Technical Information TI 193C/24/ae

Conductivity Measurement liquisys M CLM 223 / 253


Conductivity and Resistance Controller
Universal application Conductivity and specific resistance via software switching Transmitters for contacting or electrodeless sensors Ultrapure water compensation acc. to IEC 746 (contacting only) USP 24 Stage 1 compliant Simple handling Logically arranged menu structure with plain text in 6 languages facilitates instrument configuration Large, two-line display indicates measured value and temperature at the same time Calibration via CAL key Safe Operation Overvoltage (lightning) protection according to EN 61000-4-5 Direct access for manual contact control User-defined alarm configuration for alarm contact and error current CSA GP and CE approved The basic unit can be extended with: Due to the modularity of its design, the Liquisys M CLM 223 / 253 controller can be adapted to a wide range of customer needs. The basic version, which provides simple measuring and alarm signaling functions, can be equipped with additional software and hardware modules to match specific applications. Retrofitting of expansion modules is also possible. Areas of application Ultrapure water Water treatment Ion exchangers Reverse osmosis Cooling water desalinization USP and WFI water Benefits at a glance Measuring controller in field or panel-mounted housing 2 or 4 contacts for use as: Limit contacts (also for temperature)

Liquisys M CLM 223

Liquisys M CLM 253

PID controller Timer for simple rinse processes

Complete cleaning with Chemoclean Plus package: Any current output configuration via table Automatic initiation of cleaning in case of alarm or limit violation Ultrapure water monitoring with pre-alarm acc. to USP 24 (contacting only) Polarization detection (contacting only)

Concentration measurement
(% / TDS)

Temperature compensation via table Live check of measuring cell


HART or Profibus PA and DP 2nd current output for temperature

The Power of Know How

Details

Liquisys M CLM 223 / 253 provides a solution for all drinking water processing, process water and waste water treatment applications. Contacting or electrodeless Two instrument versions for measurement with contacting (twoelectrode) measuring cells or electrodeless sensors are available. The use of electrodeless sensors that are less sensitive to soiling than contacting measuring cells is recommended for concentration measurement or adhering media. Features of the basic version (CD / ID): Measurement of conductivity or specific resistance This is selected via the menu. During measurement, the value measured can be displayed in the other measuring mode. The temperature is displayed at the same time if desired. Temperature compensation selections available are linear ( = 0 means no temperature compensation) or special ultrapure water compensation (NaCl curve) according to IEC 746. The reference temperature is 77 F (25 C).
2.4 / 22 mA E 017 E 123 ---on off on off on off

The CS / IS version provides additional functions: Electrodeless sensors must normally be installed in pipes at a required minimum distance from the pipe wall. If this minimum distance is not observed, the measured value changes. The built-in adaptive calibration feature can be used to compensate for this once the sensor is installed.
20

In order to display wide measuring ranges while still achieving a high resolution in certain ranges, the current output can be configured as required via a table. This permits bilinear or quasi-logarithmic curves.

Different alarms are required depending on application and operator. Therefore the Liquisys M CLM 223 / 253 permits independent configuration of the alarm contact and error current for each individual error. Unnecessary or undesirable alarms can be suppressed in this manner.
Timer
on off on off on off

The sensor and process check system also indicates any polarization of the contacting measuring cell. Since there is no subdivision into several measuring ranges (the full range from 0 to 2000 mS/cm is available at all times), the measuring range is limited by the cell constant only. Polarization detection then determines whether or not the working range determined by the measuring cell is exceeded.
20

mA

mA

Chemoclean

The live check issues an alarm when the sensor signal does not change over a defined period of time. This may be caused by blocking, passivation, or separation from the process.

