F.S.D Field Support Dept TML 3005 DSM 700


20 Pages
(This page included)

PLEASE PASS ON THIS INFORMATION

TO ALL YOUR PERSONNEL INVOLVED

file:///W|/2006/Cd Software 2/2 Sensors/dsm700/MEMO/DSM_700.HTM (1 of 9)8/20/2006 8:53:25 PM

TML DSM 700 - HTML

Issued by : F.S.D Date: 01/10/03 Rev: 1

Approved by: R&D and Technology. Date: 01/10/03 Rev: 1

DSM 700

Section 1 DESCRIPTION
1.1. Sensor overview 1.2. DSM700 Sensor + GZG Degasser Overview

Section 2 INSTALLATION
2.1. On GZG 2.2. Electrical Connection

Section 3 SOFTWARE PROLINK


3.1. Software Installation 3.2. Utilisation

Section 4 SPARE PARTS Section 5 APPENDICES


5.1. Converter Frequency / Current 5.2. Communication Cord DSM 700: 700.BR.698

SECTION 1 - DESCRIPTION
1.1. Sensor Overview
The DSM700 will replace the DSM600 , the frame are exactly the same and can be use for the DSM700. The DSM700 is made of a Sensor and a Transmitter mounted on the top of the sensor itself. The transmitter will generate 3 independent signals, 2 currents and 1 pulse. Those 3 outputs can be configure with a matrix which is available via a software (Prolink II version 1.2). The sensor is installed in the mud circuit of the Geoservices degasser GZG to avoid cuttings.

file:///W|/2006/Cd Software 2/2 Sensors/dsm700/MEMO/DSM_700.HTM (2 of 9)8/20/2006 8:53:25 PM

TML DSM 700 - HTML

The measurement of the flow is based on the coriolis effect.

1.2. DSM700 Sensor + GZG Degasser Overview

SECTION 2 - INSTALLATION
2.1. On GZG
1. Switch off the degasser GZG 2. Put the sensor support 5 vertically on degasser frame 1 towards the pump from bottom to top as indicated by the arrow on the side of the sensor. 3. Block the sensor support with nuts. 4. Disconnect the hose-rubber from the degasser tank 2, then connect it to the bottom 3 of the sensor. 5. Connect the output 4 of the sensor on the degasser tank 2 with a hose rubber. 6. Switch on the degasser.

Notes: If impossible to install the sensor on degasser frame 1, hang it close to the degasser, vertically from bottom to top as indicated by the arrow on the sensor. The output 4 of the sensor must be higher than the degasser tank 2. You need to avoid air going inside the sensor when you will switch off the GZG pump.

file:///W|/2006/Cd Software 2/2 Sensors/dsm700/MEMO/DSM_700.HTM (3 of 9)8/20/2006 8:53:25 PM

TML DSM 700 - HTML

2.2. Electrical Connection

The sensor is supplied with 24VDC The sensor has 3 outputs: 1 x pulse 2 x 4-20mA

A serial link is connected on the transmitter to check config the sensor (the default configuration is already done at the Control Department)

The configuration of the sensor is as follow

Output
Current Output 1 Current Output 2 Pulse Output

Configuration
Density Temperature Volumique Flow

Min
0.75 Kg/L 0 Celcius 0

Max
2.75 Kg/L 100 Celcius -

Signal Output
4-20 m A 4-20 m A 100 Hz/10 litres

4-20 mA Ouptut

Those 2 outputs are for measuring the mud weight and the Temperature. The 4-20 mA outputs are passives ones (to be Intrinsic safety). So you need to power them up to recover the 420 mA signal. (you need to connect the output 1 and 2 as a normal 4-20 mA sensor like a Probe sensor).

Connections to the ALS:

Ch 9, Ch 10, Ch 11 Ch 12, Ch 13 UTB Ch 15, Ch 16

1+ 43+ 11+ 2-

file:///W|/2006/Cd Software 2/2 Sensors/dsm700/MEMO/DSM_700.HTM (4 of 9)8/20/2006 8:53:25 PM

TML DSM 700 - HTML

Ch 17 > Ch 24 MEB Channel 4-20 mA

1+ 31+ 2-

Analogical output wiring:

Pulse Ouptut

The pulse output can be use to have the flow signal. The transmitter is sending a frequency signal .This signal can be read on the ALS with a frequency/current converter. The frequency ouput has been configu re to have 100 Hz(100 pulse /sec) for a flow of 10 liter/min.

Ch 9, Ch 10, Ch 11 UTB Ch 17 > Ch 24 MEB Channel 4-20 mA

4+ 33+ 22+ 3-

If you put a 500 Ohm resistor between 14 and 15 of the converter you have a voltage signal so you can connect on channel 1 to 7 like this 4 + 2-

SECTION 3 PROLINK SOFTWARE


Needed Parts: - A computer with win2000 or NT operating system. - Communication Cord DSM 700: 700.BR.698 - DB9F-DB25M interface: 700.AH.053 The setup of the transmitter is done through a computer with the Prolink Software (version 1.2 ) The computer is linked to transmitter via a RS 485 link. This link arrive to RS485-RS232 converter and db25-

file:///W|/2006/Cd Software 2/2 Sensors/dsm700/MEMO/DSM_700.HTM (5 of 9)8/20/2006 8:53:25 PM

TML DSM 700 - HTML

db9 changer which is connected on the Com port of the computer loading with the Prolink software.

