace them if any
defects are apparent.

8H8442 – 2012-08-13 1
TROUBLESHOOTING GUIDE

• The images provided by this system are intended as tools for the trained user. They are explicitly not to be
regarded as a sole incontrovertible basis for clinical diagnosis.
• Be aware of the product specifications and of system accuracy and stability limitations. Consider these limitations
before making any decision based on quantitative values. If you have any doubts, consult your Sales
Representative.
• This system is Class I continuous operated stationary equipment without applied parts and has one signal input/
output part.
• The appliance coupler of the flexible cord is used as a disconnecting device.
• In order to guarantee medical-grade leakage current, the computer that is connected to the system should be a
medical-grade computer or connected to the mains through a medical-grade isolation transformer.
• To achieve grounding reliability in USA installations, the equipment must be connected to an equivalent
receptacle marked Hospital Only or Hospital Grade.

Important
When the System is not in use, disconnect the power cable from the electrical outlet.

Electrical Hazards

Caution
• Do not remove or open system covers or plugs. Internal circuits use high voltage capable of causing serious
injury.
• Fuses blown within 36 hours of being replaced by a qualified technician may indicate malfunctioning electrical
circuits within the system. Have the system checked by qualified service personnel. Do not attempt to replace
any fuse.
• Fluids that seep into the active circuit components of the system may cause short circuits that can result in
electrical fires. Therefore, do not place any liquid or food on any part of the system.
• To avoid risk of electric shock, this equipment must only be connected to mains supply with protective earth.

Electromagnetic Emissions/Immunity
• Electromagnetic Compatibility Precautions
– Medical electrical equipment requires special precautions regarding electromagnetic compatibility (EMC).
Medical equipment must be installed and put into service according to the EMC information provided in the
following documentation.
• Communications Equipment
– Portable and mobile radio frequency (RF) communications equipment can affect medical electrical
equipment EMC performance.
• Replacement of Cables, Accessories or Transducers
– The use of cables, accessories, or transducers, other than those specified below with the exception of
transducers or cables sold by the manufacturer of the equipment as replacement parts for internal
components, may result in increased emissions or decreased immunity of the medical equipment.
• Other Equipment
– The Vita CR System should not be used adjacent to or stacked on other equipment. If adjacent or stacked
use is necessary, verify normal operation in the configuration in which the CR Scanner will be used.

2 2012-08-13 – 8H8442
Safety and Regulatory Information

Cable, Accessory, and Transducer Information for the Vita CR System:

Port
Port Type Description From To Cable Type Cable Length
Power AC Power Vita CR System AC Mains Unshielded 3.00 m (9.84 ft.)
Signal USB Vita CR System PC Shielded 3.00 m (9.84 ft.)

Guidance and manufacturer’s declaration—electromagnetic emissions


The Vita CR System is intended for use in the electromagnetic environment specified below. The customer or
the user of the Vita CR System should ensure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment—guidance
RF emissions Group 1 The Vita CR System uses RF energy only for its internal
CISPR 11 function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF emissions Class B The Vita CR System is suitable for use in all
CISPR 11 establishments, including domestic establishments and
those directly connected to the public low-voltage power
Harmonics emissions Class A
supply network that supplies buildings for domestic
IEC 61000-3-2 purposes.
Voltage fluctuations/ Complies
flicker emissions
IEC 61000-3-3

Guidance and manufacturer’s declaration—electromagnetic immunity


The Vita CR System is intended for use in the electromagnetic environment specified below. The customer or the
user of the CR Scanner should ensure that it is used in such an environment.
Electromagnetic environment—
Immunity test IEC 60601 test level Compliance level guidance
Electrostatic ±6 kV contact ±6 kV contact Floors should be wood, concrete or
discharge (ESD) ±8 kV air ±8 kV air ceramic tile. If floors are covered
IEC 61000-4-2 with synthetic material, the relative
humidity should be at least 30 %.
Electrical fast ±2 kV for power supply ±2 kV for power Mains power quality should be that
transient/burst lines supply lines of a typical commercial or hospital
IEC 61000-4-4 ±1 kV for input/output ±1 kV for input/output environment.
lines lines
Surge ±1 kV line to line ±1 kV line to line Mains power quality should be that
IEC 61000-4-5 ±2 kV line to earth ±2 kV line to earth of a typical commercial or hospital
environment.
Voltage dips, < 5 % UT ( > 95 % dip <5 % UT ( > 95 % dip Mains power quality should be that
short in UT) for 0.5 cycle in UT) for 0.5 cycle of a typical commercial or hospital
interruptions 40 % UT (60 % dip in 40 % UT (60 % dip in environment. If the user of the Vita
and voltage UT) for 5 cycles UT) for 5 cycles CR System requires continued
variations on operation during power mains
power supply 70 % UT (30 % dip in 70 % UT (30 % dip in interruptions, it is recommended
lines UT) for 25 cycles UT) for 25 cycles that the Vita CR System be powered
IEC 61000-4-11 <5 % UT ( > 95 % dip in < 5 % UT ( > 95 % dip from an uninterrupted power supply.
UT) for 5 sec. in UT) for 5 sec.

8H8442 – 2012-08-13 3
TROUBLESHOOTING GUIDE

Guidance and manufacturer’s declaration—electromagnetic immunity


The Vita CR System is intended for use in the electromagnetic environment specified below. The customer or the
user of the CR Scanner should ensure that it is used in such an environment.
Electromagnetic environment—
Immunity test IEC 60601 test level Compliance level guidance
Power 3 A/m 3 A/m Power frequency magnetic fields
frequency (50/ should be at levels characteristic of
60 Hz) magnetic a typical location in a typical
field commercial or hospital environment.
IEC 61000-4-8
Note: UT is the AC mains voltage prior to application of the test level.

Guidance and manufacturer’s declaration—electromagnetic immunity


The CR System is intended for use in the electromagnetic environment specified below. The customer or the
user should ensure that it is used in such an environment.
Compliance
Immunity test IEC 60601 test level level Electromagnetic environment—guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the Vita CR
System, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance

Conducted RF IEC 3 Vrms 3 Vrms d = 1.17 ÷P


61000-4-6 150 kHz to 80
MHz
Radiated RF IEC 3 Vrms 3 v/m d = 1.17 ÷P 80 MHz to 800 MHz
61000-4-3 80 MHz to
2.5 GHz
d = 2.33 ÷P 800 M Hz to 2.5 GHz
where P is the maximum output rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
surveya,should be less than the compliance level in
each frequency rangeb.

4 2012-08-13 – 8H8442
Safety and Regulatory Information

Guidance and manufacturer’s declaration—electromagnetic immunity


The CR System is intended for use in the electromagnetic environment specified below. The customer or the
user should ensure that it is used in such an environment.
Compliance
Immunity test IEC 60601 test level level Electromagnetic environment—guidance
Interference may occur in the vicinity of equipment
marked with the following symbol:

Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.

Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.

a.Field from fixed transmitters, such as base station for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the CR scanner is used exceeds the applicable
RF compliance level above, the CR scanner should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the CR
scanner.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 v/m.

Recommended separation distance between portable and mobile RF communications equipment and the
Vita CR System
The Vita CR System is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Vita CR System can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communication equipment (transmitters) and
the Vita CR System as recommended below, according to the maximum output of the communications equipment.
W Rated maximum output
power of transmitter m Separation distance according to frequency of transmitter
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
d = 1.17÷P d = 1.17÷P d = 2.33÷P
0.01 0.117 0.117 0.233
0.1 0.37 0.37 0.737
1 1.17 1.17 2.33
10 3.7 3.7 7.36
100 11.7 11.7 23.3
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.