Up to four contacts can be used as limit contacts (also for temperature), to implement a PID controller or for cleaning functions. Direct manual operation of the contacts (bypassing the menu) provides quick access to limit, control or cleaning contacts, permitting speedy correction of deviations.
Instrument DATA

S e r Nr
1356348

I1

The serial numbers of the instrument and modules and the order code can be called up on the display. The cell constant can be edited and even calibrated for demanding special applications. 2

Ultrapure water monitoring according to USP 24 means that the uncompensated conductivity and the temperature are measured and compared to a table. The Liquisys M CLM 223 / 253 (contacting) comes with this function, and the user-adjustable pre-alarm indicates undesirable operating values in due time. Full compliance with USP 24 requires the use of a precisely calibrated measuring cell, for example, the CLS 15. The conversion from conductivity to concentration is effected using four user-definable concentration curves. This permits concentrations to be displayed directly in % or ppm, i.e. as TDS (total dissolved salts solids).

Basic Version and Plus Package

Basic Version Alarm signaling MEASUREMENT CALIBRATION of cell constant CALIBRATION of residual coupling (toroidal sensors) Read instrument DATA Linear CURRENT OUTPUT CURRENT OUTPUT simulation 1 programmable ALARM CONTACT (contact and error current) Additional features Controlling 2 CHANGEOVER CONTACTS for - Measuring parameter limit - Temperature limit - PID controller - Timer for cleaning Additional features Cleaning 2 more CHANGEOVER CONTACTS (total of 4) for - Measuring parameter limit - Temperature limit - PID controller - Chemoclean cleaning (water and cleaning agent) Cleaning Controlling

With plus package (CS / IS version) Alarm signaling Adaptive CALIBRATION (electrodeless sensors) POLARIZATION detection (contacting cells) LIVE CHECK of sensor CURRENT OUTPUT programmable (table)

Additional features

Measurement according to USP 24 with pre-alarm (contacting cells) Automatic cleaning triggered by alarm or limit violation Additional features

Cleaning triggered externally or automatically by alarm or limit violation

Measuring and Control System


Complete measuring systems with Liquisys S CLM 223 / 253

A complete measuring system comprises: the conductivity controller Liquisys M CLM 223 or CLM 253 a conductivity measuring cell with or without integrated temperature sensor
Liquisys M CLM 253
ENDRESS+HAUSER LIQUISYS

an immersion, flow or retractable holder Options: an extension cable CYK 71 (contacting sensors) or CLK 5 (electrodeless sensors) and junction box VBM
Up to 4 relays Alarm contact

CLS 15
REL 1 ALARM CAL REL REL 2 REL 1

REL 2

Signal output 1: conductivity Signal output 2: temperature (optional) Aux. 15 V supply voltage

Immersion holder CYA 611

2 digital inputs Power supply (mains) Measuring Cable

CLS 50 Liquisys M CLM 223


ENDRESS+HAUSER LIQUISYS

REL 1 ALARM CAL REL REL 2 REL 1

REL 2

Measuring Ranges of Conductivity Sensors

Operation
ENDRESS+HAUSER LIQUISYS M

Field for user labeling

m S/ cm

1734
Liquid crystal display for display of measured values, configuration data and current menu field
CAL

Set nt poi
REL 1 ALARM REL REL 2 REL 1

Key for manual relay switching

LED indicator for alarm function

REL 2

User interface: Display and keys

4 main control keys for calibration and instrument configuration

Key for switching between automatic and manual relay operation

LED indicators for switched limit outputs

Everything at a glance The display simultaneously shows the current measured value and the temperature the essential process data. Brief informational texts in the configuration menu provide assistance with parameter configuration.

Intelligent and simple All instrument control functions are arranged in a logical menu structure. Following access code entry, the individual parameters can be easily selected and modified as needed.

Overview of Liquisys M CLM 223 / 253 menu. This covers all the options that can be installed (see page 2 Details).

Electrical Connection

All connections to the panel-mounted instrument CLM 223 are established via the terminal strips on the rear. In the case of the field instrument CLM 253, all wires (including the measuring cell cable) are connected to terminals in a separate wiring compartment.

All the wiring can remain in place if the instrument needs repair because all repairs are limited to assembly replacement. So, dismantling the instrument and rewiring are no longer necessary.

