You are on page 1of 9

English

TECHNICAL DESCRIPTION OF THE CATIOVITAL LIFT APPLIANCE Double Ionisation

CATIOVIonisation LIFT ITAL Double


SYSTEM SPECIFICATIONS
General specifications
Manufacturer: Model: Classification: ROSS ESTETICA Y ELECTROMEDICINA S.A. CATIOVITAL LIFT - MARY COHR CLASS 1 TYPE BF EQUIPMENT (COSMETIC APPLIANCE)

Physical specifications
Dimensions: Weight: Length 395 mm / Width 465 mm / Height 230 mm 7.2 kg

Electrical specifications
Voltage: Frequency: Type T fuses: Power: 230 V 60 Hz 0.5 A / 250 V 57 W

Protections
2 0.5 A / 250 V Type T fuses for surge protection. They are located at the base input.

Application specifications
IONISATION: IMPULSE: Maximum current produced 0,3 mA D.C. Maximum output pulse: 20 Vp (without demand) Output frequency: 49 - 150 Hz Pulse amplitude: 140 s 37 V pulsed current Maximum 30 000 / Minimum 9 000

HEAT MASK: IMPEDANCE:

Cleansing
After each use of the electrodes, rinse, clean them with terrycloth toweling moistened with 70 ethyl alcohol and dry with a sterile cloth. After each treatment, clean the mask with terrycloth toweling moistened with dermatological soap and dry immediately with a sterile cloth.

IMPORTANT NOTICE
Before using the CATIOVITAL LIFT appliance be sure to obtain a special authorization for heart pacemaker patients because they are subject to a certain risk caused by the interferences or deterioration of the pacemaker. Never simultaneously use the CATIOVITAL LIFT equipment on patients undergoing other high frequency treatments. In addition, do not use this appliance at a distance of 2 meters or less from short wave or microwave equipment. In case of breakdown, contact the technical service nearest you, or ROSS ESTETICA, S.L. C/ Del Molinot 46-48 Pol.Ind. Cami Ral - 08860 Castelldefels (Espaa) - Tl : 00 34 93 42 66 017

THE APPLIANCE CATIOVIonisation LIFT ITAL Double


The CatioVital Lift appliance, MARY COHR exclusiveness, has three functions.

1. THERMIC FUNCTION
Thanks to the hot mask: in-depth cleansing of the skin with a Sauna Effect
The heat released by the hot mask has direct action on: The sebaceous glands: targets sebum secretion. The sweat glands and canals: activating perspiration, and purifying. The pores: dilating to facilitate the extraction of blackheads.

Results: Freed of impurities, the skin is cleansed, purified and ready to receive the
benefits of the skin care products.

2. IONISATION FUNCTION
Thanks to electrodes which make it possible to treat the skin through the activation of ingredients
Ionisation is ensured by a low-intensity, continuous and direct galvanic current: (+) Pole to (-) Pole. The CATIOVITAL LIFT DOUBLE IONISATION appliance includes 2 electrodes:

1 negative electrode 1 positive electrode


In Phase 2 of the CATIOVITAL treatment the ionised gel serums, selected according to the skins needs, with a polarity compatible with that of the current, are enabled to release their active ingredients (attracted by the opposite charge), for better penetration into the skin.

3. STIMULATION FUNCTION
Thanks to the electrodes that enable the muscular stimulation
Stimulation is ensured by low-frequency variable micro-currents with short impulses and low tension. The gel serum applied is only used for its conducting function. The electrodes are moved according to the specific MARY COHR protocol. The first session is conducted with a low intensity, which can be increased, depending on the clients sensitivity.

Results: Stimulated and invigorated, the skin regains greater firmness tone and volume. The face appears to be visibly lifted and is more youthful-looking.

CONNECTING
AND USING THE APPLIANCE
Plug the machine into a power outlet
Tension: 230 V. The cord is situated in the back of the appliance.

1. THERMIC FUNCTION
Plug in the appliance
Press the switch to position 1 (on the back of the appliance). The green ON/OFF light (A) comes on.

Activate the THERMIC function


Plug the hot mask into the outlet in the space for electrodes. Set the timer, located on the left side of the appliance, to the skin care treatment time (7 minutes) using the ADJ button (B). Press the ON/OFF switch (C), then turn the THERMIC (+) (D) knob to the desired heat (0% to 100%).

Special cases: Thick, oily skin: pre-heat the mask at 100%. Sensitive skin: set to 50%.
The end of the treatment time is signalled by a beep.

Turn off the appliance


Turn the THERMIC (-) (E) knob to 0 then push the ON/OFF switch (C). Press the switch to position 0. The green ON/OFF light (A) goes out. Disconnect the hot mask.

2. IONISATION FUNCTION
Plug in the appliance
Press the switch to position 1 (on the back of the appliance). The green ON/OFF light (A) comes on.

Activate the IONISATION function


Insert the positive electrode (red) and the negative electrode (black) into the dedicated receptacles, in the space for electrodes. Set the timer to the treatment time (10 minutes) desired using the ADJ button (B). Press the ON/OFF switch (F). Apply the gel serum and place the electrodes on the clients face to establish contact. Turn the (+) knob for the IONISATION function (G) (from 0 to 10, depending on the clients sensitivity). Slowly move the electrodes until you hear the beep signalling the end of the ionisation time.

Turn off the appliance


Press the switch to position 0. The green ON/OFF light (A) goes out. Disconnect and clean the electrodes.

Interruption/Pause
You can adjust the intensity of the current during the treatment by pressing on (+) (G) or (-) (H). To cancel the time displayed, press the STOP button (I) then reset it using the ADJ button (B).

DO NOT USE IN THE FOLLOWING CASE


A client with a pacemaker.

3. STIMULATION FUNCTION
Plug in the appliance
Press the switch to position 1 (on the back of the appliance). The green ON/OFF light (A) comes on.

Activate the STIMULATION function


Insert the positive electrode (red) and the negative electrode (black) into the dedicated receptacles, in the space for electrodes. Set the timer for the treatment time (20 minutes) using the ADJ button (B). Press the ON/OFF switch (F). Apply a thick coat of the gel. Place one of the electrodes on the clients face to establish contact. Turn the (+) knob for the STIMULATION function (J) (from 0 to 10, depending on the clients sensitivity). Move the electrodes according to the protocol until the beep signals the end of the stimulation session.

Turn off the appliance


Press the switch to position 0. The green ON/OFF light (A) goes out. Disconnect and clean the electrodes.

Interruption/Pause
You can adjust the power of the stimulating function during the treatment by pressing on (+) (J) or (-) (K), or use the ball side for the more sensitive areas and the flat electrodes for the less sensitive areas. To cancel the time displayed press the STOP button (I), then reset using the ADJ button (B).

DO NOT USE IN THE FOLLOWING CASE


The same as for the ionisation treatment.

Each of our beauty therapists clients who gets a CATIOVITAL LIFT


Double Ionisation

appliance will previously have followed a compulsory training course from our distributor. During this course she will be given a training manual explaining how to use the CATIOVITAL LIFT
Double Ionisation

appliance and the treatment products.