8H8442 – 2012-08-13 5
TROUBLESHOOTING GUIDE

Explosion and Implosion Hazards

Caution
• Do not operate the equipment in the presence of explosive liquids, vapors, or gases.
• Do not plug in or turn on the system if hazardous substances are detected in the environment. If these
substances are detected after the system has been turned on, do not attempt to turn off the unit or unplug it.
Evacuate and ventilate the area before turning off the system.

Overheating

Caution
Do not block the air circulation around the unit. Always maintain at least 15 cm (6 in.) clearance around the unit to
prevent overheating and damage to the system.

Vita CR System Laser Safety Instructions


The Vita CR System is classified as a Class 1 Laser Product (with the outer cover in place).

Laser Warning
• The Vita CR System contains a Class 3 laser. During normal operation, always keep the unit enclosed in its
protective cover.
• Do not attempt to remove the cover. Only a qualified technician may remove the cover to service this product.

Complies with 21 CFR 1040.10 and 1040.11 except for deviations pursuant to Laser Notice No. 50, dated
2007-06-24.
Class 1 Laser Product, and IEC/EN 60825-1.
Class 1 Equipment, and IEC/EN 60601-1.
Intended for continuous operation.
Product is provided with ordinary protection against the harmful ingress of water.
Not suitable for use in the presence of a flammable anesthetics mixture with air or with oxygen or with nitrous oxide.
The use of accessory equipment not complying with the equivalent safety requirements of this equipment may lead
to a reduced level of safety of the resulting system. Consideration relating to the choice shall include:
• Use of the accessory in the patient vicinity
• Evidence that the safety certification of the accessory has been performed in accordance with IEC 60601-1 or
the system to IEC 60601-1-1 or local equivalent

Device-Specific Safety Information

LIFTING HAZARD

The Vita CR System weighs <36 kg (79 lb). Do not try to lift the unit by
yourself. Always seek assistance from another person. Lifting equipment
that is too heavy may result in serious injury and/or damage to
equipment.

6 2012-08-13 – 8H8442
Safety and Regulatory Information

Health and Safety Compliance


Regulatory Information
The Vita CR System was examined for compliance and has classifications and licenses as follows:

Safety

USA ANSI/AAMI ES6061-1:2005—Part1: Medical Electrical Equipment General Requirements for


safety and essential performance
Canada CAN/CSA 22.2 No. 60601-1-08—Medical Electrical Equipment-Part1: General Requirements for
safety and essential performance
European EN 60601-1: 2006—Medical Electrical Equipment -Part1:General Requirements for safety and
Union essential performance
EN 60825-1: 2007—Safety of Laser Products
International IEC 60601-1: 2005—Medical Electrical Equipment—Part1: General Requirements for safety and
essential performance
IEC 60825-1: 2007—Safety of Laser Products

EMC
IEC 60601-1-2: 2007 Medical Electrical Equipment—Electro-Magnetic Compatibility Requirements & Tests.
EN 60601-1-2: 2007 Medical Electrical Equipment—Electro-Magnetic Compatibility Requirements & Tests.

CE Conformity
This product conforms to the requirements of Council Directive 93/42/EEC. The system bears the following mark of
conformity.

0086

8H8442 – 2012-08-13 7
TROUBLESHOOTING GUIDE

IEC Symbols Used


The system may have labels with one or more of the following symbols. These symbols indicate the IEC standards
to which the system conforms.

Caution—consult accompanying documents

Protective earth points

Power On

Power Off

Caution—Electrical shock hazard

8 2012-08-13 – 8H8442
Safety and Regulatory Information

Labelling Summary

External Labels
Product data plate
—Vita/Vita LE

Product data plate


—Vita XE

8H8442 – 2012-08-13 9
TROUBLESHOOTING GUIDE

Internal Safety Labels

Laser-emitting product

Class 3B laser product inside


unit

High voltage

Protective earth point (chassis


stud)

Refer to the instruction


manual

Recycling the CR Scanner

In the European Union, this symbol indicates that when the last user wishes to discard this product, it must be sent
to appropriate facilities for recovery and recycling. Contact your local representative for additional information on the
collection and recovery programs available for this product.

10 2012-08-13 – 8H8442
How to Use the Vita CR System Troubleshooting Guide

Section 2: How to Use the Vita CR System Troubleshooting Guide

Introduction
The purpose of this manual is to assist field and service engineers when providing service and supporting the Vita
CR Scanners.
This manual is a supplementary document to the CARESTREAM Vita CR Family Service Guide, 8H8442. Please
review the Service Manual before applying the procedures described in this manual.
The manual lists various types of problems, error messages, and detailed explanations for suggested solutions.

The Chapters in this Manual


Hardware Error Messages
Provides information concerning hardware failure-driven error messages and measures to take to
correct them. Each error message contains a code number. A detailed description of the error can be
found in the CarestreamCR.log file, in the User tab.
For each failure, there are recommendations for the user and recommendations for the field/service
engineer.

Software Error Messages


Provides information for all software error messages and measures to take to correct them. For each
failure there are recommendations for the user and recommendations for the field or service engineer.

General CR System Failures


Provides information for general failures in the CR System Scanner and how to fix them.

Image Artifacts
Provides images of all known artifacts, the causes for each of them, and the steps to take to fix the
problem.

Tests
Provides information for general tests that are used in some of the chapters above.

8H8443 – 2012-08-13 11
TROUBLESHOOTING GUIDE

12 2012-08-13 – 8H8443
Vita CR System Hardware Error Messages

Section 3: Vita CR System Hardware Error Messages

Troubleshooting the Vita CR System Hardware Error Messages


You can access the specific failure directly by clicking the relevant line in the table below.

Error No. Error Message Text


7 #7 Linear motor stall
5, 6, 9, 10 <#5, #6, #9, #10> Rotation motor encoder failure
11 #11 W0 sensor is in wrong state
12 #12 W1 sensor in wrong state
13 #13 Z0 sensor in wrong state
14 #14 Z4 sensor in wrong state
15 #15 Loader motor didn't reach back sensor
21, 22, 23 <#21, #22, #23> Roller motor encoder failure
24 #24 Screen didn’t reach W0 sensor
41 #41 Cassette release failure. Manually remove the cassette
61 #61 Linear homing cycle was not completed
62 #62 Loader homing cycle was not completed
69 #69 The loaded cassette is not secured properly. Please insert the cassette,
making sure that the inserted cassette reached the end Cassette size:
<dimensions read> Cassette SN: <serial number read>
71 #71 Erasing LED module is disconnected
73 #73 Erasing LED module temperature out of range
74 #74 Erasing LED module works at reduced capacity
77 #77 Scanner cover is not in place
78 #78 Internal Controller board communication failure
No License Detected. Click OK for Settings Screen. Click Cancel to Exit

8H8443 – 2012-08-13 13
TROUBLESHOOTING GUIDE

#7 Linear motor stall


Scenario
The stall detection sensor does not detect linear motor movement before beginning a scan cycle. Either the linear
motor is not moving, the stall detection sensor is not functioning, or the linear screw is not spinning.

Hardware and Software Response


[1] The CR Scanner is stalled.
[2] The error message appears.
[3] The CR Scanner operating status changes to State stalled.

Recommended Actions
User Response
[1] Click OK.
The system resets itself.
[2] Insert the cassette again.
[3] Does the error repeat?
Yes No
Contact Service. Continue using the CR Scanner.

Service Response
[1] Install V3.2.2 or higher

Note
This may be a faulty error message that has been solved by V3.2.2
[2] Remove the CR Scanner cover.
[3] Insert the technician key.
[4] Activate the linear motor using the Diagnostics screen.

14 2012-08-13 – 8H8443
Vita CR System Hardware Error Messages

[5] Does the linear screw move?


