You are on page 1of 24

WenHsin Huang

Principles of Chromatography
Principles of
Chromatography •Introduction to Analytical Separations
Introduction to –Solvent Extraction
Analytical Separations –What is Chromatography?
–A Plumber’ s View of Chromatography
–Efficiency of Separation
Wen-Hsin Huang
–Why Bands Spread
Ph. D.
NDMC

2006/9/7 WenHsin Huang, NDMC 1 2006/9/7 WenHsin Huang, NDMC 2

Simple Separations Solvent Extraction


Solute particles Solute particles Solute particles
•Extraction only have one now have two choose preferred
–transfer of solute from one phase to another choice of solvent choices of solvent solvent
–phase can be gas, solid, liquid
•Liquid/liquid extraction
–2 immiscible solvents used
add second
–typically aqueous solvent and organic solvent immiscible
solvent
•you know water and oil don’
t mix
–organic solvents less dense than water
Legend: shake
•diethyl ether, toluene, hexane aqueous solvent
–organic solvents more dense than water organic solvent
•chloroform, carbon tetrachloride, dichloromethane solute particles
2006/9/7 WenHsin Huang, NDMC 3 2006/9/7 WenHsin Huang, NDMC 4

NDMC 1
WenHsin Huang

Solvent Extraction Phase Partitioning


If we have a solute (S) that is partitioned between two
phases (1 & 2) then we can write an equilibrium expression
•Like dissolves like for this equilibrium.
–solute chooses solvent most like itself
–polar compounds (and ionic compounds) S1  S 2
choose water (which is very polar)
–nonpolar compounds choose nonpolar organic AS [S ]
solvents K 2  2
AS1 [ S1 ]
Partition Coefficient (K)

Convention: Organic phase = phase 2
2006/9/7 WenHsin Huang, NDMC 5 2006/9/7 WenHsin Huang, NDMC 6

Distribution Coefficient Example of Extraction with n-Hexane


The distribution coefficient for compound Z between
The distribution coefficient (D) is used in place of the n-hexane and water is 6.25. Calculate the percent Z
partition coefficient when dealing with a species that has remaining in 25.0 mL of water that was originally
more than one chemical form. 0.0600 M in Z after the extraction with the following
volumes of n-hexane.
A) One 25.0-mL portion 13.8%
total _ concentration _ in _ phase _ 2
D B) Two 12.5-mL portions 1.90%
total _ concentration _ in _ phase _ 1
C) Five 5.00-mL portions 4.99 x 10-3%
D) Ten 2.50-mL portions 2.49 x 10-7%

It is more efficient to do several small extractions


rather than one large extraction.
2006/9/7 WenHsin Huang, NDMC 7 2006/9/7 WenHsin Huang, NDMC 8

NDMC 2
WenHsin Huang

Solvent Extraction (pH effects) Solvent Extraction (pH effects)


•Charges of acidic and basic species change •Distribution of HA between two solvents
with pH –two equilibria to consider
–neutral species generally more soluble in •Ka equilibrium
organic solvent
•D equilibrium
–charged species generally more soluble in
aqueous solution
organic HA H+ + A-
•Distribution coefficient (D) describes
distribution of species between two phases
–Takes into account all forms of a compound Ka
(i.e. H2A, HA-, A2-) aqueous HA H+ + A-
–Different from partition coefficient (K)
2006/9/7 WenHsin Huang, NDMC 9 2006/9/7 WenHsin Huang, NDMC 10

Solvent Extraction (pH effects) Solvent Extraction (pH effects)


•Distribution of HA between two phases
•Distribution of acids and bases between two
phases
[H][ A ] K [HA ]
total conc. in phase 2 (organic) Ka   [ A ]  a 
[HA ] [H ]
D=
total conc. in phase 1 (aqueous) •Also have K expression for [HA]

[HA]2 [B]2 [HA ]2


D= D= K
[HA]1 + [A-]1 [B]1 + [BH+]1 [HA ]1
•Substitute and rearrange to get
K[H]
•Ionic species only found in aqueous layer D 
[H ] K a Eq’
n