3.1. Software Installation


On the CD double click on this button

Choose run this program from its current location then click [OK] Press [YES] Then give the answers bellows during the installation. Prolink II wellcome [next] Prolink II License Agreement [Yes] Prolink II connection information [next] Choose destination location [next] Select program folder [ next] Start copying files [ next] Installshield wizard complete [finish] A new Icon will appear on your desktop :

3.2. Software Utilisation


We will use only a parts of the software. -mapping of the current and pulse outputs. -configuration of the current outputs. -configuration of the pulse outputs. -calibration of the zero flow. -reloading of the original settings. Double click on the Prolink icon to launch the program Then on the tool bar go to Connection\Connect to Device. In Connect to Device Menu Select choose : Protocol : Universal Service Port COM port : Port connect to Black Box converter Then Press on [Connect] button. Attempting connection message will then appear

Connection Not OK Invert the connection going to the Black Box (RS 485 to RS 232 converter)

file:///W|/2006/Cd Software 2/2 Sensors/dsm700/MEMO/DSM_700.HTM (6 of 9)8/20/2006 8:53:25 PM

TML DSM 700 - HTML

Then try again to connect. Connection OK, this Output screen will come up.

Mapping of the current and pulse outputs.

Prolink /configuration/variable mapping

Configuration of the current outputs.

Prolink /configuration/analog ouput

Configuration of pulse outputs.

Prolink /configuration/frequency-discrete output

Calibration of the zero flow.

Prolink/Calibration/zero calibration When you do the calibration of the zero , the sensor need to be full with mud , and mud should not move (pump of the GZG off) . Click on [zero] , the calibration will be perform on the period enter on the Zero time windows (keep it to 20 secondes)

During the calibration , the calibration progress led will come red, as shown below.

When the calibration is finish click on the [DONE] button.

Reloading of the original settings

When the sensor is checked at the Control Department, the configurations (see paragraph 3.2 for sensor configurations) are saved on a file. This file is on the CD and named configuration_master. If you need to reinstall the original configuration follow those steps. -Copy configuration_master file on your PC harddrive -on the Prolink main page go to File/Send to Xmtr from file A windows will open highlight the configuration_master file then click on the button [OPEN] , you will have a progress bar then a similar screen will appear ,click [load transmitter] then the configuration uploading will be
file:///W|/2006/Cd Software 2/2 Sensors/dsm700/MEMO/DSM_700.HTM (7 of 9)8/20/2006 8:53:25 PM

TML DSM 700 - HTML

finish.

SECTION 4 SPARE PARTS

ITEM 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21

ORDER CODE 700.BB.773 700.AB.534 700.AD.535 700.BL.850 700.AD.398 100.CL.050 100.UL.010 100.VL.010 100.NL.010 100.CK.025 100.UK.010 100.VK.010 312.AK.100 312.AK.110 307.AF.013 307.AE.016 302.CC.220 304.CC.820 700.AQ.177 132.AA.160 112.AB.024

1. DESIGNATION Sensor DSM 700 Stainless Steel Frame Flange Gasket flat Pipe connection 1/8 Screw M12x50 Stainless Steel Washer flat 12mm Stainless Steel Washer serrated lock 12mm Stainless Steel Nut Hexagonal 12mm Stainless Steel Screw M10x25 Stainless Steel Washer flat 10mm Stainless Steel Washer serrated lock 10mm Stainless Steel Junction Box 5 Terminals Junction Box 10 Terminals Plug PG13 Polyamide Gland PG16 Polyamide Cable 4x0.75 UL L=100m Cable 12x0.75 UL L=100m Gland M20S Exproof TH 6.5-12mm Hose Rubber D16 Clamping Ring 8mm Serflex

ITEM 22 23 24 25

ORDER CODE 700.AI.760 304.ZB.111 700.AL.489 700.BR.698

DESIGNATION Power supply 24V/5A Power cable IEC 220V Converter Frequency/Current Cord DSM700 communication

file:///W|/2006/Cd Software 2/2 Sensors/dsm700/MEMO/DSM_700.HTM (8 of 9)8/20/2006 8:53:25 PM

TML DSM 700 - HTML

26 27 28 29

700.BR.690 700.AH.053 700.BR.705 700.AB.401

Converter RS485-RS232 DB9F-DB25M Interface Soft CD ROM DSM700 Brush

SECTION 5 APPENDICES
5.1. Converter Frequency / Current

5.2. Communication Cord DSM 700: 700.BR.698