Yes No
Replace the Verify that the linear motor coupling is firmly fastened from both sides.
controller board. Verify that the installed stall detection wheel is of the following configuration:

Does the stall detection wheel appear as shown?


Yes No
a. If there is still a problem, check that the power Replace it according to
supply is producing 24 V (dc) and 5 V (dc) “Replacing the Stall
according to the PS test, See “Controller Board Detector Wheel” section
Test Point—Voltages” on Page 61., OR in the service manual.
b. Check the linear motor connectors (J6) to the
controller board, OR
c. Replace the linear motor, OR
d. Replace the controller board.

8H8443 – 2012-08-13 15
TROUBLESHOOTING GUIDE

<#5, #6, #9, #10> Rotation motor encoder failure


Scenario
The CR Scanner checks the rotation motor during the initialization stage and before beginning a scan cycle. If the
rotations motor does not respond, or it is below 10 RPS, the scanning cycle is not performed.

Hardware and Software Response


[1] The CR Scanner is stalled.
[2] The error message appears.
[3] The CR Scanner will not begin a scanning procedure until the problem is corrected.

Recommended Actions
User Response
[1] Reset the CR Scanner by turning it OFF and ON again.
[2] Does the error repeat?
Yes No
Contact Service. Continue using the CR Scanner.

Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Turn the CR Scanner OFF and ON.
[4] Activate the rotation motor by means of the Diagnostics screen.
[5] Does the rotation motor spin?
Yes No
a. Exit the Diagnostic a. Check all 5 cable connections on the laser driver board.
mode. b. Check that the orange flex rotation motor cable is connected to the optical head
b. Continue using the and all corresponding pins are in place (see the graphic following this table).
CR Scanner. c. Check the blue flex cable connector between the laser driver board J1 and the
controller board.
d. Check fuse F1 2.5 A. Is the fuse still fucntional?

fuse F1

blue flex cable

Yes No
Replace the laser driver board. Replace fuse F1.
If the fuse burns again, then replace the
blue flex cable.

16 2012-08-13 – 8H8443
Vita CR System Hardware Error Messages

Orange Flex Rotation Motor Cable and Corresponding Pins

corresponding
orange flex
pins
rotation motor cable

8H8443 – 2012-08-13 17
TROUBLESHOOTING GUIDE

#11 W0 sensor is in wrong state


Scenario
The CR Scanner indicates that the W0 sensor is in a failed state—it is in the 1 position instead of the 0 position. The
sensor is shown as being blocked.

Hardware and Software Response


[1] The screen is not loaded.
[2] The CR Scanner is halted.
[3] The CR Scanner operating status changes to State stalled.

Recommended Actions
User Response
[1] Click OK.
[2] Reset the CR Scanner by turning it OFF and ON again.
[3] Does the error repeat?
Yes No
Contact Service. Continue using the CR Scanner.

Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Check for an existing screen inside the drum that is blocking the sensor W0.
[4] Clean the feeder rollers.
[5] Check that the cable W0 is connected to the erase LED board. If the cable is connected, proceed to the next
action.

Note
In some cases, the loader pin is not at the correct height and doesn’t pull out the screen from the cassette during the
loading process.
[6] Use Jig MP001520 to verify the loader pin height and adjust it if needed. See Loader Pin Height Adjustment
in the Service Guide.

18 2012-08-13 – 8H8443
Vita CR System Hardware Error Messages

[7] If the erase LED type is as shown below, replace the LED module.

[8] If the erase LED type is as shown below, replace the sensor W0.

8H8443 – 2012-08-13 19
TROUBLESHOOTING GUIDE

#12 W1 sensor in wrong state


Scenario
The CR Scanner indicates that the W1 sensor is in failed state—it is in the 1 position instead of the 0 position. The
sensor is shown as being blocked.

Hardware and Software Response


[1] The screen is not unloaded from the cassette.
[2] The CR Scanner is halted.
[3] The CR Scanner operating status changes to State stalled.

Recommended Actions
User Response
[1] Click OK to acknowledge the error message.
[2] Reset the CR Scanner by turning it OFF and ON again.
[3] Does the error repeat?
Yes No
Contact Service. Continue using the CR Scanner.

Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Check for an existing screen inside the drum that is blocking the W1 sensor.
[4] Replace the erase LED assembly.

20 2012-08-13 – 8H8443
Vita CR System Hardware Error Messages

#13 Z0 sensor in wrong state


Scenario
The screen carriage gets stuck while the screen is being unloaded or while performing an erase sequence. As a
result, the screen carriage does not reach the Z0 sensor and it does not change its position from 1 to 0.

Note
The error message will appear after 25 seconds of unsuccessful unloading sequence attempts.

Hardware and Software Response


[1] The screen is returned to the cassette.
[2] The cassette is still locked in the Scanner.
[3] The CR Scanner operating status changes to State stalled.

Recommended Actions
User Response
[1] Click OK.
[2] Manually release the cassette using the release knob under the cassette tray.
[3] Remove the cassette.
[4] Reset the CR Scanner by turning it OFF and ON again.
[5] Insert the cassette.
[6] Does the error repeat?
Yes No
Contact Service. Continue using the CR Scanner.

Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Is the screen carriage stuck?
Yes No
a. Manually move the screen carriage to its homing position. Check/replace the Z0 sensor (“Sensors
b. Adjust the drum rails with Jig AS000723 according to the Tests” on page 59.
instructions in the Service Manual
c. Turn the CR Scanner ON.
d. Insert a cassette. Is there an object interfering with the
screen carriage movement?
Yes No
Remove the object that is Replace the screen carriage.
interfering.

8H8443 – 2012-08-13 21
TROUBLESHOOTING GUIDE

#14 Z4 sensor in wrong state


Scenario
While loading the screen, the Z4 sensor is in the wrong position, preventing the screen from being loaded into the
drum. The screen is displayed as being in the drum.

Hardware and Software Response


[1] The screen is not loaded.
[2] The CR Scanner operating status changes to State stalled.

Recommended Action
User Response
[1] Click OK.
[2] Reset the CR Scanner by turning it OFF and ON.
[3] Does the error repeat?
Yes No
Contact Service. Continue using the CR Scanner.

Service Response
[1] Click OK to acknowledge the error message.
[2] Insert the technician key.
[3] Check/replace the Z4 sensor (see “Sensors Tests” on page 59).

22 2012-08-13 – 8H8443
Vita CR System Hardware Error Messages

#15 Loader motor didn't reach back sensor


Scenario
• While loading the screen, the CR Scanner indicates that the loader pin did not return to its back position (sensor
L1). Time-out is 4 seconds.
• Stepper motor failure

Hardware and Software Response


[1] There might be a mechanical noise.
[2] The error message appears.
[3] Click OK.
The screen is returned to the cassette and the cassette is ejected.
[4] The CR Scanner operating status returns to “Insert cassette”.

Recommended Actions
User Response
[1] Reset the CR Scanner by turning it OFF and ON again.
[2] Insert the cassette again.
[3] Does the error repeat after unloading the screen?
Yes No
Contact Service. Continue using the CR Scanner.

Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Activate the stepper loader motor by means of the Diagnostics screen.
[4] Is there a mechanical noise while operating the motor?
Yes No
Check/replace the L1 sensor (see “Sensors Tests” a. Remove the CR Scanner base cover
on page 59). (refer to the Service Manual).
Or b. Verify that the loader stepper motor
Replace the controller board. coupling is firmly fastened from both
sides.
c. Replace the loader assembly.

8H8443 – 2012-08-13 23
TROUBLESHOOTING GUIDE

<#21, #22, #23> Roller motor encoder failure


Scenario
During the load/unload sequence, the encoder does not detect the roller motor speed. As a result, the CR Scanner
cannot complete the load/unload sequence.