2006/9/7 WenHsin Huang, NDMC 11 2006/9/7 WenHsin Huang, NDMC 12

NDMC 3
WenHsin Huang

Solvent Extraction (pH effects) Solvent Extraction (pH effects)


D for an acid (HA) D for a base (B) •Adjust pH to effectively extract an acid or
base (and its conjugates) into aqueous layer
K[H] KK a •Consider base B with a pKa of 9.0
D  D Eq’
ns
[H ] K a K a [H] –pH < 9.0, B hydrolyzes so majority of B (in the
form BH+) in water layer (low D value)
•Distribution of acids and bases between two 0

phases is pH-dependent [ B]
K  org -2

log D
[ B ]aq pKa
•To find D, need to know
[ H ][ B]aq -4
–pH of solution Ka  mainly mainly
[ BH ]aq -6 BH+ B
–Ka of acid or conjugate acid [ B ]org K Ka
D  KB
–K of acid or base [ B ]aq [ BH ]aq K a [ H ] 2 4 6 8 10 12
2006/9/7 WenHsin Huang, NDMC 13 2006/9/7 WenHsin Huang, NDMC
pH Fig.14

History of Chromatography Tswett’


s Experiment
Mikhail Tswett is credited with the
invention of chromatography. Ground up plant leaves and added
petroleum ether.
Developed a technique that separated
various plant pigments (e.g.,
chlorophylls and xanthophylls) by
passing solutions through glass Filled column with chalk.
columns filled with finely ground
M. S. Tswett
CaCO3.
Russian Scientist
(1872-1919)
The separated species appeared as Method forgotten for many years.
colored bands on the column.

Hence Greek chroma = color and graphein = writing


2006/9/7 WenHsin Huang, NDMC 15 2006/9/7 WenHsin Huang, NDMC 16

NDMC 4
WenHsin Huang

What is Chromatography?
Tswett’s
•Method to separate components in a
Experiment mixture based on different distribution
coefficients between the two phases
•Same principle as solvent extraction (like
dissolves like), but one phase is “
stationary”
and one phase is “ mobile”
–no longer working with organic and aqueous
layers in a separatory funnel
–working in a column (which is just a tube)
2006/9/7 WenHsin Huang, NDMC 17 2006/9/7 WenHsin Huang, NDMC 18

What is Chromatography?
Types of Chromatography
•Two phases: mobile and stationary
•Mobile phase is solvent moving through •Adsorption Chromatography
column
•Partition Chromatography
–liquid (Methanol, water, buffer)
•Ion-exchange Chromatography
–gas (He, H2, N2)
•Molecular Exclusion Chromatography
•Stationary phase fixed inside column
–viscous liquid coated on inside of column •Affinity Chromatography
–solid particles packed inside column
•Solutes have different affinities for mobile
phase and stationary phase
2006/9/7 WenHsin Huang, NDMC 19 2006/9/7 WenHsin Huang, NDMC 20

NDMC 5
WenHsin Huang

Adsorption Chromatography Partition Chromatography


Stationary phase Mobile phase Stationary phase Mobile phase
Solid Liquid or Gas Liquid on solid support Liquid or Gas

How it separates How it separates


–solute adsorbs on to –solute dissolves into
stationary phase surface liquid coating

2006/9/7 WenHsin Huang, NDMC 21 2006/9/7 WenHsin Huang, NDMC 22

Molecular Exclusion
Ion-Exchange Chromatography
Chromatography
Stationary phase Mobile phase
Anions or cations covalently Liquid Stationary phase Mobile phase
bound to solid stationary phase Porous gel Liquid

How it separates
–small molecules
How it separates
trapped in pores of
solute ions of opposite charge stationary phase
attracted to stationary phase Also called
gel filtration,
gel permeation,
or
size-exclusion
chromatography
2006/9/7 WenHsin Huang, NDMC 23 2006/9/7 WenHsin Huang, NDMC 24