Left: Liquisys M CLM 223, connections on the rear of the instrument Right: Liquisys M CLM 253, rear of instrument with replaceable electronics box (E)

84

Contacting measuring cell (alternative)

83

84

Electrodeless sensor

15 16 83

41 42 43

Alarm*

47 48 49

Relay 1*

11

Temperature sensor Pt 100

12 13

57 58 59

Relay 2*

31

51

Signal output 1: conductivity

mA

52

32

Relay 3*

53

33

54

Signal output 2: temperature (optional)

mA

55

34

Relay 4* *Contact status shown: no current or fault present

56

81

Digital input 1

10 to 50 V
82

optional
~ =

93

L L+ N L

Digital input 2

10 to 50 V
94

Power supply

85

Aux. voltage output Electrical connection of Liquisys M CLM 223 / 253

86

15 V

Dimensions
ENDRESS+HAUSER LIQUISYS

3.78 (96)
REL 1 ALARM CAL REL REL 2 REL 1

REL 2

3.54 (90)

0.24 (6)

5.47 (139)

Dimensions of Liquisys M CLM 223

0.43 (11)

ENDRESS+HAUSER LIQUISYS

REL 1 CAL REL REL 2 REL 1 REL 2

ALARM

6.18 (157) 0.24 (6) 2.75 (70)

9.72 (247)

4.53 (115)

2.75 (70) 6.06 (154) 6.69 (170)

Dimensions of Liquisys M CLM 253

Mounting of CLM 223

The panel-mounted version is mounted using the supplied tensioning screws. The required overall mounting depth is approximately 6.89 in. (175 mm).

Flat gasket (included in delivery)

Mounting cutout according to DIN 43700

3.62+0.02 (92+0.5) 3.54 (90) 2.24 (57)1.30 (33) 3.78 (96) 3.62+0.02 (92+0.5)

Installation of panelmounted housing

approx. 0.98 (25)

5.47 (139)

max. 1.77 (45)

0.24 (6)

Mounting of Liquisys M CLM 253

3.54 (90) 2.76 (70) 1.18 (30) 0.31 (8)

3.54 (90) 2.76 (70) 0.31 (8)

0.24 (M 6)

2.36 (60)

Post mounting kit for installation of field housing on cylindrical pipes (horizontal or vertical, max. 2.36 (60 mm); see Accessories)

2.76 (70) 3.54 (90)

0.24 (6)

Mounting on cylindrical pipes (shown with weather protection cover installed on right)

6.18 (157) 0.31 (8)

Wall mounting Screws: 0.24 (6 mm) Anchors: 0.31 (8 mm)

0.63 (16) 4.53 (115)

Technical Data
General specifications

Manufacturer Product designation

Endress+Hauser Liquisys M CLM 223 Liquisys M CLM 253 conductivity, resistance, temperature

Measured quantities Input Contacting conductivity/resistance measurement Measuring range Usable cell constant Maximum length of cable to measuring cell Measuring frequency

conductivity: 0 to 2000 mS/cm (uncompensated) resistance: 0 to 86 mcm k = 0.0025 to 99.99 cm-1 conductivity: 330 ft. (100 m) resistance: 50 ft. (15 m) conductivity: 300 to 1000 Hz resistance: 30 to 400 Hz

Toroidal conductivity measurement Measuring range Usable cell constant Maximum length of cable to measuring cell Measuring frequency Minimum span for 0 / 4 to 20 mA signal 0 to 2000 mS/cm (uncompressed) k = 0.0025 to 99.99 cm-1 200 ft. (60 m) (CLK 5) 2 kHz meas. value betw. 0 and 199.9 S/cm: meas. value betw. 200 and 1999S/cm: meas. value betw. 2 and 19.99 mS/cm: meas. value betw. 20 and 2000 mS/cm: 20S/cm 200S/cm 2mS/cm 200 mS/cm