Hardware and Software Response


[1] The load/unload sequence is halted.
[2] The error message appears.
[3] The CR Scanner operating status changes to “State stalled”.

Recommended Actions
User Response
[1] Click OK to acknowledge the error message.
[2] Reset the CR Scanner by turning it OFF and ON again.
[3] Does the error repeat?
Yes No
Contact Service. Continue using the CR Scanner.

Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Turn the CR Scanner OFF and ON.
[4] Activate the roller motor by means of the Diagnostics screen.
[5] Does the roller motor spin?
Yes No
a. Check the encoder reader cable a. Check the roller motor cable connector.
connection to the controller board b. Check that the power supply is producing 24 V (dc)
J1. and 5 V (dc).
b. Replace the roller encoder reader. c. Adjust the encoder and reader encoder of the
roller motor with Jig AS000703 and MP001890,
See Roller Reader Encoder Positioning in the
service guide.
Is there still a problem after adjusting the encoder
and reader encoder of the roller motor?
Yes No
a. Replace the controller Replace the 24 V (dc) or 5
board. OR V (dc) power supply.
b. Replace the roller
motor.

24 2012-08-13 – 8H8443
Vita CR System Hardware Error Messages

#24 Screen didn’t reach W0 sensor


Scenario
The user inserts a cassette and one of the following occurs:
• The CR Scanner fails to load the screen to the drum.
• The screen is loaded to the drum but the CR Scanner reports that the W0 sensor did not change its position when
the screen was loaded.

Note
The error message will appear if one of the above problems occurs.

Hardware and Software Response


[1] The screen is not loaded.
[2] The CR Scanner is halted.
[3] The CR Scanner operating status changes to “State stalled”.

Recommended Actions
User Response
[1] Check if the screen is in the cassette.
Yes No
a. Insert the cassette into the CR a. Insert the screen into the cassette. Make sure
Scanner. that you insert the screen right side up.
b. Go to Step 2. b. Insert the cassette into the CR Scanner.
c. Go to Step 2.

[2] Does the error message repeat?


Yes No
a. Insert the cassette. Continue Using the CR Scanner.
b. Does the error message repeat?
Yes No
Contact Service. Continue using the CR
Scanner.

Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Insert the cassette into the CR Scanner.

8H8443 – 2012-08-13 25
TROUBLESHOOTING GUIDE

[4] Is the screen loaded to the drum?


Yes No
Check if the erase LED type is as shown in a. Repeat Step 1 of the User Response.
the picture below, replace the Erase LED b. Check that the roller motor coupling is
Module. firmly fastened from both sides.
c. Check the function of the loader motor.

Check if the erase LED type is as shown in


the picture below, replace the W0 sensor.

26 2012-08-13 – 8H8443
Vita CR System Hardware Error Messages

#41 Cassette release failure. Manually remove the cassette


Scenario
• The loader moves toward the forward end of travel (sensor L0) in order to release the cassette, but when it
reaches L0 the cassette is still detected as present in the CR Scanner by the Casette Presence Sensor (CPS).
An error message is generated.
• Sensor L0 is not working or not adjusted, the motor is not working, or the motor coupling is disconnected.

Hardware and Software Response


[1] The error might be accompanied by a mechanical noise.
[2] The screen is inside the cassette.
[3] The CR Scanner is stuck until the cassette is released.

Recommended Actions
User Response
[1] Manually release the cassette using the release knob under the cassette tray.
[2] Reset the CR Scanner by turning it OFF and ON.
[3] Try to load the cassette again.
[4] Does the message appear again?
Yes No
a. Upload a different cassette. Continue working with the CR
b. Does the message re-appear? Scanner.
Yes No
Contact Service. a. There is a problem with the first
cassette. Inform Service.
b. Continue working with the CR
Scanner.

Service Response
Is it a cassette problem or a CR Scanner problem?

CR Scanner Problem Cassette Problem


a. Check/replace the CPS sensor. Replace the cassette only. Insert
b. Is the problem solved? the old screen into a new cassette.
Yes No
Check the other cassettes. a. Adjust sensor L0 to the right
position so that the loader
motor moves all the way to its
forward position.
b. Check the other cassettes.

8H8443 – 2012-08-13 27
TROUBLESHOOTING GUIDE

#61 Linear homing cycle was not completed


Scenario
The linear assembly is stuck at the left limit sensor (X1) and therefore does not reach the right limit sensor (X0). Time-
out is 55 seconds.
The linear assembly gets stuck at the right limit sensor (X0).

Hardware and Software Response


[1] The CR Scanner operating status is changed to Linear motor homing.
[2] The CR Scanner will perform a homing loop until the problem is corrected.

Note
The error message will appear every 55 seconds after clicking OK.

Recommended Actions
User Response
[1] Reset the CR Scanner by turning it OFF and ON.
[2] Does the error repeat?
Yes No
Contact Service. Continue using the CR Scanner.

Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Go to the Settings>Diagnostics tab.
[4] Does the X1, X0 LED illuminate according to the linear position on the pictogram?
Yes No
Check/replace the right limit sensor. Check/replace the left limit sensor.

28 2012-08-13 – 8H8443
Vita CR System Hardware Error Messages

#62 Loader homing cycle was not completed


Scenario
• While loading the screen, an error message indicates that the loader pin did not return forward to its start position
(time-out is 5 seconds) and did not activate the loader forward sensor (L0).
• While the loader performs the homing sequence, the loader moves away from sensor L0 toward L1. If the axis
does not deactivate the sensor within 4 seconds, an error message is generated.

Hardware and Software Response


[1] The error might include a mechanical noise.
[2] The screen is returned to the cassette and the cassette is ejected.
[3] The error message appears.

Recommended Actions
User Response
[1] Reset the CR Scanner by turning it OFF and ON again.
[2] Insert the cassette again.
[3] Does the error repeat after unloading the screen?
Yes No
Contact Service. Continue using the CR Scanner.

Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Activate the stepper loader motor by means of the Diagnostics screen.
[4] Does the stepper loader motor move?
Yes No
Is there any mechanical noise while operating the motor? a. Check the stepper motor connection J6
Yes No on the controller board.
Check/replace the L0 sensor a. Open the CR Scanner b. If it still does not work, replace the
(see “Sensors Tests” on base cover (refer to the loader assembly.
page 59). CARESTREAM Vita CR
Family Service Guide).
b. Verify that the loader
stepper motor coupling is
firmly fastened from both
sides.

8H8443 – 2012-08-13 29
TROUBLESHOOTING GUIDE

#69 The loaded cassette is not secured properly. Please insert the
cassette, making sure that the inserted cassette reached the end
Cassette size: <dimensions read>
Cassette SN: <serial number read>
Scenario
When the cassette is inserted into the Scanner, the system will wait 1.5 seconds before starting the roller motor. This
will ensure the user pushes the cassette further into lock position (some users push the cassette slowly into the
reader slot and stop when the roller motor starts so the cassette might be still unlocked).
In Light-QC single-mode, the QC application will wait 5 seconds before closing, to ensure error #69 is displayed to
the user if needed. The message will indicate to the user to properly insert the cassette so the screen can unload
properly.

Recommended Actions
User Response
Verify that the cassette you are attempting to insert into the system matches the same size and number reported in
the error message. Doing so will ensure normal operation.

Note
Clicking OK on the error message without verifying that the correct cassette is being used may result in:
• A problem while ejecting the cassette, and the possibility of error #41 occurring
• The screen will not fully return into the cassette

30 2012-08-13 – 8H8443
Vita CR System Hardware Error Messages

#71 Erasing LED module is disconnected


Scenario
Prior to performing an erase sequence, the LED module does not receive input from the controller board and the
scan is stalled.