NDMC 6
WenHsin Huang

Classification Specific Method Stationary Phase Type of equilibrium


Liquid Liquid-liquid or Liquid adsorbed onto Partition between immiscible
Affinity Chromatography chromatography
(LC) (Mobile
partition solid liquids

phase a liquid) Liquid-bonded Organic species bonded Partition between liquid and bonded
phase to solid surface surface
Stationary phase Mobile phase
Liquid-solid, or Solid Adsorption
Immobilized molecules on Liquid adsorption
liquid or solid stationary phase
Ion exchange Ion-exchange resin Ion exchange

Size exclusion or Interstices of polymeric Sieving


How it separates molecular- solid
exclusion
–Molecules with specific
Affinity Covalently bonded SelectiveAf f
ini
ty–( egs pe ci
fi c
shape dock with ligands Chromatography molecule (eg antibody) Protein/antibody bind together)

Gas Gas-liquid Liquid adsorbed onto Partition between gas and liquid
chromatography solid surface
(GC) (Mobile
phase a gas) Gas-bonded Organic species bonded Partition between gas and bonded
phase to solid surface liquid

Gas-solid Solid Adsorption


2006/9/7 WenHsin Huang, NDMC 25 2006/9/7 WenHsin Huang, NDMC 26

Basic Principles Simple Column


Two phases considered:
Fresh eluent

1) Mobile Phase: solvent moving through the column. Initial band


of A and B

2) Stationary Phase: stays in place inside of the column.


Column
Packing


Eluent” COLUMN “Eluate”
Porous Disk

Process is called “
elution”

2006/9/7 WenHsin Huang, NDMC 27 2006/9/7 WenHsin Huang, NDMC 28

NDMC 7
WenHsin Huang

Chromatogram Chromatographic Terms


A chromatogram is a
graph showing the tm = “dead time”which is
detector response the minimum time for
(proportional to mobile phase to pass
through the column
concentration) as
solutes emerge from a
tr = “retention time”
chromato-graphic which is the time required
column as a function of for the solute (analyte) to
time or volume. pass through the column.
Detector
t’r= “adjusted retention
time”describes time

All solutes spend equal time in mobile phase solutes spend in stationary
phase
t rt r t m
2006/9/7 WenHsin Huang, NDMC 29 2006/9/7 WenHsin Huang, NDMC 30

Chromatographic Terms A Plumber’


s View…
•kcapacity factor describes how long
= “relative retention” solute retained on column
which is the ratio of

t t
adjusted retention times.
time spent in stationary phase
t k'  r m =
 r 2 tm time spent in mobile phase
t r
1
>1 always
then separation
•Higher kindicates solute retained longer
between two components
•If k= 0, solute unretained
k’= “capacity factor” •If k= 1, solute spent same amount of time
time _ in _ stationary t r in stationary phase as in mobile phase
k
 
time _ in _ mobile tm
2006/9/7 WenHsin Huang, NDMC 31 2006/9/7 WenHsin Huang, NDMC 32

NDMC 8
WenHsin Huang

A Plumber’
s View… Partition Coefficient Relationship
•kdescribes time spent in two phases
time _ in _ stationary moles _ in _ stationary tr
•kalso describes number of moles of solute k
  
time _ in _ mobile moles _ in _ mobile tm
partitioned between two phases
CV [S ] V V t
k'  s s Eq’
n k
 s s K s  r
Cm Vm [ S ]m Vm Vm t m
•K Partition coefficient compares t k  K
concentration of solute in one phase to  r 2  2  2
t r
1 k1 K1
concentration of solute in the other
[ S] C V
K  2  s  k' K s
What this says is that analytes with different partition
Eq’
n coefficients will have different retention times.
[S]1 Cm Vm
2006/9/7 WenHsin Huang, NDMC 33 2006/9/7 WenHsin Huang, NDMC 34