Temperature measurement Temperature sensor Measuring range Temperature offset range Temperature compensation Compensation types Range Reference temperature Digital inputs 1 and 2 Voltage Current consumption Output Conductivity/resistance signal output Current range Load Output range Isolation voltage Overvoltage (lightning) protection Temperature signal output (optional) Current range Load Output range Isolation voltage Overvoltage (lightning) protection Auxiliary voltage output Output voltage Output current Contact outputs (floating changeover contacts) Switching current with ohmic load (cos = 1) Switching current with inductive load (cos = 0.4) Switching voltage Switching power with ohmic load (cos = 1) Switching power with inductive load (cos = 0.4) max. 2 A max. 2 A max. 250 V AC, 30 V DC max. 1250 VA AC, 150 W DC max. 500 VA AC, 90 W DC 15 V 0.6 V max. 30 mA 0 / 4 to 20 mA, galvanically separated max. 500 adjustable, 10 to 100% of upper range value max 350 Vrms / 500 V DC acc. to EN 61000-4-5: 1995 0 / 4 to 20 mA, galvanically separated; error current 2.4 / 22 mA max. 500 adjustable max 350 Vrms / 500 V DC acc. to EN 61000-4-5: 1995 10 to 50 VDC max. 10 mA linear, NaCl, table, (with contacting cells only: ultrapure water) -31 to +482 F (-35 to +250 C) 77 F (25 C) Pt 100 -31 to +482 F (-35 to +250 C) 68 F ( 20 C)

Technical Data (continued)


Output (continued) Limit contactor Pickup / dropout delay Controller Function (adjustable) Controller response Proportional band Period for pulse length controller pulse length / pulse frequency controller PID Kp: 0.10 to 10.00 0.5 to 999.9 seconds 0 to 2000 seconds

Frequency for pulse frequency controller


Alarm Function (switchable) Alarm threshold adjustment range Alarm delay Accuracy Conductivity measurement Deviation of indication1 Reproducibility Measurement deviation1, conductivity signal output Temperature measurement Resolution Deviation of indication1 Measurement deviation1, temperature signal output Ambient conditions Nominal operating temperature Ambient temperature (limit operating conditions) Storage and transport temperature Relative humidity (nominal operating conditions) Protection class of panel-mounted unit Protection class of field housing Electromagnetic compatibility

60 to 180 min *

steady / fleeting contact; normally closed / open conductivity / resistance / temperature: complete measuring range 0 to 2000 seconds

max. 0.5% of measured value 4 digits max. 0.2% of measured value 2 digits max. 0.75% of current output range

0.2 F or 0.1 C max. 1.0% of measuring range max. 1.25% of current output range 14 to 131 F (-10 to +55 C) -4 to 140 F (-20 to +60 C) -13 to 149 F (-25 to +65 C) 10 to 95%, non-condensing NEMA 3S (IP 54 front, IP 30 housing) NEMA 4X (IP 65) interference emission acc. to EN 50081-1: 1992 interference immunity acc. to EN 50082-2: 1995 3.78 x 3.78 x 5.71 in. (96 x 96 x 145 mm) approx. 6.89 in. (175 mm) 9.72 x 6.69 x 4.53 in. (247 x 170 x 115 mm) max. 1.54 lb. (0.7 kg) max 5.07 lb. (2.3 kg) LC display, two lines, five and nine digits, with status indicators

Physical data / design

Dimensions of panel-mounted unit (H x W x D) Installation depth Dimensions of field housing (H x W x D) Weight of panel-mounted unit Weight with field housing Display

Materials Housing of panel-mounted unit Front membrane Field housing Power requirements Supply voltage Power consumption Fuse protection polycarbonate polyester, UV-resistant ABS PC Fr 100 / 115 / 230 V AC +10 / -15%, 48 to 62 Hz 24 V AC/DC +20 / -15% max. 7.5 VA fine-wire fuse, medium time-lag, 250 V / 1 A

* acc. to IEC 746-1, for nominal operating conditions Subject to modifications.

