Professional Documents
Culture Documents
I. INTRODUCTION
instruments when connected to the same signals produce percentile value (i.e., that value which is exceeded for 1% of
matching results within the specified accuracy shown in Table the measurement period) should be calculated for each 24
1. hour period for comparison with the recommend limits in the
revision. For the Tsh interval measurements, the 95th and
Table 1 – Accuracy requirements for current, voltage and 99th percentile values (i.e., those value which are exceeded
power measurements for 5% and 1% of the measurement period) are calculated for
each 7 day period for comparison with the recommended
limits. These statistics should be used for both voltage and
current harmonics and interharmonics with the exception that
the 99th percentile Tsh value is not recommended for use with
voltage harmonics.
VII. REFERENCES
[1] IEC 61000 4-7, Electromagnetic compatibility (EMC) – Part 4-7: Testing
and measurement techniques – General guide on harmonics and
interharmonics measurements and instrumentation, for power supply
systems and equipment connected thereto.
[2] IEEE Std. 519-2000 Draft 0 - Recommend Practices and Requirements for
Harmonic Control in Electric Power Systems.
V. IMPLEMENTATION ISSUES
The omission of the required anti-aliasing filter defined in
the standard is one of the most often observed reasons why
harmonic / interharmonic measurements (especially current
measurements near power electronic devices) fail to compare
with a known good reference instrument. The mains
synchronization mechanism is another common source of
measurement error when not properly implemented (or
omitted).
The IEEE 519 draft also specifies how the data obtained
using the 61000-4-7 method and aggregated as described
above is to be evaluated statistically. The Tvs and Tsh values
should be accumulated over periods of one day and one week,
respectively. For Tvs interval measurements, the 99th