A Plumber’
s View… Review
Partition coefficient describes C
•q fraction of solute in mobile phase K s
conc. of solute in SP and MP Cm
molesm Cm Vm
q 
moless molesm Cm Vm Cs Vs Capacity factor describes t t V
1 1 time solute spends (or moles of k'  r m K s
q  solute) in SP and MP
tm Vm
CV V
1 s s 1 K s
Cm Vm Vm
Relative retention compares
1 k' retention times (or partition t ' k' K
q (1 q)   r 2  2  2
1 k' 1 k ' coefficients or capacity factors) t r1' k '1 K 1
fraction in MP fraction in SP of solute 1 and 2
2006/9/7 WenHsin Huang, NDMC 35 2006/9/7 WenHsin Huang, NDMC 36

NDMC 9
WenHsin Huang

Chromatographic Peaks Resolution of peaks


•Solutes elute from a column in a Gaussian peak
shape Resolution is a measure t ( t t )
of how well two peaks are resolution  r  r1 r 2
–w width of peak at base separated. w av ( w 1 w 2 ) / 2
–w1/2 width of peak at half-height
R=0.50 R=0.75
As tr grows,
average
resolution
retention time
between two
 
h peaks 
w1/2=2.35
improves.
1/2h R=1.00 R=1.50
w=4
time
t0 tr
Peak width (w) is defined as the baseline width (4).
2006/9/7 WenHsin Huang, NDMC 37 2006/9/7 WenHsin Huang, NDMC 38

Plate Height (H)


Resolution Example
Bands broaden as they travel through the column.

Two solutes have retention times and widths of tr1 =


235 s, w1 = 8 sec; tr2 = 250 s, w2 = 10 s.

Resolution = 1.7

Plate height (H) relates the amount of broadening to


the linear distance traveled.
2
H
H = plate height
x = distance traveled
2 = variance of peak x
2006/9/7 WenHsin Huang, NDMC 39 2006/9/7 WenHsin Huang, NDMC 40

NDMC 10
WenHsin Huang

Evaluating Separation Efficiency Evaluating Separation Efficiency


•“N”in chromatography is analogous to “
n”
•Plate theory in liquid-liquid extractions
–Breaks separation up into many discrete stages
–each extraction represents a theoretical plate
–Stages represent individual equilibria

[S]2
•N (Theoretical plate) represents each
equilibrium between MP and SP
[S]1
•Each time a solute molecule
•Higher N corresponds to better separation
enters the SP from the mobile –indicates solute molecules enter SP a higher
phase is a theoretical plate number of times
2006/9/7 WenHsin Huang, NDMC 41 2006/9/7 WenHsin Huang, NDMC 42

Evaluating Separation Efficiency Chromatographic Peaks


•N is dependent on •Distribution of retention times for a solute
–column •Peak shape represents differences in
–analyte behavior of solute molecules during
–retention time of analyte separation
–width (at base or at half-height) of peak average
2 2 retention time
t 16t  
N  r 2  2r N is dimensionless so h
 w w1/2=2.35
measure tr and w (or w1/2)
2 1/2h
5.55t in same units (i.e. time)
N 2 r w=4
w 1/ 2 time
t0 tr
2006/9/7 WenHsin Huang, NDMC 43 2006/9/7 WenHsin Huang, NDMC 44

NDMC 11
WenHsin Huang

Evaluating Separation Efficiency Evaluating Separation Efficiency


•Do not want solute band to spread out as it
travels through column •H related to N and length of column (L)
•H (Plate height) represents the relationship L Report H in units
between the width of a solute band to the H of length (i.e. cm)
N
distance traveled through column •Resolution related to N
•Lower H corresponds to better separation
(higher efficiency)  k '2 
N 1
R  

 
•H also called HETP 4  1 k'av 
 Eq’
n
–Height Equivalent to a Theoretical Plate –R is dimensionless
–One solute equilibrium between SP and MP is –k
2 = capacity factor for solute retained longer
one theoretical plate (N) on column (higher tr)
–Equilibrium occurs in one HETP
2006/9/7 WenHsin Huang, NDMC 45 2006/9/7 WenHsin Huang, NDMC 46

Evaluating Separation Efficiency Number of Theoretical Plates (N)


N 1  k '2 
R  
 

The number of theoretical plates (N) increases as
4   1 k ' av 
Eq’
n the efficiency of the separation increases.

