You are on page 1of 13

HYBRID PICA

SYSTEM DATA

Hybrid PICA
brings a new dimension to MR Imaging.

TIME MEDICAL SYSTEMS


Vision Beyond Imagination.

Hybrid

TABLE OF CONTENTS

Specifications

Site Planning Room Requirements Typical MRI Floor Plan Chassis & Patient Table Magnetic Fringe Field

7 9 10 11 12

Performance Highlights Staff Productivity Patient Comfort System Reliability

13 15 17 19

Radio Frequency (RF) Coils

21

Hybrid

Specifications

Magnet System Type: Permanent, active shims Material: Neodymium Ferrite Boron (NdFeB) Field Strength: 0.35 Tesla Opening: 400 mm Fringe Field: 5 Gauss line 2.5 m axial x 2.5 m radial See diagrams on page 14. Size: 1770 x 2220 x 1590 mm (L x W x H) Field Stability: < 0.2 ppm/hr Magnet Weight: 17,000 kg / 37,480 lbs Gradient System Amplitude: 25 mT/m Rise Time: 0.5 ms Slew Rate: 50 T/m/s upgradable to Amplitude: 33 mT/m Rise Time: 0.37 ms Slew Rate: 90 T/m/s

RF System Frequency: 15 MHz 200 kHz RF Power: 6000 W peak rms, 600 W average Image BW: 3.3 MHz Preamp Noise Figure < 0.5 dB Pre-amplifier integrated in all receiving coils Automatic receiving coil recognition RF Receive Coils Quad Head Coil Quad Neck Coil Quad Shoulder Coil Quad Wrist Coil Quad Body Coil Quad Knee Coil Quad Foot/Ankle Coil Large FOV Spine Coil Pulse Sequences 2D Spin Echo, 2D & 3D Gradient Echo, FSE, IR-FSE, STIR, FLAIR, Turbo-FLAIR, 2D & 3D MRA-TOF.

Computer System Host Computer: Apple Mac Pro with Quad Core processor, multi-processing, 4 GB RAM, 2 TB RAID hard drive OS: Windows XP Professional Imaging Software: Prodiva MRI Platform with Patient Flow Technology Data Archive: DVD-RW (DL) 9.4 GB or Blue-ray 25 GB (optional) Display: Apple Cinema HD LCD Monitor 24" (1920x1200) Data Transfer & Handling (HIS/RIS): DICOM 3.0 Printing Support: DICOM Print, All Windows Printer & PDF Patient Table Size: 2600 x 800 x 767 mm (L x W x H) IN/OUT movement 2000 mm LEFT/RIGHT movement 120 mm each side UP/DOWN movement 200 mm Min height from floor 550 mm Loading capacity 200 kg / 440 lbs

Imaging Parameters FOV: 5 cm 42 cm, 0.1 mm increments Slice Thickness: 2D: 1 mm 10 mm, 0.1 mm increments 3D: 0.25 mm 10 mm, 0.1 mm increments Slice Spacing: 0.1 mm increment Min / Max Phase Matrix: 64 / 512 Min / Max Frequency Matrix: 64 / 512 Bandwidth: 5 kHz 3.33 MHz Rectangular FOV (increment in % in full FOV): 0.1% Types of Recon filters: Ring Filter (On/Off), Histogram Equalization, Image Uniformity Image Reconstruction Multithreaded reconstruction 2D > 1000 images/s, 256 x 256 3D > 200 planes/s, 256 x 256 x 64 Image Visualization Features: OpenGL based visualization with auto Window/Level, move, zoom, multiple image display, ROI, annotate, measurements, multi-planar reformat, real-time MIP, CINE, multiple viewport.