Hardware and Software Response


[1] The CR Scanner does not perform an erase sequence.
[2] The error message appears.

Recommended Actions
User Response
[1] Click OK.
The screen is returned to the cassette and the cassette is ejected.
[2] The CR Scanner operating status is changed to State stalled.
[3] Reset the CR Scanner by turning it OFF and ON again.
[4] Insert the cassette again.
[5] Does the error repeat after unloading the screen?
Yes No
Contact Service. Continue using the CR Scanner.

Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Activate the erase LED by means of the Diagnostics screen.
[4] Do any of the LED indicators illuminate red?
Yes No
Replace the erase
LED assembly.

a. Check/replace the cable J7 on


the LED driver board.
b. Replace the erase LED
assembly.

[5] Is the 3.3 V (dc) at red wire of J7 connector with erase LED off?
Yes No
Replace the controller board

[6] Is the 0 V (dc) at red wire of J7 connector with erase LED on?
Yes No
Replace the controller board

8H8443 – 2012-08-13 31
TROUBLESHOOTING GUIDE

#73 Erasing LED module temperature out of range


Scenario
The erase LED temperature is over 70 °C (158 °F).

Hardware and Software Response


[1] The screen is not unloaded into the cassette.
[2] The CR Scanner is halted.
[3] The CR Scanner operating status changes to “State stalled”.

Recommended Actions
User Response
[1] Click OK.
[2] Turn the CR Scanner OFF and wait 45 minutes before turning it ON again.

Note
This is the time needed to reduce the temperature of the erase LED.
[3] Check room temperature according to site specifications.
[4] Insert the cassette again, after 45 minutes, and scan.
[5] Does the error repeat?
Yes No
Contact Service. Continue using the CR Scanner.

Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Activate the erase LED by means of the Diagnostics screen.
[4] Does the LED indicator light up red to the right of the word “Temperature”?
Yes No
Replace the LED assembly.

a. Check the J1 connector on the


erase LED driver board.
b. Replace the erase LED assembly.

32 2012-08-13 – 8H8443
Vita CR System Hardware Error Messages

#74 Erasing LED module works at reduced capacity


Scenario
One or more segments of the erase LEDs burned out.

Hardware and Software Response


The CR Scanner will continue to operate normally. However, the screen will be erased partially in every scanning
cycle.

Recommended Actions
User Response
[1] Click OK.
[2] Manually erase the cassette after every scanning cycle.
[3] Contact Service.

Note
The erase LED assembly needs to be replaced.

Service Response
[1] Remove the CR Scanner cover.
[2] Insert the technician key.
[3] Turn ON the CR Scanner.
[4] Activate the erase LED from the Diagnostic screen
[5] Does the LED indicator to the right of the word Current light up red?
Yes No
Replace the Erase LED assembly.

a. Check the J2 and J6 connectors on


the erase LED board.
b. Check the F1 fuse on the erase
LED board.

[6] Check for 24 V (dc) input at the F1 fuse on the erase LED assembly.
Yes No
Replace the erase LED assembly. Replace the erase 24 power supply.

8H8443 – 2012-08-13 33
TROUBLESHOOTING GUIDE

#77 Scanner cover is not in place


Scenarios
The CR Scanner does not operate when the cover is removed.

Hardware and Software Response


[1] The CR Scanner will not operate.
[2] The error message appears.

Recommended Actions
Service Response
[1] Click OK to acknowledge the error message.
[2] Insert the technician key.
[3] Operate the CR Scanner. Does the message appear:

Yes No

a. Check sensor cover is connected to Check the existing magnet on the system
J16 on the controller board. cover and its position.
b. If it is, then replace the cover sensor.

34 2012-08-13 – 8H8443
Vita CR System Hardware Error Messages

#78 Internal Controller board communication failure


Scenario
While updating the motion version, there is no internal communication between the controller board and the motion
chip.

Hardware and Software Response


The CR Scanner will not operate.

Recommended Actions
User Response
[1] Reset the CR Scanner by turning it OFF and ON again.
[2] Insert the cassette again.
[3] Does the error repeat after unloading the screen?
Yes No
Contact Service. Continue using the CR Scanner.

Service Response
[1] Go to the About screen.
[2] Is the motion version compatible with the CR Systems software?
Yes No
Replace the controller board. a. Update the motion version by means of the SW Update &
Backup screen.
b. Does the motion version update fail?
Yes No
Replace the controller board. Check the system operation.

8H8443 – 2012-08-13 35
TROUBLESHOOTING GUIDE

No License Detected. Click OK for Settings Screen. Click Cancel to Exit


Scenario
The error appears while trying login to the CR System software, and the proper license is not detected.

Hardware and Software Response


The CR Scanner will not operate.
1. Click Cancel to close the CR System Software. OR
2. Click OK, go to Settings>Technical Mode.

Recommended Actions
User Response
Verify that the proper licenses exist.
[1] In the CR System software, select Settings>Settings>About.
[2] Click License Manager.
[3] Check that the relevant licenses appear checked in the License Options window.

Service Response
[1] Open the iButton cover on the Scanner’s rear panel. It is next to the USB connector.
[2] Verify that an iButton is inserted correctly into the socket.

[3] If the iButton is installed correctly, verify that the CR System software detects the iButton and the licenses. See
the User Response above.

If the CR System Software Does Not Detect a License


[1] Open the CR Scanner’s rear panel by removing the four screws.
[2] Verify that the interface cable between the iButton board and the Controller board is properly connected, and
no wires are loose. See below.
[3] Verify that the CR System software detects the iButton and the licenses. See User Response above.

[4] If the CR System still does not detect the iButton or licenses.
(a) Replace the Internal USB cable.
(b) Replace the iButton electronic board.
(c) Replace the iButton. Verify that the iButton has the right licenses.
[5] Connect a WIBUKEY (license dongle) to the PC. Verify that a license is present on the System.
36 2012-08-13 – 8H8443
Software Error Messages

Section 4: Software Error Messages

Troubleshooting Vita CR System Software Error Messages


The CR Scanner software sends various event information, such as wrong steps taken by the user, notes, and error
messages. These system events are presented to the user in pop-up screens.

Software Error Descriptions


Errors are reported by the system in several categories:
• “CR Systems Software User Interface” on Page 38
• “Setting Screen in the Technician Interface” on Page 41
• “Exam Screen in the Software User Interface” on Page 40
• “Procedure Mapping Screens in the Software User Interface” on Page 41

8H8443 – 2012-08-13 37
TROUBLESHOOTING GUIDE

Vita CR Systems Software Error Messages


CR Systems Software User Interface
Note
The error numbers in the list below are for user convenience only.

Error
No. Error Message Text Possible Reason and Solution
1 Cassette was ejected due to two minute time-out. The user inserted a cassette, but didn’t scan within
two minutes. The software is notified and generates
an abort command.
Insert the cassette again to perform a scan.

2 The following image(s) should be Accepted/ The user didn’t choose the Reject or Accept option.
Rejected first.
3 CR Scanner is ready for rollers cleaning. Please The technician/user needs to complete the rollers
insert cleaning tray and then insert cleaning cleaning process.
Plate.
4 Pull out the cleaning Plate and then release the The technician/user needs to complete the rollers
cleaning tray. cleaning process.
5 The cassette does not match the selected sub- The cassette type is not compatible to the sub-
organ. Please either replace the cassette or re- organ or to BMD scanning type.
select sub-organ.
6 Press Cassette Size button and select cassette. There is a barcode failure or an unrecognized
cassette barcode.
7 No available demo image for the selected sub- There is no existing demo image for the selected
organ. sub-organ in the database. Not all sub-organs have
demo images.
8 The machine is disconnected. Changes cannot 1. Check the USB cable connections.
be saved, except the change to the scan mode. 2. Check power cord connections.
3. Restart the scanner. If the error repeats, call for
service.