•Increase L to increase R
16t 2 t 2
L
H N
L N  2r  r 2 Derived

N H w 
•Increase to increase R or
t ' k' K
 r 2  2  2 5.55t 2 t 2
t r1' k'1 K 1 N  2 r  r2
–to change , must change nature of MP and SP w1 
2
to change relative affinity analytes have for SP
2006/9/7 WenHsin Huang, NDMC 47 2006/9/7 WenHsin Huang, NDMC 48

NDMC 12
WenHsin Huang

Example Factors Affecting Resolution


The greater the resolution, the better the separation.
The more theoretical plates, the better the separation.
A solute with a retention time of 302 s has a width of
11 s on a column that is 15 m long. Find both N and N   k 2 
1 
H for this separation. R  

 

4    
 ave 
1 k

R N R  L
N = 1.2 x 10-4
H = 1.2 mm
 1 R0
k 2 0 R0

2006/9/7 WenHsin Huang, NDMC 49 2006/9/7 WenHsin Huang, NDMC 50

Principles of Chromatography Why Do Bands Broaden?


Solute invariably spreads apart as it travels
•Introduction to Analytical Separations (diffusions) through a chromatographic
–Solvent Extraction column.
–What is Chromatography?
–A Plumber’ s View of Chromatography The observed variance is a sum of variances
from all the broadening mechanisms.
–Efficiency of Separation
Why Do Bands Spread
obs
2
i2

2006/9/7 WenHsin Huang, NDMC 51 2006/9/7 WenHsin Huang, NDMC 52

NDMC 13
WenHsin Huang

Where Does Broadening Occur? Where Does Broadening Occur?


Inside the column:
Outside the column: Multiple paths
Width of injection plug Longitudinal diffusion
Mixing in detector dead volume Equilibration time

Pump Injector Pump Injector


I I
Column Detector Column Detector

2006/9/7 WenHsin Huang, NDMC 53 2006/9/7 WenHsin Huang, NDMC 54

Multiple Paths (A) Multiple Paths (A)


•Solute molecules can choose many different Band of 3 solute
paths through packed columns molecules traveling Band spreads
through packed column
open tubular packed
column column

layer of SP SP particles
coated on packed inside
inside of column column
time
Occurs only in packed columns.
Compare to solute Compare to solute Smaller particles reduce the plate height.
traveling through traveling through HA
a straw a bean bag • Open tubular columns do not provide multiple paths for solute
during tm (no A term)
2006/9/7 WenHsin Huang, NDMC 55 2006/9/7 WenHsin Huang, NDMC 56

NDMC 14
WenHsin Huang

Longitudinal Diffusion (B /ux) Equilibration Time (Cux)


•S molecules travel through column in a band •Finite time required to allow solute molecules to
•[S] varies throughout band equilibrate between MP and SP
–some solute molecules get ahead of band •If flow rate too high, MP will “ leave behind”
–some solute molecules lag behind band molecules in SP (i.e., Solute in stationary remains
“stuck”while solute in mobile phase moves forward.)
Caused by diffusion of solute in the mobile phase. •Keep SP thin to decrease equilibration time
Faster flow rate means less time, thus less diffusion.
MP slow MP
Direction
start H B/ux so, of travel
SP eq. SP
increase ux (flow rate)
as band bandwidth bandwidth
to reduce problem
travels
2006/9/7 WenHsin Huang, NDMC 57 2006/9/7
H Cux so decrease ux to reduce problem
WenHsin Huang, NDMC 58

Van Deemter Equation Band Spreading


Putting the three factors together yields the Van
Deemter equation that helps predict how the •Three terms in van Deemter equation
column flow rate will affect the theoretical plate correspond to three sources of band
height. B
H A  Cu x spreading or broadening
ux
•A term Multiple Paths term
Multiple Equilibration (also called Eddy Diffusion)
Paths Longitudinal Time
Diffusion •B term Longitudinal Diffusion term

Break equation up into A, B, and C terms
•C term Equilibration Time term
–A H independent of flow rate
–B/ux H inversely proportional to flow rate
–Cux H proportional to flow rate
2006/9/7 WenHsin Huang, NDMC 59 2006/9/7 WenHsin Huang, NDMC 60