Hybrid

Site Planning

Hybrid

Room Requirements

Typical MRI Floor Plan

Specification
Recommended room size (inside dimensions) *

Equipment Room
3000 x 3000 mm (9.8 9.8 ft) 2800 mm

Magnet Room
7000 x 5000 mm (23 16.4 ft) 3300 mm

Control Room
3000 x 4000 mm (9.8 13.1 ft)

Minimum ceiling height (9.2 ft) Floor requirements Total Floor Loading antistatic 417.7 lbs / sqft Floor Leveling 5 mm / 3 m 90 dB attenuation over RF shielding Magnetic field shielding not required not required 3 x 110 Vac or 220 Vac 3 x 110 Vac or 220 Vac Power outlets (depends on country) (depends on country) 1 x 380 Vac (3 phase) Air conditioning Humidity without condensation Network outlets 1 (directly connected to Control Room) without condensation No requirement 3 kW 40%-70% 15 kW 40%-70% 10 100 MHz depends on site (11 ft) antistatic, level 20 kPa

no requirement

antistatic -

not required

not required at least 6 x 110 Vac or 220 Vac (depends on country) 3 kW 40%-70% without condensation 1 (directly connected to Equipment Room)

* Remarks: The minimum space requirements vary based on the specific site survey.

10

Hybrid

Chassis & Patient Table Dimensions

Magnetic Fringe Field

(Front)

(Side)

11

12

Hybrid

Performance Highlights

13

14

Hybrid

Staff Productivity

Scan

View Mode

View

Manage

Scan Mode

Print Mode

15

16

Hybrid

Patient Comfort

Six Direction Automated Patient Table

Its ok Teddy, the PICA is wide-open. And the scan only takes 20 minutes.

UP/DOWN travelling distance 200 mm

IN/OUT travelling distance 2000 mm

LEFT/RIGHT travelling distance 120 mm

Laser Positioning Device 1 mm accuracy

Patient Table
Cooling Coolant thermal loading Air 25 W Ambient (-10C to +40C) 450 kg (992 lbs)

Unit weight

Patient loading weight

200 kg (440 lbs)

Shipping weight

600 kg (1322.8 lbs)

Power Input

DC 24V - 40A

17

18

Hybrid

System Reliability

System / Power Supply Cabinets


Cooling Coolant thermal loading Operating temperature Line voltage AC power requirements Dimensions (LWH) Unit weight Shipping weight Air cooled with internal fans 2.0 kW / 2.2 kW 21 3C, limits 17C and 27C 380 Vac 5%, 3N~ , 50/60 Hz 63 A / 42 kVA 1000 x 600 x 1549 mm / 640 x 900 x 975 mm (39.4 x 23.6 x 61.0 in.) / (25.2 x 35.4 x 38.4 in.) 375 kg (827 lbs) / 390 kg (860 lbs) 420 kg (926 lbs) / 420 kg (926 lbs)

19

20

Hybrid

Radio Frequency (RF) Coils

Item 1. Coil type 2. Tuning Method 3. Decoupling Quad Head Array Coil Quad C-Spine Array Coil 4. Q Factor 5. Uniformity for Optimal IQ 6. Connector

Specification 2 channel, Receive only Passive Passive & active >100 >90% Mixed D type

Remark Integrated preamp

From transmission coil F = 15 MHz 200 kHz, loaded

Coils 1. Quad Head Array Coil Quad Body Array Coil 2. Quad Cervical Spine Array Coil 3. Quad Body Array Coil: Quad Shoulder Array Coil Quad Wrist Array Coil 20 coil 17 coil 14 coil 4. Quad Knee Array Coil 5. Quad Foot / Ankle Array Coil 6. Quad Wrist Array Coil Large FOV Spine Coil 7. Quad Shoulder Array Coil 8. Large FOV Spine Coil

FOV (L x W x H mm) 260 x 215 x 250 180 x 190 x 215 300 x 550 x 325 300 x 420 x 280 300 x 350 x 240 250 x 180 x 180 180 x 115 x 175 130 x 115 x 70 195 x 200 x 215 200 x 250 x 280

Quad Knee Array Coil

Quad Ankle Array Coil

21

22

Hybrid

23

24