9 The scanner is disconnected 1. Check the USB cable connections.


2. Check the power cord connections.
3. Restart the scanner. If the error repeats, call for
service.
10 A restart is needed for all changes to take effect Click OK, then exit and restart the CR System
Software.
11 Erasing process was stopped. Clicking Ok wil Acknowledge the error. Manually erase the
eject the cassette. cassette. If the error repeats, call for service.
Note: The screen will not be erased. Manually
erase the cassette after error message is
received.
12 Scanner internal temperature is out of range. Check the room temperature according to the site
specifications. Turn off the scanner. After
adjusting the room temperature, wait a few
minutes before turning on the scanner.

38 2012-08-13 – 8H8443
Software Error Messages

Error
No. Error Message Text Possible Reason and Solution
13 Another instance of CR Systems 1. Wait a few seconds to see if the CR Systems
Software already exists! Software runs.
2. Try to open the CR Systems Software again.
3. Restart the PC.
14 Failed to read the machine’s settings. The Restart the scanner. If the error repeats, call for
application can’t proceed. Please contact service. service.
15 Failed to store image information into the 1. Check the disk space.
database 2. Re-launch CR Systems Software; check if
Image Storage reprocessed images have
unread status.
3. If the error repeats, call for service.
16 The following image(s) should be Accept or reject the pending images before
Accepted/Rejected first: continuing.
17 Warning: Not enough space on disk. Please clear 1. Check the disk space.
some disk space. 2. Free at least 1 GB of disk space.
18 This User does not have the required User permission is denied. If necessary,
permissions to access this utility. reconfigure the user permissions.
19 A patient with the same ID already exists in the When editing a patient, the ID may have been
System. changed to an ID that already exists. Edit the patient
data again to correct the ID information.

8H8443 – 2012-08-13 39
TROUBLESHOOTING GUIDE

LLI Cassette Error Messages


Note
The error numbers in the list below are for user convenience only.

Error
No. Error Message Text Possible Reason and Solution
1 Can’t choose BOTTOM before TOP was scanned Insert LLI 1—top before LLI 2—bottom or insert
or scan TOP twice. Operation is cancelled. LLI 2—bottom after scanning LLI 1—top.
2 The cassette is not an LLI cassette. Please either The cassette type is not compatible to the sub-
replace the cassette or re-select sub-organ. organ.
3 Can’t choose a sub-organ which is not for LLI If working within an LLI procedure, the user can’t
during an LLI study. Operation is cancelled. choose a sub-organ that is not part of a LLI
procedure.
4 Can’t choose LLI sub-organ after regular sub- To perform an LLI sub-organ, go to the patient list,
organ was chosen. Operation is cancelled. reselect the patient, and perform only LLI sub-
organ.
5 The cassette Barcode ID does not match the top Check that the Barcode ID label on the LLI 1
side. Operation is cancelled. cassette is readable.

Exam Screen in the Software User Interface


Note
The error numbers in the list below are for user convenience only.

Error No. Error Message Text Possible Reason and Solution


1 Search in database failed The query to the database failed.
2 Image Search in database failed The query of an image from the database failed.
3 Global status count failed The system failed to get the number of studies
from the database.
4 Destinations/Jobs Search in database failed The system failed to find a destination or job in the
database.
5 Failed to delete processed image An incomplete job may exist.
6 There is no image selected (or) The user did not choose an image.
Make sure there is only one image selected
7 Failed to send one or more images Send all failed images operation failed.
8 Failed to send image The system failed to send an image to the printer.
9 Failed to show image thumbnail A failure occurred when trying to display an image
in the thumbnail.
10 Log file not found The log file was not found when selecting Log
File.
11 No such patient in database The patient was not found in the database.
12 Update procedure failed! The user is trying to edit the patient details.
13 Patient with same ID already exists. Click OK to continue.

40 2012-08-13 – 8H8443
Software Error Messages

Procedure Mapping Screens in the Software User Interface


Note
The error numbers in the list below are for user convenience only.

Error No. Error Message Text Possible Reason and Solution


1 This User does not have the required permissions User permission is denied.
to access this utility. If necessary, reconfigure user permissions
2 Procedure Mapping Tool application is already The user tries to open Procedure Mapping tool
running. more than once.
3 Please fill a procedure name, The user needs to write the procedure name
inside the tab frame.
4 A procedure with this name already exists The procedure name is duplicated.
5 Please select an item No item was selected when clicking ADD.
6 A restart is needed for all changes to take effect Restart the CR Systems Software.

Setting Screen in the Technician Interface


Note
The error numbers in the list below are for user convenience only.

Error No. Error Message Text Possible Reason and Solution


1 Failed opening the file. The system failed to open the files that were
requested for updating. (FW = iic, FPGA = rpd and
Motion=s).
2 The file is in a wrong format. The selected file for import operation is in an
incorrect format (for example: not an XML file or
switching between Anatoms and ScannerSettings).
3 Barcode CR Scanner fails. Check the Barcode operation.
4 Value is out of range. This can happen during calibration. Recalibrate.

DICOM Error-related Messages


Note
The error numbers in the list below are for user convenience only.

Error
No. Error Message Text Recommended Action
1 Failed to store the DICOM Send information for 1. Check the network settings.
this image. 2. Repeat the DICOM Send action.

2 Please note: Modality query has failed. 1. Check the RIS network settings.
Information retrieved from the last successful 2. Check the Modality Worklist settings.
query.
3. Repeat the RIS query.
3 Please note; Modality Query has failed. There is no access to RIS\HIS. Cannot present a
study list.
4 Note: Host name and Port number are invalid. The DICOM Settings needs to be reconfigured.
MWL mode will be disabled until those values
configured via the DicomSetting tool.

8H8443 – 2012-08-13 41
TROUBLESHOOTING GUIDE

Parameters Saved on System Memory


Note
The error numbers in the list below are for user convenience only.

Error No. Error Message Text Possible Reason and Solution


1 This operation will require re-calibration of the CR The technician selected Restore to Factory.
Scanner. Are you sure you want to continue?
2 This operation will override existing CR Scanner The technician selected Import Scanner
parameters. Are you sure you want to continue? Settings.
3 Failed to read the machine’s settings. The The system cannot read the machine S.N. from
application can’t proceed, please contact service. the System memory. The settings will be loaded
from the PC (calib folder).
4 Error 81- No FPGA The FPGA may be corrupted, reload firmware and
FPGA.

42 2012-08-13 – 8H8443
General Operation Failure

Section 5: General Operation Failure

When the cassette is pushed, the CR Scanner does not load the screen
Suggested Solution
Is there a “Screen didn’t reach W0 Sensor” error message?

Yes No
Go to “#24 Screen didn’t reach W0 a. Open the cover.
sensor” on Page 25 b. Check the Casette Presence Sensor (CPS) See “Sensors Tests” on
Page 59

No communication with the CR scanner, disconnected message


Solution
[1] Check the power cable connection and that CR Scanner is turned ON.
[2] Connect to a different PC with the required instructions according to the specifications.
[3] Replace the USB cables connecting the PC and the CR Scanner.
[4] Check the cable between the USB socket and the controller board.
[5] Replace the controller board.

Note
If one of the steps solves the problem, do not continue with the following steps.

No power in the CR scanner


Solution
[1] Check the power cable connection and that CR Scanner is turned ON.
[2] Turn the CR Scanner OFF.
[3] Check the two fuses (3A) in the power cable connector.
[4] Remove the CR Scanner cover.
[5] Check the power supply voltage (see “Controller Board Test Point—Voltages” on Page 61).