NDMC 15
WenHsin Huang

van Deemter plot Column Type Affects H


open tubular packed
•Plot dependence of H on individual terms
column column
•Sum and find optimum ux to use to minimize band
spreading layer of SP SP particles
coated on packed inside
inside of column column
H
B Cux •Advantages •Advantages
ux –useful for large-scale
–high resolution (low H)
Hmin analyses
–high sensitivity
A –low analysis time •Disadvantages
Flow •Disadvantages –A term in van Deemter
uopt rate equation increases H
–not useful for large-scale
2006/9/7 WenHsin Huang, NDMC 61 2006/9/7
analyses WenHsin Huang, NDMC 62

Asymmetric Band Shapes Asymmetric Band Shapes


•Overloaded column (high [S])
–so much S enters SP that the SP starts to look
•Expect Gaussian peak shape from solute S more like S than it looks like original SP
eluting from column, independent of [S]total –most S molecules retained longer on column
•Separation based on K Observed peak
C
K s Band shape in chromatogram
Cm
•In reality, K changes as [S]total changes
•If [S] is too high or too low, peak shape
Majority of S retained Faster-moving (less
deviates from Gaussian shape
on column longer retained) S reaches
detector first
2006/9/7 WenHsin Huang, NDMC 63 2006/9/7 WenHsin Huang, NDMC 64

NDMC 16
WenHsin Huang

Asymmetric Band Shapes Asymmetric Band Shapes


•Underloading (low [S])
•Load less sample to reduce overloading
–“hot sites”on SP more available when less S
problem
molecules trying to enter SP
–some (minority) S get “stuck”on hot sites •Protect groups on SP to reduce number of
–some S retained longer on column
hot sites to reduce tailing problem
Observed peak Constant K (slope) results
Band shape in chromatogram overloaded in ideal peak shape
Cs
tailed

Slower-moving (more
Minority of S retained Cm
retained) S reaches Fig. 23-19
on column longer
2006/9/7 WenHsin Huang, NDMCdetector last 65 2006/9/7 WenHsin Huang, NDMC 66

Chromatographer’s Triangle Quantitative Analysis


Relationship •In general, detectors can tell us
–“
Yes, something is eluting from the column.”
Resolution •Use calibration methods to determine how much
of a compound is eluting from column

Finding Peak Area

Peak Area
•A = ½ wbaseh
•A w½ h
-b
Speed Capacity
x= m •cut-and-weigh
concentration (M) •computer integration
2006/9/7 WenHsin Huang, NDMC 67 2006/9/7 WenHsin Huang, NDMC 68

NDMC 17
WenHsin Huang

Quantitative Analysis Example


•Generally use an internal standard
–internal standard is a known amount of a When 1.06 mmol of 1-pentanol and 1.53 mmol
compound different from analyte of 1-hexanol were separated by GC, they gave
–compare analytical signal from analyte to relative peak areas of 922 and 1570 units,
analytical signal from internal standard respectively. When 0.57 mmol of pentanol was
–not the same as method of standard additions added to an unknown containing hexanol, the
–internal standards discussed in Ch. 4 relative chromatographic peak areas were
Ax area of analyte peak 843:816 (pentanol:hexanol). How much
Ax A As area of standard peak
F s [X] concentration of analyte
hexanol did the unknown contain?
[ X] [ S]
[S] concentration of standard
2006/9/7
F response factor
WenHsin Huang, NDMC 69 2006/9/7 WenHsin Huang, NDMC 70

Liquid Chromatography LC Columns


GUARD

•Analyte
–Soluble in MP (more likely than being volatile) •Constructed of steel or plastic
•Mobile phase •5 –30 cm length
A
–Liquid •1 –5 mm i.d. N
•Methanol, Water, Acetonitrile, Hexane
•Easily contaminated and degraded A
–Plays more active role in separation
•Guard column L
•Stationary phase Y
–Contains same SP
–Liquid (partition) or solid (adsorption) T
–Dust, other particles, strongly adsorbed
•High Performance Liquid Chromatography(HPLC) solutes retained on guard column
I
C
–High pressure used to force solvent through column (as –Expendable A
opposed to gravity) –Extends life of analytical column L