10

Accessories
Mounting accessories Type Features Pendulum For pendulum suspension of assemblies frame e.g. CYA 611 Material: 304 SS Post mounting Kit for mounting of field housing on kit horizontal or vertical pipes, max. diameter, 2.5 (60 mm) Material: 304 SS Order number 50080196

50086842

Holders Type CYA 611

Features Immersion assembly with G 1, G 3/4 or 3/4 NPT thread

Applications Waste water, water

Two-electrode conductivity measuring cells (selection) Type Features CLS 15 Cell constant: k = 0.1 cm-1 or 0.01 cm-1 Process connections: 3/4 NPT, 1-1/2 Tri-clamp CLS 19 Cell constant: k = 0.1cm-1 or 0.01 cm-1 Process connections: 3/4 NPT CLS 21 Cell constant: k = 1 cm-1 Process connections: 1 NPT and 2 Tri-clamp

Applications Reverse osmosis filter, ion exchanger Boiler feed, control Service water, boiler feed water

Electrodeless conductivity measuring sensors Type Features CLS 50 Cell constant: k=2 cm-1 Process connections: G 3/4, 1 NPT 2 ANSI flange incl. 16.4 ft (5 m) attached cable CLS 52 Cell constant: k=5.9 cm-1 Process connections: 2 Tri-clamp and Perlick

Applications Concentration measurement, product monitoring, chemical industries Concentration measurement, product monitoring, chemical industries, food industries

Cables Type CYK 71

CLK 5 VBM

Features Standard measuring cable for connection of two electrode measuring cells with and without Pt 100 (also usable for extension) Extension cable for CLS 50 and CLS 52 sensors Junction box NEMA 4X

Order number 50085333

50085473 51500177

11

Ordering Information
Liquisys M 223
1 CLM 223 1 Version CD Conductivity/resistance measurement (two-electrode cell) CS Conductivity/resistance measurement (two-electrode cell) with additional functions ID Conductivity measurement electrodeless (inductive) IS Conductivity measurement electrodeless (inductive) with additional functions Power Supply 0 230 VAC 1 115 VAC 2 230 VAC, CSA general purpose 3 115 VAC, CSA general purpose 5 100 VAC 7 24 VAC/VDC, CSA general purpose 8 24 VAC/VDC Measurement output 0 Conductivity/resistance 1 Conductivity/resistance and temperature 3 Profibus PA 4 Profibus DP 5 Conductivity/resistance with HART 6 Conductivity/resistance with HART and temperature Contacts 05 No additional contacts 10 2 contacts (limit / PID / timer) 15 4 contacts (limit / PID / Chemoclean) 16 4 contacts (limit / PID / timer) 2 3 4 CLM 253 1 Version CD Conductivity/resistance measurement (two-electrode cell) CS Conductivity/resistance measurement (two-electrode cell) with additional functions ID Conductivity measurement electrodeless (inductive) IS Conductivity measurement electrodeless (inductive) with additional functions Power Supply 0 230 VAC 1 115 VAC 2 230 VAC, CSA general purpose 3 115 VAC, CSA general purpose 5 100 VAC 7 24 VAC/VDC, CSA general purpose 8 24 VAC/VDC Measurement output 0 Conductivity/resistance 1 Conductivity/resistance and temperature 3 Profibus PA 4 Profibus DP 5 Conductivity/resistance with HART 6 Conductivity/resistance with HART and temperature Contacts 05 No additional contacts 10 2 contacts (limit / PID / timer) 15 4 contacts (limit / PID / Chemoclean) 16 4 contacts (limit / PID / timer)

Liquisys M 253
1 2 3 4

For application and selection assistance, in the U.S. call 888-ENDRESS For total support of your installed base, 24 hours a day, in the U.S. call 800-642-8737 Visit us on our web site, www.us.endress.com
United States Endress+Hauser, Inc. 2350 Endress Place Greenwood, IN 46143 Phone: (317) 535-7138 888-ENDRESS FAX: (317) 535-8498 Canada Endress+Hauser Canada Ltd. 1440 Grahams Lane Unit 1, Burlington ON, L7S 1W3 Phone: (905) 681-9292 800-668-3199 FAX: (905) 681-9444 Mexico Endress+Hauser Paseo del Pedregal No. 610 Col. Jardines del Pedregal 01900, Mexico D.F. Mexico Phone: (525) 568-2405 FAX: (525) 568-7459

Endress+Hauser
The Power of Know How

TI 193C/24/ae/05.02 2002 Endress+Hauser, Inc.

You might also like