Barcode misreading
Workaround
Barcode misreading occurs when the system fails to read the cassette label to indicate cassette size.
To continue, use the manual cassette size setting as follows:
[1] Log in to Settings>User tab.
[2] Select the Display cassette size button check box.
[3] Load the screen. The cassette size list is displayed.
[4] Select the correct cassette size from the list.

Solution
[1] Remove the tray assembly (refer to the Service Manual).
[2] Check if the system has the most current software installed.

8H8443 – 2012-08-13 43
TROUBLESHOOTING GUIDE

[3] Check if the flex cable is connected correctly to the electronic board as shown below.

correct flex cable setup

incorrect flex cable setup

[4] Check that the barcode grounding cable is attached to the barcode reader bracket.

Note
Record this update in the Mod label as Mod 2. In case one of the above updates is not implemented, update the
system according to AT000243.
[5] Make sure the barcode assembly is secured well.
[6] Adjust the barcode position with Jig MP001692. See “Barcode Reader Adjustment” in the CARESTREAM Vita
CR Family Service Guide.
[7] Test the barcode reader is positioned correctly according to the “Barcode Reader Test” procedure in the
CARESTREAM Vita CR Family Service Guide.
[8] Replace the barcode assembly.

The Screen is left outside the cassette after the unloading cycle
Solution
[1] Does this failure occur with every case?

Yes No
a. Check Loader Pin height and its Replace the Cassette
travel with Jigs MP001689 and
MP001691.
b. See “Loader Pin Height
Adjustment” in the
CARESTREAM Vita CR Family
Service Guide.
[2] Check Last software version update.

44 2012-08-13 – 8H8443
General Operation Failure

The Vita CR System works as Vita at 40 cassettes per hour, when it should
work as Vita XE at 60 cassettes per hour
Checking that the system is recognized as Vita XE
The system is recognized as Vita XE under the following criteria:
• Existing new Erasing LED module
• Existing cable RS232
• License Vita XE applied on iButton
• CARESTREAM CR System SW is V3.2.2 or higher
• Firmware FPGA and Motion version
To verify that these criteria are correct, go to the About tab and check as follow:

[1] In Software Information, check:

Testing Criteria Should be In case it doesn’t…


Software V3.2.2 Install V3.2.2 or higher SW
[2] In the Embedded Software information, check:

Testing Criteria Should be In case it doesn’t…


Controller Type 02 Ensure the controller version is AS000786-00 or higher
on the white label located at the top left of the controller
board.
Erasing LED 100005/01 Check you replace the erase LED module for Vita XE
configuration
Controller 10.2.1.01-112 Perform a manual update from the SW Update &
Firmware Backup tab.
For more information, check the service manual

8H8443 – 2012-08-13 45
TROUBLESHOOTING GUIDE

Testing Criteria Should be In case it doesn’t…


Controller FPGA 10.3.1.02-001 Perform a manual update from the SW Update &
Backup Tab.
For more information, check the service manual
Controller Motion 10.4.1.02-02 Perform a manual update from the SW Update &
Backup Tab.
For more information, check the service manual
[3] Click on License Manager and check:

Testing Criteria Should be In case it doesn’t…


Vita XE Marked with V The system doesn’t recognise the iButton , See ““No
License Detected. Click OK for Settings Screen. Click
Cancel to Exit”” on Page 36.
Standard Speed Marked with V 1. The Previous system was Vita LE and it
should be upgraded first to Vita , contact the
sales department for this kit.
2. The system doesn’t recognise the iButton ,
See ““No License Detected. Click OK for
Settings Screen. Click Cancel to Exit”” on
Page 36.

[4] If all of the above criteria are as they “Should be”, then the Vita system should be recognized by the software
as Vita XE, Check in Scanner information – Machine Type Vita XE, else recheck all the above again.

46 2012-08-13 – 8H8443
Image Artifacts

Section 6: Image Artifacts

Black Dots on the Image

Example of Black Dots

Cause
These dots (streaks) are caused by an Electrostatic Discharge (ESD). These ESD dots generate very short sharp
and high light sparks that are coupled into the photo multiplier through the blue filter. The ESD effect is induced by
several factors, mainly environmental conditions such as dry air and low humidity, and the rubbing of the screen by
the mechanical CR Scanner (rollers).

Recommended Actions
Clean the screen with a screen solution that contains an anti-static agent. See the CARESTREAM Vita CR Family
Service Guide for cleaning instructions.

8H8443 – 2012-08-13 47
TROUBLESHOOTING GUIDE

White Dots on the Image

Example of White Dots

Cause
White dots appear on the scanned image. This results from accumulated dust.

Recommended Actions
Clean the screen. See the Service Manual for screen cleaning instructions.

48 2012-08-13 – 8H8443
Image Artifacts

Shifted Image

Example of Shifted Image

Cause
Lines and/or distorted image may be due to incorrect computer specifications.

Recommended Actions
[1] Check that the computer’s USB chipset is the INTEL model.
[2] Reinstall the computer’s USB chipset driver.

Note
The USB chipset driver can be downloaded from the computer model Web site.
[3] Verify that the issue is with the computer by activating the CR Scanner via another computer.
[4] Replace the controller board.

Note
If one of the steps solves the problem, do not continue with the following steps.

8H8443 – 2012-08-13 49
TROUBLESHOOTING GUIDE

Black Surround Mask (BSM) Failure

Examples of Failed Black Surround Mask

Cause
The BSM algorithm failed to detect the collimators blades (in the above left sample image, the right hand side blade
was not detected).
• In cases where an apron is used to collimate the radiation, the curved line will not be detected by the BSM
automatically. It will leave a white collimation area (see the sample image on the above right).
• Implants in the clinical body might be detected as collimation border and the clinical image will be cropped.

Recommended Actions
[1] Select Reprocess in the Scan screen.
[2] Select the Edit Mask checkbox to set the BSM manually.
[3] Click and drag the BSM blades to redefine the Region of Interest (ROI).
[4] Once defined, click Process. The software will apply the BSM only to the area within the frame.
[5] Accept the image in order to send it to the PACS.

Note
For details on using Black Surround Mask, see the Vita CR System Software User Guide.

50 2012-08-13 – 8H8443
Image Artifacts

Grid Lines on the Image

Example of Grid Lines on Image

Note
The size of the grid lines varies with the magnification.

Cause
There are grid lines on the image.

Recommended Actions
[1] Go to the Technician>Settings>Setup screen.
[2] Check that Grid Suppression Filter is selected.
[3] Check that the grid is 40 lines per cm (103 lines per in.).

8H8443 – 2012-08-13 51
TROUBLESHOOTING GUIDE

Inverted Screen

Example of an Inverted Screen

Cause
The screen was inserted the wrong way in the cassette.

Recommended Actions
[1] With the screen extraction tool, remove the screen from the cassette.
[2] Re-insert the screen with the correct side up.

52 2012-08-13 – 8H8443
Image Artifacts

Vertical White Lines

Note
The lines are vertical to the scan. In the example on the right, the image was rotated.
Examples of Vertical White Lines

Cause
The USB chipset on the computer mother board is not working.

Recommended Actions
[1] Check that the PC USB Chipset is the Intel model.
[2] Reinstall PC USB Chipset driver. It can be downloaded from the PC model Web site.
[3] Verify that the issue is with the PC by activating the CR Scanner via another PC.

Note
If one of the steps solves the problem, do not continue to the following steps.

8H8443 – 2012-08-13 53
TROUBLESHOOTING GUIDE

The Image Becomes Black after Reprocessing

Example of Raw Image Becomes Black after Reprocessing

Cause
The image appears black after image processing even though the raw image is fine. This can be caused by the
monitor having the incorrect resolution or the wrong DICOM tag.