2006/9/7 WenHsin Huang, NDMC 71 2006/9/7 WenHsin Huang, NDMC 72

NDMC 18
WenHsin Huang

LC Columns LC Columns
•Almost exclusively use packed columns •Packing particle diameter (dp) crucial
–Typical particle diameters are 3 –10 m
•Solutes move much slower through liquids
–Decrease particle diameter to decrease H
than through gases
•Provide more uniform flow (low A)
–Time needed to diffuse to SP in open tubular •Less time needed for solute to get to particle (low C)
column is too long 60 10m

40

H (m)
Fig. 21-14 20 5m
S S
10
3m

0 2 4 6 8
2006/9/7 WenHsin Huang, NDMC 73 2006/9/7
Flow WenHsin
rate (mL/min)
Huang, NDMC 74

LC –Adsorption Chromatography
Microporous Silica Particles LC –Adsorption Chromatography
•Solid SP –Silica gel Aggregate of Particles Sponge-like Structure

•Solvent molecules compete with solute molecules


–Pure, spherical, microporous
particles for sites on SP
•Permeable to solvent MP
•Very high surface area Flow SP
•Use when pH < 8 OH OH Solvent
OH
Solute
O Si O Si
Silica gel with silanol O Si O
groups on surface Si O
HO O O
Si Si Fig. 21-10
Si
Deprotonated silanols O Si O •Solute is displaced by solvent molecule
(Si –O-) are “hot sites” OH
OH OH
that can lead to tailing OH
2006/9/7 WenHsin Huang, NDMC 75 2006/9/7 WenHsin Huang, NDMC 76

NDMC 19
WenHsin Huang

LC –Adsorption Chromatography Isocratic vs. Gradient Elution


•Forcible desolvation of solute by solvent is nearly •Isocratic elution
independent of solute identity –One solvent or one constant solvent mixture used as
–Dependent on solvent identity MP throughout entire separation
•Gradient elution
•Eluent strength (
o)
–Adjust solvent mixture through separation
–Ability of solvent to displace solute
–Adjusting eluent strength
•Elutropic series –May speed up separation process
–Relative eluent strengths of common solvents
–Must know about relative polarity of SP/MP •Table 21-2 includes 
o values of solvents for

adsorption chromatography on silica


•Eluent strength increases as MP becomes more
like SP –polarity, o

2006/9/7 WenHsin Huang, NDMC 77 2006/9/7 WenHsin Huang, NDMC 78

Isocratic vs. Gradient Elution Isocratic vs. Gradient Elution


Fig. 21-18
Hexane 
°=0.01 &
Fig. 21-19 •Previous example illustrated balance of good
isocratic resolution and reasonable retention times
elutions
Acetonitrile 
°=0.52
•Gradient elution may help
–Improve resolution
–Shorten retention times (and total analysis time)
Start w/ 100% Benzene –Improve peak shapes
then add Acetonitrile gradient
elution

2006/9/7 WenHsin Huang, NDMC 79 2006/9/7 WenHsin Huang, NDMC 80

NDMC 20
WenHsin Huang

LC –Partition Chromatography
LC Phases
•Liquid SP coated on solid support
–Often, liquid SP covalently attached to surface •Normal-phase Chromatography
of silica gel particle –SP polar
CH3 –MP weakly polar or non polar

Si –More polar solvent has higher 


o
O Si R
•More attracted to SP and able to displace solute
CH3 •Reversed-phase Chromatography
support liquid SP –SP nonpolar or weakly polar
–MP polar
Common polar R groups Common nonpolar R groups
(CH2)3NH2 (CH2)17CH3 (or C18) –Less polar solvent has higher 
o

•More attracted to SP and able to displace solute


(CH2)3CN (CH2)7CH3 (or C8)
(CH2)2OCH2CH(OH)CH2OH (CH2)3C6H5
2006/9/7 WenHsin Huang, NDMC 81 2006/9/7 WenHsin Huang, NDMC 82