Recommended Actions
[1] Check that the monitor settings are 1024 x 768 with a frequency of 74 Hz.
[2] For a high resolution monitor (1–3 m), select grayscale.
[3] Ensure that the DICOM tag (00028,00103) pixel representation value is “0” or “1”.

54 2012-08-13 – 8H8443
Image Artifacts

The Image has White Spots

Example of Image

Cause
The image presented with white spots even after exposure—no image was shown.

Recommended Action
Replace the PM board.

8H8443 – 2012-08-13 55
TROUBLESHOOTING GUIDE

White Artifact Line


White line appears across the image

Cause
The image appears with a white band artifact across the image. It may be caused due to damage on the screen or
to an interfering object in the optical laser head assembly.

Recommended Action
[1] Remove the screen from the cassette, check and clean the screen, or replace it if necessary.
[2] Replace the optical laser head assembly.

Note
After replacing the optical head assembly, you must recalibrate the system: Offset and E-gain calibration.

56 2012-08-13 – 8H8443
Image Artifacts

White Band Across Image

White Band Appears on the Image

Cause
The image appears with a white band artifact across the image. It may be caused due to an interfering object in the
optical laser head assembly.

Recommended Actions
Replace the optical laser head assembly.

Note
After replacing the optical head assembly, you must recalibrate the system: Offset and E-gain calibration.

8H8443 – 2012-08-13 57
TROUBLESHOOTING GUIDE

Vertical White Strips Across the Image

Cause
A radio therapy generator may be near the Vita system and was influencing the system while it was switched On.

Recommended Action
[1] Clear the Vita system environment from any powerful generator. For more information, see Electrical Hazards
in the CARESTREAM Vita Safety and Regulatory Guide (8H8439).

58 2012-08-13 – 8H8443
Vita CR System Troubleshooting Tests

Section 7: Vita CR System Troubleshooting Tests

Sensors Tests
Introduction
Perform this test to check if a sensor is functioning. This test is valid for all sensors.

Procedure
[1] Remove the CR Scanner cover.
[2] Turn the CR Scanner ON.
[3] Go to the Settings>Diagnostics tab.

[4] Insert a piece of paper into the sensor where the flag passes during the procedure.

Note
This blocks the sensor signal. The green LED adjacent to that sensor on the pictogram should light up if the sensor
is functioning properly. If the specific sensor that is blocked does not light up, the sensor is not functioning properly.

[5] Remove the piece of paper to unblock the sensor. The green LED should turn OFF.

8H8443 – 2012-08-13 59
TROUBLESHOOTING GUIDE

sensor is blocked Sensor is unblocked


[6] If the sensor is not functioning properly, replace the sensor and repeat the test.

Note
To perform this test on the W0 or W1 sensor: The sensor is divided into two parts; the top is the transmitter and the
bottom is the receiver. Insert a screen into the drum and check the W0 or W1 sensor LED position change. During
the screen upload, the LED should change from green to OFF.

Controller Fuses Test


Introduction
The controller board has fuses to protect it from high voltage and to prevent short circuits from occurring.

Fuse Number Function Size


F1 5V 1.5 A
F2 -15 V 1A
F3 +15 V 1A
The LED diodes D18, D20, and D21 correspond to F1, F2, and F3. If a fuse is blown, the LED diode will light up
orange. If one of the LEDs is red, replace the corresponding fuse.
The LED is located above its corresponding fuse.
Controller Board LEDs

F1
F2
F3

60 2012-08-13 – 8H8443
Vita CR System Troubleshooting Tests

Controller Board Test Point—Voltages


Introduction
The controller board has test points responsible for checking the current voltages while the system is operating. The
controller board reports the current voltage status to the CR System software application and can be viewed in the
Diagnostic screen.

There are six test points located on the controller board that confirm the power supply voltages and the controller
converting voltage status. These points can be tested using a digital voltmeter (DVM).

Test Points Voltage Voltage Source


TP9 -15 V Power supply
TP13 +15 V Power supply
TP14 5V Power supply
TP11 24 V Power supply
TP10 3.3 V Converted from 5 V power supply in the
controller board
TP2 1.2 V Converted from 5 V power supply in the
controller board

8H8443 – 2012-08-13 61
TROUBLESHOOTING GUIDE

Controller Board Test Points and Diagram

TP2

TP9

TP13

TP10
TP14
TP11

In addition to the test points, the controller board indicates normal voltage status using LED indicators.

(D10)+15 V 5 V (D11)
(D13)3.3 V 24 V (D12)
-15 V (D17)
(D11)+5 V

If one of the voltages does not fall within the ±10 % definition range listed in the Power Supply section in the
Diagnostic screen, the LED indicators on the controller board will not light up.

Voltage Voltage Source Replace


-15 V Power supply ±15 V and 5 V power supply
+15 V Power supply ±15 V and 5 V power supply
5V Power supply ±15 V and 5 V power supply
24 V Power supply 24 V power supply
3.3 V Converted from 5 V power supply Controller board
in the controller board

62 2012-08-13 – 8H8443
Vita CR System Troubleshooting Tests

Voltage Voltage Source Replace


1.2 V Converted from 5 V power supply Controller board
in the controller board
+5V While connected to the PC USB cable

±15 V and 5 V power supply


24 V power supply

8H8443 – 2012-08-13 63
TROUBLESHOOTING GUIDE

Controller Board Dip Switch and LED Indication


Introduction
There are LED indicators on the controller board that indicate the system’s components status.
These LED indicators are controlled by DIP switches. The DIP switches turn the LED indicators ON and OFF in order
to prevent any light from penetrating the system during a scan cycle.
To turn ON the LED indicators, move the DIP switch to the ON position.

Important
The LED indicators should be turned ON using the DIP switch only when testing the controller board function.
Make sure to turn the LED indicator(s) OFF following the test to prevent any light from penetrating the system during
a scan cycle.

LED Indicator DIP Switch Indication ON Stats Indication


D10 SW3 +15 V Steady green
D11 SW3 5V Steady green
D12 SW3 24 V Steady green
D13 SW3 3.3 V Steady green
D16 SW3 5V Steady green
D17 SW3 -15 V Steady green
D19 SW1 CPU SW loading Blinking green
Firmware
D6 SW1 Motion SW Blinking green
loading FPGA
D2 SW1 ALTERA SW Blinking green
loading
No LED SW2 (push Reset the None
button) controller board

Note
To initialize the system, plug in the interlock key and then press the Reset button once.

64 2012-08-13 – 8H8443
Vita CR System Troubleshooting Tests

DIP Switch Locations

D2
D6

SW1

SW2
SW3

D19

8H8443 – 2012-08-13 65
TROUBLESHOOTING GUIDE

66 2012-08-13 – 8H8443
Electric Schematic Diagrams

Section 8: Electric Schematic Diagrams


Rotation, PM Schematics

8H8443 – 2012-08-13 67
TROUBLESHOOTING GUIDE

Power Supply, Motors Schematics

68 2012-08-13 – 8H8443
Electric Schematic Diagrams

Sensors Schematics

8H8443 – 2012-08-13 69
TROUBLESHOOTING GUIDE

70 2012-08-13 – 8H8443
Publication History

Publication History

Publication Publication Changed


Date No. Pages File Name Notes
2010-07-28 8H8443 8H8443.fm New Publication
2012-08-13 8H8443 various 8H8443.fm Update to VitaXE
Update using the jig adjustment tool
Image artifact failures added

2012-08-13 – 8H8443 71
CARESTREAM and DIRECTVIEW are trademarks of Carestream Health, Inc.

Printed in U.S.A.
Carestream Health, Inc.
Rochester, N.Y. 14608
CARESTREAM is a trademark of Carestream Health.
© Carestream Health, Inc., 2012

72 2012-08-13 – 8H8443

You might also like