LC Phases HPLC Instrumental Design


•Normal-phase developed first •Already covered columns
•Focus now on
•Reversed-phase is more common
–Injector
–Many choices of polar solvents in which solutes –Detectors
are commonly soluble
•Water, methanol Pump Injector
–Nonpolar SP less likely to have hot sites
•Less peak tailing I
Column Detector

2006/9/7 WenHsin Huang, NDMC 83 2006/9/7 WenHsin Huang, NDMC 84

NDMC 21
WenHsin Huang

HPLC 6-way Injection Valve Detectors


Fig. 21-17 •Questions to ask when evaluating an LC
detector:
waste waste
–Is the detection universal?
Sample loop Sample loop
–Is the detector response linear?
–What is the limit of detection (LOD)?
column
–Is the detector useful with gradient elution?
•In other words, is the detector insensitive to solvent
composition?

2006/9/7 WenHsin Huang, NDMC 85 2006/9/7 WenHsin Huang, NDMC 86

UV Detector (w/ flow cell) UV Detector


•Common HPLC detector
–Many solutes absorb UV light
•Will cover during spectrophotometry
Eluate out
–Any UV spectrophotometers discussed previously are
Flow cell
good LC detectors
•UV cutoff wavelengths
Light
source Detector –Below given , MP absorbs UV radiation
–MP absorbance overwhelms analyte absorbance

Eluate in
Fig. 21-20
pathlength
2006/9/7 WenHsin Huang, NDMC 87 2006/9/7 WenHsin Huang, NDMC 88

NDMC 22
WenHsin Huang

UV Detector Fluorescence Detector


•Applicable for UV chromophores •Similar idea to UV detector
Essentially linear •Few molecules fluoresce
–Will discuss deviations from linearity in CH 19 •Derivatize solutes with fluorescent tag
LOD ~ 0.1 ng –Derivatize mixture of solutes before separation
Good detection method with gradient –Derivatize solutes as they elute from column
elution before detection
•Post-column Derivatization
–Use solvents that do not absorb UV radiation

2006/9/7 WenHsin Huang, NDMC 89 2006/9/7 WenHsin Huang, NDMC 90

Refractive Index Detector


Fluorescence Detector
•Compare refractive indices of eluate & reference
Applicable to fluorophores
Essentially linear
–Will discuss deviations from linearity
(Reference) (MP solvent & analyte = eluate)
LOD ~ 0.001 ng
Fine with gradient elution •Cannot use with gradient elution
–Use solvents that do not fluoresce –reference composition cannot exactly match solvent
composition during experiment

2006/9/7 WenHsin Huang, NDMC 91 2006/9/7 WenHsin Huang, NDMC 92

NDMC 23
WenHsin Huang

Refractive Index Detector Comparison of LC Detectors


Type Approx LOD Approx Comments
Linear range
10 -10 g 10 4
Essentially universal UV-VIS For light absorbing
compounds
LOD ~ 100 ng Fluorescence 10-14 g 10 5 For Fluorescent
Essentially non-linear compounds

–Linear range very small MS 10-7-10-9 g 10 5 Universal detector

Disaster with gradient elution Electrochemical -8


10 g/mL 10 5
Specific detector; for
conductometric all ions
Finicky
Electrochemical 10- 10-10 –11 g 10 5 Specfic detector;
–Need constant T amperometric electroactve
compounds

2006/9/7 WenHsin Huang, NDMC 93 2006/9/7 WenHsin Huang, NDMC 94

Liquid Chromatography (LC)


Especially high-performance liquid chromatography (HPLC).

Thet erm“ high-pe rfor


ma nc e”refe r
stot
he
use of packed columns with very small
packing particles (diam. 5-10 m) -
giving greatly enhance resolution.

Note: several types of LC. In


addition to partition (as described
so far) - there are also ion-
exchange and size-exclusion
chromatography using liquid
mobile phases. We will concentrate
only on partition.

2006/9/7 WenHsin Huang, NDMC 95

NDMC 24

You might also like